FDA-cleared profile

DeepXray Spina

Alpha Intelligence Manifolds, Inc.

Authorized purpose: Opportunistic review of frontal lumbar-spine or KUB radiographs in adults aged 50 years and older to flag possible low bone mineral density at L1-L4 for clinician follow-up; a negative result does not rule out low BMD.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Opportunistic review of frontal lumbar-spine or KUB radiographs in adults aged 50 years and older to flag possible low bone mineral density at L1-L4 for clinician follow-up; a negative result does not rule out low BMD.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Opportunistic review of frontal lumbar-spine or KUB radiographs in adults aged 50 years and older to flag possible low bone mineral density at L1-L4 for clinician follow-up; a negative result does not rule out low BMD.

Evidence takeaway

Evidence from the reviewed source record: Study type: Stand-Alone Performance. Cases: 1219. Reported sensitivity: 77.0% (95% CI: 72.0%, 81.9%). Reported specificity: 93.6% (95% CI: 92.1%, 95.1%)

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Opportunistic review of frontal lumbar-spine or KUB radiographs in adults aged 50 years and older to flag possible low bone mineral density at L1-L4 for clinician follow-up; a negative result does not rule out low BMD.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Radiologists and/or physician interpreters
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Care setting and population
Adult
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

Confirm whether the tool is concurrent reading, triage, post-processing, planning, acquisition support, or another workflow role.

Required input
DeepXray Spina is intended for use with standard frontal radiographs of the lumbar spine or KUB that are suitable for L1–L4 BMD analysis.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact input restriction states the supported frontal lumbar-spine or KUB radiographs and suitability for L1–L4 BMD analysis; the current generic consultation wording is not packet text.

Output and human action
The device outputs a binary classification: 'Low BMD' (internal L1-L4 T-score 2.5) or 'Not Suspected' (internal L1-L4 T-score > 2.5). A DeepXray Spina report is generated for positive cases only, displaying a vertebral region-of-interest (ROI) overlay on the original radiograph. For negative results, no report is generated. An automated Quality-Control (QC) function detects technical and anatomical exclusion conditions and flags input images with potential quality issues.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Limitations
Exclusion criteria: Images in which one or more L1-L4 vertebrae contains: (a) metallic spinal implants or prostheses (e.g., pedicle screws, rods, cages, or total disc replacements); (b) vertebral cement augmentation (e.g., kyphoplasty or vertebroplasty); (c) severe vertebral deformity, fracture, or other structural abnormality that would preclude a valid DXA-based L1-L4 BMD measurement; or (d) extraspinal medical implants overlying the L1-L4 measurement region (e.g., abdominal aortic stent grafts or vascular grafts).
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K253192
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Anatomy
Spine/Bone Health
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2026-06-10
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

Attach protocol-level DICOM, HL7, FHIR, PACS, RIS, EHR, or orchestrator documentation.

Deployment and data flow
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for DeepXray Spina; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Study type: Stand-Alone Performance. Cases: 1219. Reported sensitivity: 77.0% (95% CI: 72.0%, 81.9%). Reported specificity: 93.6% (95% CI: 92.1%, 95.1%)
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Reported sensitivity
77.0% (95% CI: 72.0%, 81.9%)
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
93.6% (95% CI: 92.1%, 95.1%)
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

DeepXray Spina regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K253192
  • Samplen=1219
  • IndependenceIndependence not established

K253192

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

No public price or quote for the exact K253192 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-08

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K253192. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
23Fields reviewed
10Source classes checked
2Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 2 PubMed and 0 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 2 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, acr ai central product, fda decision summary, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K253192: DeepXray Spina

    U.S. Food and Drug Administration - Published 2026-06-10 - Accessed 2026-09-19

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, output, limitations, evidenceSummary, input

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. DeepXray Spina model card (K253192)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, reportedSensitivity, reportedSpecificity, evidenceSummary, anatomy, evidenceStudies

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K253192

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

Research history

  1. K253192 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K260378

Rayvolve

AZmed

Rayvolve supports clinicians by identifying suspected fractures, joint effusions, and dislocations on labeled radiographic views; it does not replace clinician interpretation.

Modality
X-ray/Radiography
Anatomy
Musculoskeletal and chest/ribs radiographs across specified views
Decision
2026-05-12
FDA-cleared K252953

Velmeni for Dentists (V4D) Endo-Perio

Velmeni, Inc.

Concurrent-read CADe to assist dentists in detecting specified tooth and jaw anatomy, bone-level landmarks, and root-canal-treated teeth in digital radiographs of permanent teeth...

Modality
X-ray/Radiography
Anatomy
Permanent teeth and jaw/maxillofacial anatomy, including enamel, dentine, pulp, root, mental foramen, inferior alveolar canal, maxillary sinus, cementoenamel junction, crestal bone, and root-canal-treated teeth
Decision
2026-05-11
FDA-cleared K251934

qXR-Detect

Qure.Ai Technologies

qXR-Detect is an adult chest-radiograph concurrent reading aid that identifies, highlights and categorizes suspicious regions of interest across lung, pleura,...

Modality
X-ray/Radiography
Anatomy
Chest/Lung; pleura, mediastinum/hila/heart, bone, hardware and other chest-radiograph regions
Decision
2026-01-16
FDA-cleared K250914

MediAI-BA

Crescom Co., Ltd.

View and quantify bone age from 2D posteroanterior left-hand radiographs to aid pediatric radiologists assessing patients aged 2-18 years.

Modality
X-ray/Radiography
Anatomy
Left hand and wrist; skeletal maturity/bone age
Decision
2025-12-18