- Integration
- The FDA package describes DICOM viewing and analysis, automated segmentation, editing and pre-printing, integration with 3D Systems printers and software packages, and virtual-reality visualization for non-diagnostic use.
FDA source Exact FDA submission Checked 2026-09-04
This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.
- Deployment and data flow
- The FDA package describes a stand-alone modular software package with output to qualified 3D-printing technologies and materials; current hosting, hardware, printer, service, security, and local configuration remain unverified.
FDA source Exact FDA submission Checked 2026-09-04
This is the release-level boundary described in the exact FDA package; current hosting, contracts, security, availability, upgrade path, and site configuration remain separate evidence.
- Security and privacy
- Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04
Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.
- Training and support
- Only surgeons, radiologists, and other trained healthcare professionals should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-04
- Monitoring and change control
- Monitor K183489 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
FDA source Exact FDA submission Checked 2026-09-04
This is buyer-facing monitoring guidance derived from the exact cleared or authorized scope; it is not a claim that the FDA package provides a live monitoring service.
Release and scope verification
Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K183489 before acceptance.
exact submission · checked 2026-09-04 Later product-family features are not evidence that this cleared or authorized release includes them.
Deployment boundary
The FDA package describes a stand-alone modular software package with output to qualified 3D-printing technologies and materials; current hosting, hardware, printer, service, security, and local configuration remain unverified.
exact submission · checked 2026-09-04 FDA-described deployment boundaries do not establish current hosting contracts, security controls, uptime, or local qualification.
Interoperability acceptance
The FDA package describes DICOM viewing and analysis, automated segmentation, editing and pre-printing, integration with 3D Systems printers and software packages, and virtual-reality visualization for non-diagnostic use.
exact submission · checked 2026-09-04 Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.
Clinical acceptance
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
exact submission · checked 2026-09-04 Define stop-use, escalation, rollback, and revalidation triggers before production use.
Economic evaluation
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-04 No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.
Lifecycle monitoring
Monitor K183489 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Physical replicas are adjunctive and require other diagnostic tools and expert clinical judgment; application, printer, material, anatomy, and output-validation limits apply.
exact submission · checked 2026-09-04 Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. DICOM image-segmentation and surgical-planning software limits remain submission-specific.