FDA-cleared profile

Clarius Ejection Fraction AI

Clarius Mobile Health Corp.

Authorized purpose: Semi-automated measurement of left ventricular ejection fraction during cardiac ultrasound.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: Semi-automated measurement of left ventricular ejection fraction during cardiac ultrasound.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: Semi-automated measurement of left ventricular ejection fraction during cardiac ultrasound.

Evidence takeaway

Evidence from the reviewed source record: FDA packet: retrospective clinical verification of 279 exams from a multi-center anonymized dataset, with three expert reviewers and separate clinical design validation in simulated use. The verification excluded inadequate/incomplete images and subjects for which the model failed to generate a measurement; it was a measurement comparison without follow-up clinical examination.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Semi-automated measurement of left ventricular ejection fraction during cardiac ultrasound.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Healthcare professionals trained and qualified in ultrasound.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Care setting and population
Professional healthcare setting such as a hospital or clinic.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Workflow role
Assistive real-time/semi-automatic EF measurement during cardiac ultrasound, with clinician adjustment and final responsibility.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Required input
Ultrasound image data acquired by Clarius phased-array or curvilinear scanners, including PLAX, PSAX, AP4, and AP2 views; the algorithm automatically captures ED/ES frames and is integrated into the Clarius App on iOS/Android.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Output and human action
Semi-automatic LV EF measurement using Simpson single-plane/apical, fractional area change/PSAX, and Teicholz/PLAX methods; segmentation masks or landmark markers identify the LV, ED/ES frames are captured, and the user can manually adjust caliper measurements.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Limitations
Adults only; not stand-alone software and limited to the listed Clarius scanner/app ecosystem and cleared transducers. Inadequate-quality images, incomplete anatomy/views, or failure to generate a measurement were excluded from the measurement comparison. The FDA packet also describes a PCCP for future scanner/input and algorithm changes, with data-skew and concept-drift risks requiring verification.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K253593
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Ultrasound
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Anatomy
Heart, specifically the left ventricle
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2026-03-02
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Integrated into the Clarius App and complete Clarius Ultrasound Scanner system, not a stand-alone device; compatible iOS/Android app versions and PA HD3, PAL HD3, and C3 HD3 transducers are listed in the packet. Generic DICOM/PACS integration is not established for the subject device.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Deployment and data flow
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for Clarius Ejection Fraction AI; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission ยท checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
FDA packet: retrospective clinical verification of 279 exams from a multi-center anonymized dataset, with three expert reviewers and separate clinical design validation in simulated use. The verification excluded inadequate/incomplete images and subjects for which the model failed to generate a measurement; it was a measurement comparison without follow-up clinical examination.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

Clarius Ejection Fraction AI regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K253593
  • Samplen=270
  • IndependenceIndependence not established

K253593

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

No public price or quote for the exact K253593 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-08

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K253593. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
21Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, acr ai central product, fda decision summary, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K253593: Clarius Ejection Fraction AI

    U.S. Food and Drug Administration - Published 2026-03-02 - Accessed 2026-09-08

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. Clarius Ejection Fraction AI model card (K253593)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, evidenceSummary, anatomy, evidenceStudies

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K253593

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

Research history

  1. K253593 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay