FDA-cleared profile

BriefCase

Aidoc Medical, Ltd.

Authorized purpose: BriefCase analyzes frontal chest X-rays in adults or transitional adolescents at least 18 years old to triage suspected vertically malpositioned endotracheal tubes relative to the carina; it flags tip positions more than 5 cm above, less than 2 cm above, or below the carina and requires standard-of-care image review.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: BriefCase analyzes frontal chest X-rays in adults or transitional adolescents at least 18 years old to triage suspected vertically malpositioned endotracheal tubes relative to the carina; it flags tip positions more than 5 cm above, less than 2 cm above, or below the carina and requires standard-of-care image review.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: BriefCase analyzes frontal chest X-rays in adults or transitional adolescents at least 18 years old to triage suspected vertically malpositioned endotracheal tubes relative to the carina; it flags tip positions more than 5 cm above, less than 2 cm above, or below the carina and requires standard-of-care image review.

Evidence takeaway

Evidence from the reviewed source record: A retrospective, blinded, multicenter pivotal study evaluated 921 cases from 5 US clinical sites against a 2-of-3 majority reference of senior board-certified radiologists. Sensitivity was 86.81% and specificity 92.02%; BriefCase time-to-notification was 4.93 minutes in 182 true-positive cases.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
BriefCase analyzes frontal chest X-rays in adults or transitional adolescents at least 18 years old to triage suspected vertically malpositioned endotracheal tubes relative to the carina; it flags tip positions more than 5 cm above, less than 2 cm above, or below the carina and requires standard-of-care image review.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Intended users
Appropriately trained medical specialists in hospital networks; the notification supports triage while full-image review and professional judgment remain required.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Care setting and population
Hospital networks and acute radiology workflows for adults and transitional adolescents aged 18 years and older.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Workflow role
Frontal CXR triage and notification for suspected vertical ETT malposition
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Required input
Frontal chest X-ray in DICOM 3.0 from compatible X-ray scanners; the carina must be well visualized for a result.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Output and human action
Desktop triage notification for suspected vertical ETT malposition with a compressed, unannotated grayscale preview labeled not for diagnostic use; the original image remains unchanged.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Limitations
The device assesses only vertical ETT-tip position relative to the carina, does not factor patient positioning, cannot detect esophageal intubation, provides no result when the carina is not well visualized, and may falsely alert on a properly positioned double-lumen ETT; the notification is informational and not diagnostic.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K221330
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Anatomy
Chest/Lung
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2022-11-18
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact system comprises Aidoc Hospital Server/Orchestrator for image acquisition, Aidoc Cloud Server for processing, and Aidoc Desktop Application for workflow integration. DICOM images are received, filtered, and de-identified before processing, with notifications sent to the desktop application and shared PACS workflows supported.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Deployment and data flow
On-cloud
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07

The FDA packet describes an Aidoc Cloud Server processing component; current hosting region and contractual data-flow controls were not established.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-07

The reviewed sources do not establish a scoped security audit, certification, or control set.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-07

Training, credentialing, and implementation support terms were not established.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-07

Current post-deployment monitoring and change-control practices were not established.

DICOM/PACS/cloud triage workflow

The exact submission describes DICOM 3.0-compatible X-ray scanners, PACS and radiology workstations, Aidoc Hospital Server acquisition, Aidoc Cloud Server processing, and Aidoc Desktop workflow integration with filtered/de-identified image handling.

exact submission · checked 2026-09-07

Current Aidoc radiology workflow context

Aidoc's current official radiology page lists ETT malposition among radiology solutions and describes a Widget interface with key images/summary, workstation access, PACS worklist integration, and care-team communication; current family capabilities are not assumed to prove the exact K221330 build.

product family · checked 2026-09-07

Current family workflow context only; exact-version compatibility requires confirmation.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
A retrospective, blinded, multicenter pivotal study evaluated 921 cases from 5 US clinical sites against a 2-of-3 majority reference of senior board-certified radiologists. Sensitivity was 86.81% and specificity 92.02%; BriefCase time-to-notification was 4.93 minutes in 182 true-positive cases.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Reported sensitivity
86.81% (95% CI 81.02%-91.36%)
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Reported specificity
92.02% (95% CI 89.20%-93.87%)
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07

Sensitivity for vertically malpositioned ETT triage

86.81 percent

  • EndpointDetection sensitivity against 2-of-3 majority radiologist reference
  • Population921 frontal chest X-ray cases from 5 US clinical sites
  • Tested versionBriefCase vertically malpositioned ETT triage submitted in K221330; build not stated

Specificity for vertically malpositioned ETT triage

92.02 percent

  • EndpointExclusion specificity against 2-of-3 majority radiologist reference
  • Population921 frontal chest X-ray cases from 5 US clinical sites
  • Tested versionBriefCase vertically malpositioned ETT triage submitted in K221330; build not stated

BriefCase time-to-notification

4.93 minutes

  • EndpointDICOM receipt, de-identification, cloud upload, analysis, and notification for a true-positive suspect case
  • Population182 true-positive cases in the five-center study
  • Tested versionBriefCase vertically malpositioned ETT triage submitted in K221330; build not stated

BriefCase vertically malpositioned ETT pivotal validation

  • DesignRetrospective blinded multicenter validation against a 2-of-3 majority senior board-certified radiologist reference
  • Population921 frontal chest X-ray cases from 5 US clinical sites; time-to-notification analysis included 182 true-positive cases.
  • Scopeexact submission
  • Tested versionBriefCase vertically malpositioned ETT triage submitted in K221330; build not stated
  • Samplen=921
  • Sites5
  • IndependenceIndependence not established

K221330

The FDA states that pivotal cases were distinct in time or center from training cases. That establishes dataset separation, not independent investigators or sponsorship.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-07

No public amount, range, currency, geography, or contract term was established.

Reimbursement and coding
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-09-07

No payer or CMS policy review was performed in this worker cohort.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-07Last searched
26Fields reviewed
10Source classes checked
20Screened PubMed leads
6Screened trial leads
0Unreviewed FDA recall leads

The stored PubMed, ClinicalTrials.gov, and exact-submission recall metadata candidates were screened by identifier and count. The records were not treated as a full paper, trial, or postmarket review. The current discovery snapshot contains 20 PubMed and 6 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, vendor product page, literature index, trial registry, fda device recall, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA 510(k) K221330: BriefCase malpositioned endotracheal tube triage

    U.S. Food and Drug Administration - Published 2022-11-18 - Accessed 2026-09-07

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. BriefCase-Malpositioned Endotracheal Tube Triage model card (K221330)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: purpose, modality, intendedUsers, careSetting, output, limitations, reportedSensitivity, reportedSpecificity, evidenceSummary, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. Aidoc radiology solutions

    Aidoc - Accessed 2026-09-08

    Scope: productFamily, clinicalWorkflow, PACS, implementationCapabilities

Research history

  1. K221330 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    human reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K243808

Rayvolve PTX-PE

AZmed

Passive study-level worklist triage for suspected pneumothorax and/or pleural effusion on adult PA or AP chest X-rays; it is not a proactive specialist alert.

Modality
X-ray/Radiography
Anatomy
Chest/Lung
Decision
2025-03-21
FDA-cleared K232410

SmartChest

Milvue

Passive worklist prioritization for suspected pleural effusion and/or pneumothorax on frontal PA or AP chest radiographs in transitional adolescents aged 18-21 and adults aged 22...

Modality
X-ray/Radiography
Anatomy
Chest/lung
Decision
2024-05-10