FDA-cleared profile

Eclipse II with Smart Noise Cancellation

Carestream Health, Inc.

Eclipse II with Smart Noise Cancellation digitally enhances radiographic images generated by supported X-ray devices for adult and pediatric use; the cleared function is for projection radiography and excludes mammography.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Eclipse II with Smart Noise Cancellation digitally enhances radiographic images generated by supported X-ray devices for adult and pediatric use; the cleared function is for projection radiography and excludes mammography.
FDA source Exact FDA submission Checked 2026-09-04
Intended users
Qualified radiologic technologists and interpreting physicians.
FDA source Exact FDA submission Checked 2026-09-04
Care setting and population
Hospital and outpatient projection-radiography workflows.
FDA source Exact FDA submission Checked 2026-09-04
Workflow role
Digital radiographic image enhancement and professional review support.
FDA source Exact FDA submission Checked 2026-09-04
Required input
Supported adult or pediatric projection-radiography acquisitions from compatible CR or DR devices.
FDA source Exact FDA submission Checked 2026-09-04
Output and human action
Digitally processed radiographic images for professional review.
FDA source Exact FDA submission Checked 2026-09-04
Limitations
The function is for supported projection radiography and excludes mammography; detector, acquisition, processing, and professional-review limits apply.
FDA source Exact FDA submission Checked 2026-09-04

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K202441
FDA source Exact FDA submission Checked 2026-08-31
Modality
X-ray/Radiography
FDA source Exact FDA submission Checked 2026-09-04
Anatomy
Adult and pediatric radiographic anatomy excluding mammography
FDA source Exact FDA submission Checked 2026-09-04
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2021-04-02
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The FDA package describes Smart Noise Cancellation as an optional ImageView function for supported CR and DR projection-radiography acquisitions.
FDA source Exact FDA submission Checked 2026-09-04

This is the interface boundary described in the exact FDA package; current local compatibility, routing, identity, archive, and downtime behavior still require acceptance testing.

Deployment and data flow
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

The exact public FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.

Security and privacy
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified radiologic technologists and interpreting physicians should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-04
Monitoring and change control
Monitor K202441 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The function is for supported projection radiography and excludes mammography; detector, acquisition, processing, and professional-review limits apply.
FDA source Exact FDA submission Checked 2026-09-04

Monitoring guidance is buyer-facing implementation guidance derived from the exact cleared scope; it is not a claim that the FDA package provides a live monitoring service.

Release and scope verification

Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K202441 before acceptance.

exact submission · checked 2026-09-04

Later product-family features are not evidence that this cleared release includes them.

Deployment boundary

The exact public deployment topology remains incomplete; establish hosting responsibility, data flow, availability, upgrade, service, and downtime boundaries locally.

exact submission · checked 2026-09-04

FDA-described deployment boundaries do not establish current hosting contracts, security controls, uptime, or local qualification.

Interoperability acceptance

The FDA package describes Smart Noise Cancellation as an optional ImageView function for supported CR and DR projection-radiography acquisitions.

exact submission · checked 2026-09-04

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The function is for supported projection radiography and excludes mammography; detector, acquisition, processing, and professional-review limits apply.

exact submission · checked 2026-09-04

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-04

No exact-product price was found in the reviewed public material; productivity or value claims require local evidence.

Lifecycle monitoring

Monitor K202441 by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The function is for supported projection radiography and excludes mammography; detector, acquisition, processing, and professional-review limits apply.

exact submission · checked 2026-09-04

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. radiographic image-processing software limits remain submission-specific.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
K202441 FDA decision material describes the cleared radiographic image-processing software, labeled use, and submission-specific verification. It does not establish product-wide sensitivity, specificity, outcomes, or independent clinical validation for this exact configuration.
FDA source Exact FDA submission Checked 2026-09-04
Reported sensitivity
Not applicable
Not applicable Not applicable Checked 2026-09-04

Sensitivity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, and source if later verified.

Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-04

Specificity is not promoted as a product-wide endpoint. Compare only an exact endpoint, threshold, population, version, and source if later verified.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K202441 FDA evidence package for Eclipse II with Smart Noise Cancellation

FDA 510(k) or PMA decision evidence and product verification described in the exact submission. · Adult and pediatric radiographic anatomy excluding mammography

exact submission · Tested version: Eclipse II with Smart Noise Cancellation configuration described in K202441 · Sponsor or vendor study

K202441

Manufacturer evidence submitted for the exact FDA decision. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging or clinical service line and verify payer, site-of-service, coding, and coverage assumptions independently.
Public source Not applicable Checked 2026-09-04

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

This is not a product-specific reimbursement determination. Build a local total-cost and value model and verify payer, site-of-service, coding, and coverage assumptions independently.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K202441 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source Exact FDA submission Checked 2026-09-04

Exact-identifier review does not prove absence of all product-family safety signals.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
26Fields reviewed
11Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Exact-name and exact-submission searches in the checked 2026-09-01 indexes returned no PubMed records and no ClinicalTrials.gov records for K202441; review status is candidate_review_pending. Literature and registry candidates remain quarantined unless exact submission and tested-version linkage is established. The checked recall disposition is no exact-submission recall record was promoted; ambiguous or family-only candidates are not published. Pricing, complete security architecture, contracted deployment, exact reimbursement, and local interoperability remain unverified. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, fda decision material, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-04.

Alternatives

Related tools to compare

FDA-cleared K251167

uDR Aurora CX

Shanghai United Imaging Healthcare Co., Ltd.

uDR Aurora CX acquires two-dimensional X-ray images of the labeled skull, spine, chest, abdomen, extremities, limbs, and trunk for review by qualified healthcare professionals; it...

Modality
X-ray/Radiography
Anatomy
Skull, spine, chest, abdomen, extremities, limbs, and trunk
Decision
2025-09-19
FDA-cleared K133259

YSIO MAX

SIEMENS MEDICAL SOLUTIONS USA, INC.

YSIO MAX is a digital radiography system for whole-body skull, chest, abdomen, and extremity imaging in pediatric, adult, and bariatric patients, including supported emergency and...

Modality
X-ray/Radiography
Anatomy
Multi-region
Decision
2014-01-24
FDA-cleared K253172

MediAI-OA

Crescom Co., Ltd.

FDA product code QIH identifies this device class as Automated Radiological Image Processing Software. To provide automated radiological image processing and analysis tools....

Modality
X-ray/Radiography
Anatomy
N/A
Decision
2026-06-15
FDA-cleared K253192

DeepXray Spina

Alpha Intelligence Manifolds, Inc.

FDA product code SAO identifies this device class as Radiology Software For Opportunistic Evaluation Of Low Bone Mineral Density. This device is software which opportunistically...

Modality
X-ray/Radiography
Anatomy
N/A
Decision
2026-06-10