FDA-cleared profile

Lunit INSIGHT MMG (v1.1.10)

Lunit, Inc.

Authorized purpose: Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.

Evidence takeaway

Evidence from the reviewed source record: The exact v1.1.10 standalone study is available with threshold and population context. Four stored family trial leads do not establish linkage to this release. The summary's AUC acceptance statement needs clarification.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Lunit INSIGHT MMG v1.1.10 aids physicians in detection, localization and characterization of suspicious breast-cancer areas on female screening mammograms from compatible FFDM systems, viewed after the physician's initial read as an adjunct to complete clinical interpretation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Intended users
Physicians interpreting screening mammograms
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Care setting and population
Female screening mammography; performance evaluated in adult women
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Workflow role
Detection, localization and characterization after the initial physician read
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Required input
Standard DICOM FFDM mammograms from compatible systems; validation used standard four-view 2D images from Hologic, GE Healthcare and Siemens.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Output and human action
Color, single-color, grayscale or combined heatmaps of suspicious areas and an abnormality score for each breast. Version 1.1.10 updates the AI model and removes the one-view Secondary Capture output mode.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Limitations
Physician review of the complete examination and patient history remains necessary. Results are viewed after the initial read; this authorization does not replace clinical judgment or establish autonomous screening or DBT analysis.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K260320
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Mammography/DBT
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05

Controlled catalog grouping. This submission supports FFDM; the grouping is not a DBT authorization claim.

Anatomy
Breast
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2026-04-23
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
DICOM mammograms are uploaded to the analysis engine; the summary identifies Secondary Capture analysis output and removal of its one-view mode.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05

Exact PACS interoperability and interface configuration need local confirmation.

Deployment and data flow
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-05

Current hosting model, region, hardware, latency and retention for v1.1.10 were not established by the reviewed sources.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-05

No current dated, scoped security attestation or installed data-flow/access-control package was established.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-05

Vendor family program support mentions training, but the v1.1.10 curriculum, responsibilities and support term are not established.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-05

Editorial acceptance question: verify the installed model, default operating point, supported FFDM systems and post-initial-read display sequence; supplied monitoring commitments remain unknown.

Vendor family commercial program

The current Lunit product page describes SecondRead AI as an optional out-of-pocket breast-screening analysis program with radiologist review, training and program materials.

product family ยท checked 2026-09-05

No exact v1.1.10 linkage, price or payer determination; regional availability requires confirmation.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The exact v1.1.10 standalone study is available with threshold and population context. Four stored family trial leads do not establish linkage to this release. The summary's AUC acceptance statement needs clarification.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05
Reported sensitivity
82.41% (95% CI 79.33-85.43%) at default operating point 0.1
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05

Standalone cancer detection in 2,412 adult female mammography cases from 28 US facilities; v1.1.10, not reader-aided sensitivity.

Reported specificity
81.54% (95% CI 79.78-83.31%) at default operating point 0.1
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-05

Same standalone 2,412-case v1.1.10 study and radiologist/pathology-informed reference standard.

Standalone sensitivity

82.41 percent

  • EndpointDetection of breast cancer at default operating point 0.1
  • Population2,412 adult female FFDM cases from 28 US imaging facilities; radiologist adjudication using radiology/pathology records
  • Tested versionLunit INSIGHT MMG v1.1.10

Standalone specificity

81.54 percent

  • EndpointClassification of non-cancer cases at default operating point 0.1
  • Population2,412 adult female FFDM cases from 28 US imaging facilities; radiologist adjudication using radiology/pathology records
  • Tested versionLunit INSIGHT MMG v1.1.10

K260320 Lunit INSIGHT MMG v1.1.10 standalone validation

  • DesignManufacturer standalone evaluation on a held-out multi-facility US FFDM dataset
  • PopulationAdult female screening mammography cases, Hologic/GE/Siemens acquisitions; enriched validation composition is not a routine screening prevalence estimate
  • Scopeexact submission
  • Tested versionLunit INSIGHT MMG v1.1.10
  • Samplen=2412
  • Sites28
  • IndependenceSponsor or vendor study

K260320

Two or three breast radiologists defined truth using clinical and pathology evidence. Printed AUC lower bound 0.896 conflicts with the claimed greater-than-0.903 criterion; no independent reader benefit for this release is inferred.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-05

The current vendor page mentions optional out-of-pocket SecondRead AI, but gives no verified v1.1.10 price, contract unit or included services.

Reimbursement and coding
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-09-05

No applicable CMS national/local policy content was reviewed. A vendor's optional out-of-pocket program does not establish payer coverage or noncoverage.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-05Last searched
23Fields reviewed
11Source classes checked
0Screened PubMed leads
4Screened trial leads
0Unreviewed FDA recall leads

All four stored trial titles and sponsors were screened. They are plausible Lunit family leads but metadata does not identify v1.1.10/K260320; no trial result is promoted. Full registry content and release linkage remain pending. No stored PubMed candidates. The current discovery snapshot contains 0 PubMed and 4 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, acr ai central product, literature index, trial registry, fda decision summary, vendor product page, stored candidate metadata, stored recall candidates, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA 510(k) K260320: Lunit INSIGHT MMG v1.1.10

    U.S. Food and Drug Administration - Accessed 2026-09-05

    Scope: identity, clinicalUse, modality, anatomy, implementationCapabilities, evidenceStudies, reportedSensitivity, reportedSpecificity

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Lunit INSIGHT MMG (v1.1.10) model card (K260320)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, reportedSensitivity, reportedSpecificity, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. Lunit INSIGHT MMG: AI-powered software for 2D mammograms

    Lunit - Accessed 2026-09-05

    Scope: productFamily, implementationCapabilities, currentCommercialContext

Research history

  1. K260320 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    human reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K253825

Saige-Dx

DeepHealth, Inc.

Concurrent detection and characterization support for screening DBT mammography

Modality
Mammography/DBT
Anatomy
Breast
Decision
2026-06-15
FDA-cleared K253796

Lunit INSIGHT DBT (V1.2)

Lunit, Inc.

Concurrent CADe/x aid for interpreting physicians to detect and characterize breast-cancer-suspicious lesions in screening DBT exams from compatible systems, marking soft-tissue...

Modality
Mammography/DBT
Anatomy
Breast
Decision
2026-03-26
FDA-cleared K243341

Genius AI Detection 2.0

Hologic, Inc.

Genius AI Detection 2.0 analyzes compatible digital breast tomosynthesis exams to identify and mark soft-tissue densities (masses, architectural distortions, asymmetries) and...

Modality
Mammography/DBT
Anatomy
Breast
Decision
2025-07-31
FDA-authorized DEN240047

Allix5

Clairity, Inc.

Allix5 generates a five-year breast-cancer-risk prediction from bilateral screening mammograms for eligible patients without known breast cancer at presentation; it is not...

Modality
Mammography/DBT
Anatomy
Breast
Decision
2025-05-30