FDA-cleared profile

Lung-CAD

Imagen Technologies, Inc

Authorized purpose: Lung-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for lung hyperinflation. The device uses a deep learning algorithm to identify regions of interest (ROIs) with lung hyperinflation and produces boxes around the ROIs. Lung-CAD is intended for use as a concurrent reading aid for physicians interpreting chest X-rays. The device is not intended for clinical diagnosis of any disease. It does not replace the role of the physician or of other diagnostic testing in the standard of care for lung parenchymal findings. Lung-CAD is indicated for adults only.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Lung-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for lung hyperinflation. The device uses a deep learning algorithm to identify regions of interest (ROIs) with lung hyperinflation and produces boxes around the ROIs. Lung-CAD is intended for use as a concurrent reading aid for physicians interpreting chest X-rays. The device is not intended for clinical diagnosis of any disease. It does not replace the role of the physician or of other diagnostic testing in the standard of care for lung parenchymal findings. Lung-CAD is indicated for adults only.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Lung-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for lung hyperinflation. The device uses a deep learning algorithm to identify regions of interest (ROIs) with lung hyperinflation and produces boxes around the ROIs. Lung-CAD is intended for use as a concurrent reading aid for physicians interpreting chest X-rays. The device is not intended for clinical diagnosis of any disease. It does not replace the role of the physician or of other diagnostic testing in the standard of care for lung parenchymal findings. Lung-CAD is indicated for adults only.

Evidence takeaway

Evidence from the reviewed source record: Imagen conducted a standalone performance assessment on 5,000 chest radiograph cases representative of the intended use population. The results demonstrated sensitivity of 0.898, specificity of 0.894, and Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve of 0.964. A fully-crossed multiple reader, multiple case retrospective reader study evaluated 244 cases.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Lung-CAD is a computer-assisted detection (CADe) software device that analyzes chest radiograph studies for lung hyperinflation. The device uses a deep learning algorithm to identify regions of interest (ROIs) with lung hyperinflation and produces boxes around the ROIs. Lung-CAD is intended for use as a concurrent reading aid for physicians interpreting chest X-rays. The device is not intended for clinical diagnosis of any disease. It does not replace the role of the physician or of other diagnostic testing in the standard of care for lung parenchymal findings. Lung-CAD is indicated for adults only.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Physicians interpreting chest X-rays
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Care setting and population
Adult
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Detection, localization, or diagnostic support
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
Digital X-ray chest radiographs.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact subject-device table identifies the image source as Digital X-ray and the study type as Chest; the current generic data and consult-instruction wording is not retained.

Output and human action
For each image within a study, Lung-CAD generates a DICOM Presentation State file (output overlay). If any region of interest (ROI) is detected, the output overlay includes 'Lung hyperinflation' label and bounding boxes surrounding each detected ROI. If no ROI is detected, the output overlay includes the text 'No Lung-CAD ROI(s)' and no bounding boxes are included.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Limitations
The device is not intended for clinical diagnosis of any disease. It does not replace the role of the physician or of other diagnostic testing in the standard of care for lung parenchymal findings. Lung-CAD is indicated for adults only.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact indications state the diagnostic, physician, testing, and adult-population boundaries; they do not state the current scanner-manufacturer or exclusion-criteria wording.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K230085
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
X-ray/Radiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Anatomy
Chest/Lung
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2023-10-03
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact-submission ACR model card explicitly references DICOM in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Deployment and data flow
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for Lung-CAD; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Imagen conducted a standalone performance assessment on 5,000 chest radiograph cases representative of the intended use population. The results demonstrated sensitivity of 0.898, specificity of 0.894, and Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve of 0.964. A fully-crossed multiple reader, multiple case retrospective reader study evaluated 244 cases.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact packet reports sensitivity, specificity, and ROC AUC, not the current rounded accuracy label, and separately reports the 244-case reader study.

Reported sensitivity
89%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
89%
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Lung-CAD regulatory performance study summary

  • DesignStand-alone performance and reader study (ACR AI Central category)
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K230085
  • Samplen=5000
  • IndependenceIndependence not established

K230085

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

No public price or quote for the exact K230085 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-08

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K230085. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
26Fields reviewed
10Source classes checked
14Screened PubMed leads
1Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 14 PubMed and 1 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 14 PubMed and 1 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K230085: Lung-CAD

    U.S. Food and Drug Administration - Published 2023-10-03 - Accessed 2026-09-19

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, output, input, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Lung-CAD model card (K230085)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, reportedSensitivity, reportedSpecificity, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K230085

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

Research history

  1. K230085 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K223811

Lung-CAD

Imagen Technologies, Inc

Lung-CAD is a concurrent-read computer-aided detection aid for identifying and marking regions of interstitial thickening on adult chest radiographs; it does not provide a...

Modality
X-ray/Radiography
Anatomy
Chest/Lung
Decision
2023-09-13
FDA-cleared K243831

Rayvolve LN

AZmed

Second-reader support for radiologists reviewing adult digital AP or PA chest radiographs, with suspected pulmonary nodules in the 6-30 mm range marked for review.

Modality
X-ray/Radiography
Anatomy
Chest/Lung
Decision
2025-03-26
FDA-cleared K243808

Rayvolve PTX-PE

AZmed

Passive study-level worklist triage for suspected pneumothorax and/or pleural effusion on adult PA or AP chest X-rays; it is not a proactive specialist alert.

Modality
X-ray/Radiography
Anatomy
Chest/Lung
Decision
2025-03-21