FDA-cleared profile

BriefCase

Aidoc Medical, Ltd.

BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of M1 Large Vessel Occlusion (M1-LVO) pathologies. Briefcase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a Stand-Alone Performance desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of M1 Large Vessel Occlusion (M1-LVO) pathologies. Briefcase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a Stand-Alone Performance desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.
Public source Exact FDA submission Checked 2026-09-01

Exact-submission indication text from the ACR AI Central model card. FDA labeling remains controlling.

Intended users
Hospital networks and appropriately trained medical specialists
Public source Exact FDA submission Checked 2026-09-01

Intended-user text reported in the exact-submission ACR model card.

Care setting and population
Adult
Public source Exact FDA submission Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Triage or notification support
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Required input
CTA images or data. Inclusion criteria: Head CTA protocol with a 64-slice scanner or higher; Scans performed on adults/transitional adults >= 18 years of age; Slice thickness 0.5 mm - 1.0 mm.
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Output and human action
The AI model flags and communicates suspected M1 Large Vessel Occlusion (M1-LVO) cases by sending notifications to a standalone desktop application. Notifications include compressed, low-quality, grayscale, unannotated preview images for informational purposes only. The device provides worklist prioritization to facilitate earlier triage. It does not alter original images or provide diagnostic markings.
Public source Exact FDA submission Checked 2026-09-01

Clinical output reported in the exact-submission ACR model card.

Limitations
Exclusion criteria: All scans that are technically inadequate, including motion artifacts, severe metal artifacts, suboptimal bolus timing or an inadequate field of view.. Scanner manufacturers: All. Slice thickness: Slice thickness 0.5 mm – 1.0 mm
Public source Exact FDA submission Checked 2026-09-01

Limitations and supported-acquisition details reported in the exact-submission ACR model card.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K221314
FDA source Exact FDA submission Checked 2026-08-31
Modality
CT/CTA
Public source Exact FDA submission Checked 2026-09-01

Imaging modality reported in the exact-submission ACR AI Central model card.

Anatomy
Head
Public source Exact FDA submission Checked 2026-09-01

Anatomy normalized from explicit exact-submission ACR model-card wording.

Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2022-06-03
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Standards-based DICOM/PACS integration was identified in the FDA submission materials; compatibility with a specific PACS, RIS, EHR, or orchestrator requires confirmation.
Public source Exact FDA submission Checked 2026-08-13

Normalized from the prior exact-submission review. Confirm interface versions and local compatibility during technical evaluation.

Deployment and data flow
On-cloud
Public source Exact FDA submission Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA authorization does not establish the current commercial security posture. Request data-flow diagrams, hosting and retention details, access controls, audit logging, attestations, and incident-response documentation.

Training and support
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Training, onboarding, competency, and support commitments are commercial implementation details that require current vendor documentation.

Monitoring and change control
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Local acceptance testing, production monitoring, drift review, update control, and incident escalation require a current vendor and site-specific governance plan.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
Study type: Stand-Alone Performance. Cases: 383. Sites: 3. Reported sensitivity: 88.8% (95% CI: 81.9%, 93.8%). Reported specificity: 87.2% (95% CI: 82.5%, 91.1%)
Public source Exact FDA submission Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Reported sensitivity
88.8% (95% CI: 81.9%, 93.8%)
Public source Exact FDA submission Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
87.2% (95% CI: 82.5%, 91.1%)
Public source Exact FDA submission Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

BriefCase-Vessel Occlusion M1-LVO Triage regulatory performance study summary

Stand-Alone Performance · Adult

exact submission · Tested version: Exact product version not reported; regulatory study summarized for FDA submission K221314 · n=383 · 3 sites · Independence not established

K221314

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA submission materials do not establish current commercial pricing. Request the pricing unit, term, minimum commitment, implementation fees, and included support from the vendor.

Reimbursement and coding
Research pending
Needs research Exact FDA submission Checked 2026-08-31

Determine whether any product-specific CPT/HCPCS pathway exists and separate it from billing for the underlying imaging service.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-04Last searched
20Fields reviewed
6Source classes checked
20Unreviewed PubMed leads
6Unreviewed trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Automated exact-name discovery found 20 PubMed and 6 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-02.

Alternatives

Related tools to compare

FDA-cleared K253560

Enhanced Boundary for PCCT

Ge Medical Systems, LLC

Enhanced Boundary for PCCT is a deep-learning CT processing function for improved soft-tissue delineation in routine head examinations with or without contrast.

Modality
CT/CTA
Anatomy
Head
Decision
2026-06-24
FDA-cleared K261317

BriefCase-Triage

Aidoc Medical, Ltd.

FDA product code QAS identifies this device class as Radiological Computer-Assisted Triage And Notification Software. Radiological computer-assisted triage and notification...

Modality
CT/CTA
Anatomy
N/A
Decision
2026-05-14
FDA-cleared K252421

JLK-NCCT

JLK, Inc.

FDA product code QAS identifies this device class as Radiological Computer-Assisted Triage And Notification Software. Radiological computer-assisted triage and notification...

Modality
CT/CTA
Anatomy
N/A
Decision
2026-03-24