FDA-cleared profile

Saige-Q

DeepHealth, Inc.

Authorized purpose: Provides passive worklist prioritization for compatible FFDM and DBT screening mammograms.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Provides passive worklist prioritization for compatible FFDM and DBT screening mammograms.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Provides passive worklist prioritization for compatible FFDM and DBT screening mammograms.

Evidence takeaway

Evidence from the reviewed source record: Retrospective blinded multicenter FFDM study of 501 malignant and 832 normal cases and DBT study of 517 malignant and 1,011 normal cases reported AUC 0.966 and 0.985 respectively; no one global accuracy applies.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Provides passive worklist prioritization for compatible FFDM and DBT screening mammograms.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.

Intended users
Interpreting radiologists.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet intended-user statement.

Care setting and population
The screening population described in the FDA intended use; the institutional care setting requires local confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Workflow role
Triage or notification support
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
DICOM files from one FFDM or DBT screening study, compatible with specified Hologic mammography equipment.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet input and system scope.

Output and human action
A Saige-Q suspicion code to PACS, EPR, RIS, or workstation; suspicious studies also receive an informational compressed preview.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet output scope.

Limitations
Passive notification only, with no diagnostic information beyond triage; not a replacement for image review or standalone decision-making and cases remain on the worklist.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Exact FDA limitation boundary.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K203517
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Mammography/DBT
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA FFDM and DBT scope normalized to catalog modality.

Anatomy
Breast
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Exact FDA breast scope.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2021-04-16
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact-submission ACR model card explicitly references DICOM, PACS, and RIS in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Deployment and data flow
Both
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Result destination

PACS and workstation are explicitly referenced as a result destination.

exact submission · checked 2026-09-01

Confirm supported products, versions, routing behavior, and display requirements during local evaluation.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for Saige-Q; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-09

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Retrospective blinded multicenter FFDM study of 501 malignant and 832 normal cases and DBT study of 517 malignant and 1,011 normal cases reported AUC 0.966 and 0.985 respectively; no one global accuracy applies.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Submission-reported performance results.

Reported sensitivity
89.40%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
86.90%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Saige-Q regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationPopulation described in the exact-submission ACR model card
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K203517
  • Samplen=2861
  • Sites14
  • IndependenceIndependence not established

K203517

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-09

No public price or quote for the exact K203517 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-09

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K203517. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-09

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
28Fields reviewed
10Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K203517: Saige-Q

    U.S. Food and Drug Administration - Published 2021-04-16 - Accessed 2026-09-19

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Saige-Q model card (K203517)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, output, limitations, reportedSensitivity, reportedSpecificity, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K203517

    U.S. Food and Drug Administration - Accessed 2026-09-09

    Scope: postmarketSafety

Research history

  1. K203517 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K252954

MammoSightAI

Neurocareai, Inc. (Dba Savelife.Ai)

Tags screening cases as suspicious or blank to aid worklist prioritization; radiologists must review every examination and it does not mark original images.

Modality
Mammography/DBT
Anatomy
Breast
Decision
2026-05-04
FDA-cleared K252482

CogNet AI-MT+

Medcognetics, Inc.

Prioritizes suspicious screening DBT examinations for MQSA interpreting physicians; every examination remains subject to full physician review.

Modality
Mammography/DBT
Anatomy
Breast
Decision
2025-12-11
FDA-cleared K233108

VinDr-Mammo

VinBigData Joint Stock Company

The VinDr-Mammo is a passive notification for prioritization-only, a parallel- workflow software tool used by MQSA qualified interpreting physicians to prioritize patients with...

Modality
Mammography/DBT
Anatomy
Breast
Decision
2024-05-23