FDA-cleared profile

Rapid Neuro3D

iSchemaView, Inc.

Authorized purpose: Rapid Neuro3D analyzes adult head-and-neck CTA DICOM from the aortic arch to the vertex, removing bone, tissue, and venous vessels to generate 3D/2D neurovascular visualizations, including MIP, volume-rendered, and CPR views of carotid and vertebral arteries for trained physicians' lesion review and planning; results are not standalone clinical decisions.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Rapid Neuro3D analyzes adult head-and-neck CTA DICOM from the aortic arch to the vertex, removing bone, tissue, and venous vessels to generate 3D/2D neurovascular visualizations, including MIP, volume-rendered, and CPR views of carotid and vertebral arteries for trained physicians' lesion review and planning; results are not standalone clinical decisions.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Rapid Neuro3D analyzes adult head-and-neck CTA DICOM from the aortic arch to the vertex, removing bone, tissue, and venous vessels to generate 3D/2D neurovascular visualizations, including MIP, volume-rendered, and CPR views of carotid and vertebral arteries for trained physicians' lesion review and planning; results are not standalone clinical decisions.

Evidence takeaway

Evidence from the reviewed source record: K243350 FDA evidence is manufacturer-submitted standalone regulatory validation. The quality endpoint used 120 CTA cases from 115 patients, while the accuracy endpoint used 50 CTA cases from 48 patients; their totals are kept separate. No independent paper or trial was promoted in this timeboxed review.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Rapid Neuro3D analyzes adult head-and-neck CTA DICOM from the aortic arch to the vertex, removing bone, tissue, and venous vessels to generate 3D/2D neurovascular visualizations, including MIP, volume-rendered, and CPR views of carotid and vertebral arteries for trained physicians' lesion review and planning; results are not standalone clinical decisions.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Intended users
Physicians and trained clinical professionals evaluating neurovascular CTA; the physician remains responsible for clinical assessment and decisions.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Care setting and population
Adult neurovascular CTA review, lesion documentation, follow-up, and treatment-planning workflows using a DICOM viewer, clinical workstation, or PACS.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Workflow role
Automated head-and-neck arterial visualization and segmentation to support physician confirmation of identified lesions, evaluation, documentation, follow-up, and treatment planning; output does not replace clinical assessment.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Required input
Conventional adult CTA DICOM images extending from the aortic arch to the vertex of the head.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Output and human action
A DICOM series with 3D rotational maximum-intensity projections, volume renders, source-series-equivalent images with bone/tissue/vein removal, and curved planar reformations of isolated left and right internal carotid and vertebral arteries.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Limitations
Excessive motion or metal implants may affect output. The module excludes inadequate intracranial arterial contrast, cone-beam C-arm CT, non-axial or non-supine series, missing or improperly ordered slices, and CTA datasets outside its stated field-of-view and acquisition constraints. Results are not intended for standalone clinical decision-making.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K243350
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
CT/CTA
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Anatomy
Brain
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2025-01-22
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The Rapid Platform accepts DICOM images and returns processed DICOM images to the source imaging modality or PACS; the resulting DICOM series can be used by DICOM viewers and clinical workstations. Local routing, identifiers, viewer behavior, and failure handling require acceptance testing.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Deployment and data flow
The exact Rapid Platform may be installed on-premises behind the customer firewall or in a hybrid on-premises/cloud configuration, on dedicated hardware or a virtual machine; the reviewed sources do not establish the current site-specific topology, storage, retention, availability, or service-level arrangement.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Security and privacy
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-07

The reviewed exact packet and official family page do not establish current scoped security controls, encryption, PHI flow, retention, access control, audit, vulnerability management, or incident response for the installed K243350 release.

Training and support
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-07

The FDA indication requires trained clinical review, but no exact-version curriculum, competency assessment, or support terms were established.

Monitoring and change control
Monitor CTA coverage and contrast eligibility, slice order and patient-position constraints, motion or metal-related failures, output review, DICOM delivery, failed or repeated analyses, reader overrides, downtime, updates, complaints, and exact-submission recalls.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07

Buyer acceptance guidance derived from the exact FDA workflow and validation scope; it is not a claimed vendor monitoring service.

Family CTA reconstruction outputs

RapidAI's current radiologist page lists automated CTA reconstruction for the head, neck, and aortic arch, with curved planar reformats, MIPs, volume renderings, bone removal, and segmentation. This is Rapid family context and does not establish the exact K243350 release or contract configuration.

product family · checked 2026-09-07

Rapid Platform DICOM exchange

The exact FDA submission describes DICOM input and return of processed DICOM images to the source modality or PACS; confirm local routing, viewer compatibility, firewall rules, and failure recovery during qualification.

exact submission · checked 2026-09-07

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
K243350 FDA evidence is manufacturer-submitted standalone regulatory validation. The quality endpoint used 120 CTA cases from 115 patients, while the accuracy endpoint used 50 CTA cases from 48 patients; their totals are kept separate. No independent paper or trial was promoted in this timeboxed review.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-07
Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

MIP clinical agreement

0.998 proportion

  • EndpointExpert-consensus agreement with source DICOM for 3D rotational MIP output
  • Population120 CTA cases from 115 patients; 1,920 MIP, VR, and SSE segments were evaluated across the quality study
  • Tested versionRapid Neuro3D configuration described in K243350

Volume-rendered clinical agreement

0.986 proportion

  • EndpointExpert-consensus agreement with source DICOM for volume-rendered output
  • Population120 CTA cases from 115 patients; quality study
  • Tested versionRapid Neuro3D configuration described in K243350

Source-series-equivalent clinical agreement

1 proportion

  • EndpointExpert-consensus agreement with source DICOM for source-series-equivalent output
  • Population120 CTA cases from 115 patients; quality study
  • Tested versionRapid Neuro3D configuration described in K243350

CPR clinical agreement

1 proportion

  • EndpointExpert-consensus agreement with source DICOM for curved planar reformation output
  • Population120 CTA cases from 115 patients; 480 CPR segments were evaluated in the quality study
  • Tested versionRapid Neuro3D configuration described in K243350

Extracranial segmentation Dice

0.89 Dice coefficient

  • EndpointAverage extracranial segmentation Dice versus ground truth
  • Population50 CTA cases from 48 patients in the segmentation-accuracy study
  • Tested versionRapid Neuro3D configuration described in K243350

Intracranial segmentation Dice

0.97 Dice coefficient

  • EndpointAverage intracranial segmentation Dice versus predicate-device reference
  • Population50 CTA cases from 48 patients in the segmentation-accuracy study
  • Tested versionRapid Neuro3D configuration described in K243350

Extracranial Hausdorff distance

0.44 mm

  • EndpointAverage Hausdorff distance between extracranial module output and ground truth
  • Population50 CTA cases from 48 patients in the segmentation-accuracy study
  • Tested versionRapid Neuro3D configuration described in K243350

Intracranial Hausdorff distance

0.44 mm

  • EndpointAverage Hausdorff distance between intracranial module output and predicate-device reference
  • Population50 CTA cases from 48 patients in the segmentation-accuracy study
  • Tested versionRapid Neuro3D configuration described in K243350

CPR centerline Hausdorff distance

0.31 mm

  • EndpointAverage centerline Hausdorff distance for CPR visualization
  • Population50 CTA cases from 48 patients in the segmentation-accuracy study
  • Tested versionRapid Neuro3D configuration described in K243350

K243350 Rapid Neuro3D segmentation-quality validation

  • DesignStandalone performance validation comparing MIP, volume-rendered, source-series-equivalent, and CPR outputs with source DICOM and expert consensus.
  • Population120 CTA cases from 115 patients (104 US, 16 OUS; 65 female and 50 male; ages 27 to 90+); major CT scanner manufacturers were represented and the dataset was independent of model-training data per internal quality-control processes.
  • Scopeexact submission
  • Tested versionRapid Neuro3D configuration described in K243350
  • Samplen=120
  • IndependenceSponsor or vendor study

K243350

Manufacturer-submitted FDA evidence; this is not an independent clinical-outcome study.

K243350 Rapid Neuro3D segmentation-accuracy validation

  • DesignStandalone performance validation comparing 3D visualization segmentation with ground truth or predicate reference and CPR centerlines with ground truth.
  • Population50 CTA cases from 48 patients (43 US, 7 OUS; 24 female and 24 male; ages 27 to 90+), assessed for extracranial and intracranial regions and CPR centerline accuracy.
  • Scopeexact submission
  • Tested versionRapid Neuro3D configuration described in K243350
  • Samplen=50
  • IndependenceSponsor or vendor study

K243350

Manufacturer-submitted FDA evidence; the intracranial comparison used the predicate-device reference for the stated endpoint.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-07

No public list price or exact contract unit for K243350 was established in the reviewed official sources.

Reimbursement and coding
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-09-07

No payer-policy research was applied to this dossier; product-specific coverage or claim eligibility is not established.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-submission FDA recall query returned no native recall record for K243350. This is not a comprehensive safety clearance and does not exclude family, later-version, or adverse-event information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-07

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-07Last searched
24Fields reviewed
11Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The stored PubMed and ClinicalTrials.gov candidate arrays were screened and were empty. No paper or trial was promoted; empty stored arrays do not establish that no external literature or trials exist. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, vendor product page, fda device recall, literature index, trial registry, stored candidate metadata, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA 510(k) K243350: Rapid Neuro3D

    U.S. Food and Drug Administration - Accessed 2026-09-07

    Scope: identity, clinicalUse, modality, anatomy, intendedUsers, careSetting, workflowRole, input, output, limitations, integration, deployment, evidenceMetrics, evidenceStudies

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Rapid Neuro3D model card (K243350)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. RapidAI for Radiologists: CTA reconstruction and neurovascular workflow

    RapidAI / iSchemaView - Accessed 2026-09-07

    Scope: productFamily, clinicalUse, implementationCapabilities

  8. openFDA exact recall query for K243350

    U.S. Food and Drug Administration - Accessed 2026-09-07

    Scope: postmarketSafety

Research history

  1. K243350 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    human reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K200921

qER

Qure.ai Technologies

qER is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT images. The device is intended to assist hospital...

Modality
CT/CTA
Anatomy
Brain
Decision
2020-06-17
FDA-cleared K253818

Annalise Enterprise

Harrison-AI Medical Pty, Ltd.

Annalise Enterprise prioritizes and notifies trained CT clinicians of suspected acute infarct on non-contrast brain CT for adults; it is a study-level, non-localizing worklist aid...

Modality
CT/CTA
Anatomy
Brain
Decision
2026-03-03
FDA-cleared K251195

BriefCase-Triage

Aidoc Medical, Ltd.

Adult contrast-enhanced brain CT triage for suspected brain aneurysm findings at least 3.0 mm.

Modality
CT/CTA
Anatomy
Brain
Decision
2026-01-27