Authorized purpose: Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.
Product overview
Start with the essentials
Authorized purpose: Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Overview status: Source-reviewed overview
What does this software do?Authorized purpose: Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Which images or clinical context are in scope?Modality: Mammography/DBT Anatomy or clinical context: Breast Input: A DICOM image folder from compatible full-field digital mammography or digital breast tomosynthesis systems, including RCC, LCC, RMLO, and LMLO views; the comparison describes FFDM and 2D synthetic mammography from DBT.
Who uses it?Intended users: Interpreting physicians.
What does the software produce?Output: A four-category ACR BI-RADS 5th Edition breast-density result for each examination, delivered as JSON and capable of presentation through a compatible web UI, PACS viewer using DICOM Structured Report or Secondary Capture, patient worklists, or reporting software.
What is its most important limitation?Key limitation: The output is complementary information, not diagnostic software, and patient-management decisions must not be based solely on it. The primary validation data were Hologic examinations from two U.S. screening centers and one French center. Very small subgroup results, including the eight-patient American Indian, Alaska Native, Native Hawaiian, or Pacific Islander group, have wide uncertainty.
Clinical takeaway
Clinical role: Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Evidence takeaway
Evidence from the reviewed source record: FDA standalone validation used 922 women and 4,274 examinations/images across two U.S. screening centers and one French center, with four views per examination and consensus ground truth from five breast radiologists. Quadratically weighted Cohen kappa was reported as 89.03 (95% CI 87.43-90.56) for the four-class task and 84.50 (95% CI 81.46-87.36) for the dense-versus-non-dense binary task. FFDM-only and 2DSM-only kappa values were 89.11 and 88.89; subgroup kappa values included 87.93, 91.22, and 87.15 by age and 53.85 (95% CI 0.00-83.33) for the eight-patient smallest race group.
Clinical details
What this tool is for
Start with the authorized purpose, then verify how it fits your service line and reading workflow.
Exact purpose
Adjunctively assess breast tissue composition and provide an ACR BI-RADS 5th Edition breast-density category for asymptomatic women aged 40 years or older undergoing screening mammography; it is not diagnostic.
Concurrent adjunct to physician mammography interpretation, producing breast-density classification for review and reporting through compatible mammography software, PACS, worklists, or reporting systems.
A DICOM image folder from compatible full-field digital mammography or digital breast tomosynthesis systems, including RCC, LCC, RMLO, and LMLO views; the comparison describes FFDM and 2D synthetic mammography from DBT.
A four-category ACR BI-RADS 5th Edition breast-density result for each examination, delivered as JSON and capable of presentation through a compatible web UI, PACS viewer using DICOM Structured Report or Secondary Capture, patient worklists, or reporting software.
The output is complementary information, not diagnostic software, and patient-management decisions must not be based solely on it. The primary validation data were Hologic examinations from two U.S. screening centers and one French center. Very small subgroup results, including the eight-patient American Indian, Alaska Native, Native Hawaiian, or Pacific Islander group, have wide uncertainty.
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
Use these fields to structure a vendor demo, security review, and implementation estimate.
Integration
DICOM folder input and JSON output with described integration through MammoScreen Web-UI, compatible PACS via DICOM Structured Report or Secondary Capture, patient worklists, and reporting software; the device is described as virtual-machine software. Specific local interface versions and compatibility are not established.
Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.
Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.
Training and support
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-08
Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Monitoring and change control
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-08
Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
ACR exact-submission catalog context
ACR AI Central provides an exact-submission model card for MammoScreen BD; FDA labeling remains controlling for clinical use, limitations, and performance.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA standalone validation used 922 women and 4,274 examinations/images across two U.S. screening centers and one French center, with four views per examination and consensus ground truth from five breast radiologists. Quadratically weighted Cohen kappa was reported as 89.03 (95% CI 87.43-90.56) for the four-class task and 84.50 (95% CI 81.46-87.36) for the dense-versus-non-dense binary task. FFDM-only and 2DSM-only kappa values were 89.11 and 88.89; subgroup kappa values included 87.93, 91.22, and 87.15 by age and 53.85 (95% CI 0.00-83.33) for the eight-patient smallest race group.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
MammoScreen BD regulatory performance study summary
DesignStand-Alone Performance
PopulationAdult
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K241561
Samplen=922
IndependenceIndependence not established
K241561
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
These are common procurement questions; unknown values remain visible until a source supports them.
Pricing and total cost
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-08
No public price or quote for the exact K241561 configuration was established in this review.
Reimbursement and coding
Not established in reviewed sources
Reviewed sources checked
Exact FDA submission
Checked 2026-09-08
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Safety and lifecycle
Postmarket record
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K241561. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
Verify security and privacy controls: Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.
Verify training: Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Verify pricing and total cost: No public price or quote for the exact K241561 configuration was established in this review.
Verify reimbursement and coding relevance: No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Verify local validation and lifecycle monitoring: Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
22Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Sources & history
How this profile was documented
Open the ledger for source dates, scope, and research-record updates.
View source ledger and history7 sources - 3 history items
Flags calcifications, masses, asymmetries and architectural distortions on compatible FFDM or DBT screening examinations; management must not be based solely on MammoScreen output.
Tags screening cases as suspicious or blank to aid worklist prioritization; radiologists must review every examination and it does not mark original images.
Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM...