NeuViz 128 Multi-Slice CT Scanner System

Neusoft Medical Systems Co., Ltd.

The NeuViz 128 Multi-Slice CT Scanner System is a whole-body computed-tomography X-ray system with a continuously rotating X-ray tube and detector array that reconstructs X-ray transmission data into cross-sectional images of the body.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
The NeuViz 128 Multi-Slice CT Scanner System is a whole-body computed-tomography X-ray system with a continuously rotating X-ray tube and detector array that reconstructs X-ray transmission data into cross-sectional images of the body.
FDA source Exact FDA submission Checked 2026-09-04
Intended users
Qualified radiologic technologists and physicians or other trained healthcare professionals within the labeled workflow.
FDA source Exact FDA submission Checked 2026-09-04
Care setting and population
Hospital, clinic, emergency, outpatient, or other labeled CT imaging environment.
FDA source Exact FDA submission Checked 2026-09-04
Workflow role
X-ray CT acquisition and reconstruction, with release-specific image-quality, spectral, or deep-learning processing functions where labeled.
FDA source Exact FDA submission Checked 2026-09-04
Required input
X-ray transmission data acquired with the exact labeled CT system, protocols, software, and hardware.
FDA source Exact FDA submission Checked 2026-09-04
Output and human action
Cross-sectional CT images, volume sets, and labeled reconstructed or processed image outputs for professional review.
FDA source Exact FDA submission Checked 2026-09-04
Limitations
Use only the exact labeled scanner, software, dose, reconstruction, patient population, anatomy, and screening criteria; local dose and image-quality decisions remain clinical responsibilities.
FDA source Exact FDA submission Checked 2026-09-04

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K230220
FDA source Exact FDA submission Checked 2026-08-31
Modality
CT/CTA
FDA source Exact FDA submission Checked 2026-09-04
Anatomy
Whole body including head
FDA source Exact FDA submission Checked 2026-09-04
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2023-10-11
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Standards-based DICOM/PACS integration was identified in the FDA submission materials; compatibility with a specific PACS, RIS, EHR, or orchestrator requires confirmation.
Public source Exact FDA submission Checked 2026-08-13

Normalized from the prior exact-submission review. Confirm interface versions and local compatibility during technical evaluation.

Deployment and data flow
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

Exact deployment topology, hosting, network boundaries, PHI flow, retention, update mechanism, and local operational requirements are not established for this submission.

Security and privacy
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

Public sources reviewed here do not establish the complete exact-release security posture, including identity, audit, encryption, retention, remote access, vulnerability management, backup, or incident response.

Training and support
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Training, onboarding, competency, and support commitments are commercial implementation details that require current vendor documentation.

Monitoring and change control
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Local acceptance testing, production monitoring, drift review, update control, and incident escalation require a current vendor and site-specific governance plan.

Workflow role

X-ray CT acquisition and reconstruction, with release-specific image-quality, spectral, or deep-learning processing functions where labeled.

exact submission · checked 2026-09-04

Submission-anchored workflow summary, not a local acceptance or deployment result.

Input

X-ray transmission data acquired with the exact labeled CT system, protocols, software, and hardware.

exact submission · checked 2026-09-04

Exact supported inputs and protocol limits must be confirmed from the complete labeling.

Output

Cross-sectional CT images, volume sets, and labeled reconstructed or processed image outputs for professional review.

exact submission · checked 2026-09-04

Output review, editing, approval, export, and downstream use remain release- and site-specific.

Deployment evidence

Exact deployment topology and operational boundaries remain unestablished.

exact submission · checked 2026-09-04

Confirm hosting, PHI flow, retention, access, updates, downtime, and support obligations before procurement or clinical use.

Integration evidence

Exact local interfaces, identity matching, routing, display, reporting, and fallback behavior remain unestablished.

exact submission · checked 2026-09-04

Validate the local workflow against the exact release and enabled options.

Vendor family context

Current vendor material is product-family context for K230220 only; exact release, enabled options, deployment, security, interoperability, pricing, and local workflow require confirmation.

product family · checked 2026-09-04

Vendor material is not exact-release evidence unless that linkage is separately proven.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
The exact FDA material describes engineering bench testing and clinical image evaluation for the ClearInfinity reconstruction function, including reader comparisons against filtered back projection; these are submission-specific image-quality evaluations, not universal clinical outcomes.
FDA source Exact FDA submission Checked 2026-09-04
Reported sensitivity
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No exact-submission and version-linked sensitivity endpoint is promoted; candidate or product-family values were not substituted.

Reported specificity
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No exact-submission and version-linked specificity endpoint is promoted; candidate or product-family values were not substituted.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

FDA submission-specific verification and validation evidence for K230220

FDA 510(k) regulatory decision material and submission-specific verification or validation evidence for K230220; this is regulatory evidence, not an independent clinical effectiveness study. · The labeled population, anatomy, images, and test materials described in the exact FDA decision material for K230220; a broader clinical population is not established here.

exact submission · Tested version: NeuViz 128 Multi-Slice CT Scanner System (K230220) · Sponsor or vendor study

K230220

The exact FDA material describes engineering bench testing and clinical image evaluation for the ClearInfinity reconstruction function, including reader comparisons against filtered back projection; these are submission-specific image-quality evaluations, not universal clinical outcomes.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No public exact-release price, license fee, subscription, contract, or implementation cost was established in this review.

Reimbursement and coding
Not applicable
Not applicable Not applicable Checked 2026-09-04

CMS coverage-index context was checked, but no product-specific coverage or reimbursement determination is inferred.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-04Last searched
30Fields reviewed
11Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact-name and exact-submission searches in the checked 2026-09-01 literature and trial indexes returned no PubMed candidate records and no ClinicalTrials.gov candidate records for K230220; the checked openFDA snapshot returned no recall candidate records. Candidate records remain quarantined unless exact product and version linkage is established. Current vendor material is product-family context only. Pricing, complete security architecture, contracted deployment, exact reimbursement, and local interoperability remain unverified. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, vendor product page, implementation documentation, reimbursement policy, peer reviewed literature, clinical trial registry, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-04.

Alternatives

Related tools to compare

FDA-cleared K260729

Vascular Assist Occlusion Triage (VAOT) (v1.0)

GuideAI Health

Passive workflow prioritization for studies with imaging patterns consistent with possible total arterial occlusion in the specified vessels; not primary diagnosis, screening,...

Modality
CT/CTA
Anatomy
Aorta and lower-extremity arteries: common iliac, external iliac, common femoral, superficial femoral, and popliteal arteries
Decision
2026-06-18
FDA-cleared K254075

Synapse Lung Nodule AI

Fujifilm Corporation

Concurrent-reader aid for a radiologist detecting pulmonary nodules during review of chest CT examinations in the lung cancer screening population under screening protocols.

Modality
CT/CTA
Anatomy
Chest/lungs; pulmonary nodules
Decision
2026-05-27
FDA-cleared K253520

Photonova Spectra, Photonova Spectra Select

Ge Medical Systems, LLC

Photonova Spectra and Spectra Select are silicon-based spectral photon-counting detector CT systems that acquire cross-sectional body images, multienergy data, monochromatic...

Modality
CT/CTA
Anatomy
Head, whole body, heart, vessels, and lungs
Decision
2026-03-20
FDA-cleared K260166

Bunkerhill Contrast CAC

BunkerHill Health

Detect, localize, and estimate the burden of coronary artery calcification on contrast-enhanced chest CT during clinical care.

Modality
CT/CTA
Anatomy
Chest/coronary arteries
Decision
2026-03-06