FDA-cleared profile

Clarius Prostate AI

Clarius Mobile Health Corp.

Authorized purpose: Clarius Prostate AI provides semi-automatic measurements of prostate volume from ultrasound data acquired by the Clarius Ultrasound Scanner.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Clarius Prostate AI provides semi-automatic measurements of prostate volume from ultrasound data acquired by the Clarius Ultrasound Scanner.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Clarius Prostate AI provides semi-automatic measurements of prostate volume from ultrasound data acquired by the Clarius Ultrasound Scanner.

Evidence takeaway

Evidence from the reviewed source record: The FDA summary reports a retrospective, multicenter clinical-verification study of 139 male subjects, separate from model training and tuning. It reports 95% view-prediction accuracy and ICC values of 0.87 for endo-cavitary and 0.67 for curvilinear probes. These are manufacturer regulatory-study results, not independent clinical-outcome evidence.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Clarius Prostate AI provides semi-automatic measurements of prostate volume from ultrasound data acquired by the Clarius Ultrasound Scanner.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Clinical-use display mirrors the exact FDA-supported purpose reconciled from the decision-summary packet.

Intended users
Healthcare professionals trained and qualified in ultrasound.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Care setting and population
Adult male patients; the FDA labeling does not restrict facility type.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-14
Workflow role
Assistive ultrasound acquisition and interpretation support through segmentation, caliper placement, and semi-automatic prostate-volume measurement.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-14
Required input
B-mode prostate ultrasound data acquired with Clarius C3 HD3 or EC7 HD3 curvilinear or endo-cavitary scanners through the Clarius App.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Output and human action
Automatic prostate length, width, and height measurements, corresponding-view detection, segmentation, adjustable caliper placement, and calculated prostate volume.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Limitations
Limited to adult male patients, B-mode use, and the named Clarius scanner and transducer configurations. It is not standalone software, does not replace clinical decision-making, and the user must confirm measurements using standard practice and clinical judgment.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K243853
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Ultrasound
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Anatomy
Prostate
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2025-04-16
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Clarius Prostate AI is integrated into the Clarius App for iOS and Android and operates with the named Clarius Ultrasound Scanner configurations; it is not standalone software. Local downstream interface compatibility is not established by the FDA packet.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-14
Deployment and data flow
On-cloud
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-14

Current security controls, data retention, access model, and assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-14

Current training, competency, and support commitments require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-14

Site-specific quality monitoring, update control, escalation, and incident-response commitments require vendor and local governance evidence.

Integrated acquisition workflow

The FDA packet describes integration with the Clarius App and C3 HD3 or EC7 HD3 Clarius Ultrasound Scanner configurations for B-mode prostate ultrasound.

exact submission ยท checked 2026-09-14

Confirm current mobile operating-system support, configurations, and downstream interfaces locally.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The FDA summary reports a retrospective, multicenter clinical-verification study of 139 male subjects, separate from model training and tuning. It reports 95% view-prediction accuracy and ICC values of 0.87 for endo-cavitary and 0.67 for curvilinear probes. These are manufacturer regulatory-study results, not independent clinical-outcome evidence.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Clarius Prostate AI regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K243853
  • Samplen=139
  • IndependenceIndependence not established

K243853

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-14

No current exact-product price, contract term, or implementation fee was established.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-14

No exact-product payer policy, coding instruction, or payment determination was established.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Candidate records await review
  • Human review pending
  • Exact FDA submission
  • Checked 2026-09-14

Exact-submission postmarket candidates remain quarantined pending a separate source review.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
24Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. K243853 FDA 510(k) decision summary

    U.S. Food and Drug Administration - Published 2025-04-16 - Accessed 2026-09-19

    Scope: purpose, modality, anatomy, intendedUsers, workflowRole, input, output, careSetting, limitations, evidenceSummary, integration, clinicalUse

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Clarius Prostate AI model card (K243853)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, evidenceSummary, anatomy, evidenceStudies

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

Research history

  1. K243853 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K170250

SmartTarget

SmartTarget, Ltd.

SmartTarget is an accessory for image-guided interventional and diagnostic prostate procedures that enhances two-dimensional transrectal ultrasound, reconstructs three-dimensional...

Modality
Ultrasound
Anatomy
Prostate
Decision
2017-06-07
FDA-cleared K252237

EdgeFlow UW20

Edgecare, Inc.

EdgeFlow UW20 is a wearable ultrasound system for noninvasive periodic bladder-volume measurement in adults with BMI no greater than 30.

Modality
Ultrasound
Anatomy
Urinary bladder
Decision
2026-04-17