FDA-cleared profile

Videa Perio Assist

VideaHealth, Inc.

Authorized purpose: Aids dental professionals in measuring and visualizing mesial and distal bone levels for each tooth on bitewing and periapical radiographs.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Aids dental professionals in measuring and visualizing mesial and distal bone levels for each tooth on bitewing and periapical radiographs.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Aids dental professionals in measuring and visualizing mesial and distal bone levels for each tooth on bitewing and periapical radiographs.

Evidence takeaway

Evidence from the reviewed source record: Bench testing included 996 radiographs and 16,131 landmarks; clinical testing included 189 radiographs and 2,350 lines, with reported bitewing sensitivity/specificity of 92.8%/89.4% and periapical sensitivity/specificity of 88.3%/87.0%.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Aids dental professionals in measuring and visualizing mesial and distal bone levels for each tooth on bitewing and periapical radiographs.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.

Intended users
Trained professionals, including dentists and dental hygienists.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet user statement.

Care setting and population
Adult and Pediatric
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Measurement or quantification
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
Bitewing and periapical radiographs from patients aged 12 years or older with permanent dentition.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Exact input and population boundary.

Output and human action
Points and connecting lines for bone-level measurements in millimeters and/or percentage of root length, or a no-identifiable-teeth result.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA output description.

Limitations
Not a substitute for full patient evaluation or the sole basis for diagnosis; tested sensor, age, and periapical subgroups have specified performance boundaries.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Exact FDA limitation boundary.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K223296
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Dental radiographs normalized to the controlled X-ray bucket.

Anatomy
Teeth and mesial/distal alveolar bone levels
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA permanent-dentition scope.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2023-02-06
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact-submission ACR model card explicitly references DICOM in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Deployment and data flow
On-cloud
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for Videa Perio Assist; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-09

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Bench testing included 996 radiographs and 16,131 landmarks; clinical testing included 189 radiographs and 2,350 lines, with reported bitewing sensitivity/specificity of 92.8%/89.4% and periapical sensitivity/specificity of 88.3%/87.0%.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet performance summary.

Reported sensitivity
82%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
81%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Videa Perio Assist regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult and Pediatric
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K223296
  • Samplen=189
  • IndependenceIndependence not established

K223296

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

No public price or quote for the exact K223296 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-09

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K223296. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-09

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
26Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K223296: Videa Perio Assist

    U.S. Food and Drug Administration - Published 2023-02-06 - Accessed 2026-09-09

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Videa Perio Assist model card (K223296)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, reportedSensitivity, reportedSpecificity, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K223296

    U.S. Food and Drug Administration - Accessed 2026-09-09

    Scope: postmarketSafety

Research history

  1. K223296 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K253192

DeepXray Spina

Alpha Intelligence Manifolds, Inc.

Opportunistic review of frontal lumbar-spine or KUB radiographs in adults aged 50 years and older to flag possible low bone mineral density at L1-L4 for clinician follow-up; a...

Modality
X-ray/Radiography
Anatomy
Spine/Bone Health
Decision
2026-06-10
FDA-cleared K260378

Rayvolve

AZmed

Rayvolve supports clinicians by identifying suspected fractures, joint effusions, and dislocations on labeled radiographic views; it does not replace clinician interpretation.

Modality
X-ray/Radiography
Anatomy
Musculoskeletal and chest/ribs radiographs across specified views
Decision
2026-05-12
FDA-cleared K252953

Velmeni for Dentists (V4D) Endo-Perio

Velmeni, Inc.

Concurrent-read CADe to assist dentists in detecting specified tooth and jaw anatomy, bone-level landmarks, and root-canal-treated teeth in digital radiographs of permanent teeth...

Modality
X-ray/Radiography
Anatomy
Permanent teeth and jaw/maxillofacial anatomy, including enamel, dentine, pulp, root, mental foramen, inferior alveolar canal, maxillary sinus, cementoenamel junction, crestal bone, and root-canal-treated teeth
Decision
2026-05-11
FDA-cleared K251934

qXR-Detect

Qure.Ai Technologies

qXR-Detect is an adult chest-radiograph concurrent reading aid that identifies, highlights and categorizes suspicious regions of interest across lung, pleura,...

Modality
X-ray/Radiography
Anatomy
Chest/Lung; pleura, mediastinum/hila/heart, bone, hardware and other chest-radiograph regions
Decision
2026-01-16