FDA-cleared profile

BrainInsight

Hyperfine, Inc. · MRI · FDA-cleared (510(k))

Authorized purpose: Automatically labels, measures, and quantifies brain structures from specified low-field MR images.

Clinical role
Segmentation or contouring
Images and patients
MRI · Brain
Key limitation
Keep in mind: Use is limited to the specified low-field scanner, protocols, sequences, and age/application boundaries.

Overview status: Source-reviewed overview

Report a correction
Clinical details Purpose, workflow fit, evidence, and safety record

Workflow navigation pilot

Workflow role has not yet been reviewed for this navigation pilot.

A tag identifies a documented role; it does not establish compatibility or local acceptance. Methods and sources

  • What does this software do?Authorized purpose: Automatically labels, measures, and quantifies brain structures from specified low-field MR images.
  • Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Brain Input: Specified low-field 0.064 T structural MRI sequences from the Hyperfine Swoop system.
  • Who uses it?Intended users: Clinicians.
  • What does the software produce?Output: Annotated and segmented images, color overlays, spatial measurements, and volumetric reports viewable through DICOM/PACS workflows.
  • What is its most important limitation?Key limitation: Use is limited to the specified low-field scanner, protocols, sequences, and age/application boundaries.

Clinical takeaway. Clinical role: Automatically labels, measures, and quantifies brain structures from specified low-field MR images.

Evidence takeaway. Evidence from the reviewed source record: Separate Swoop test data reported midline-shift mean absolute error of 0.75-1.01 mm and Dice overlap of 82-96% across listed structures and sequences; diagnostic sensitivity and specificity were not reported.

Purpose
Automatically labels, measures, and quantifies brain structures from specified low-field MR images.
  • Exact FDA submission
  • Checked 2026-09-15

Clinical-use display mirrors the exact FDA-supported purpose reconciled in this remediation finding.

Workflow role
Segmentation or contouring
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
Specified low-field 0.064 T structural MRI sequences from the Hyperfine Swoop system.
  • Exact FDA submission
  • Checked 2026-09-15

Exact scanner, protocol, and sequence scope.

Output and clinician action
Annotated and segmented images, color overlays, spatial measurements, and volumetric reports viewable through DICOM/PACS workflows.
  • Exact FDA submission
  • Checked 2026-09-15

FDA output description.

Care setting
Adult and Pediatric
  • Exact FDA submission
  • Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Intended users
Clinicians.
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet user statement.

Limitations
Use is limited to the specified low-field scanner, protocols, sequences, and age/application boundaries.
  • Exact FDA submission
  • Checked 2026-09-15

Exact packet boundary; universal scanner support was not established.

Evidence summary
Separate Swoop test data reported midline-shift mean absolute error of 0.75-1.01 mm and Dice overlap of 82-96% across listed structures and sequences; diagnostic sensitivity and specificity were not reported.
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet performance summary.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Separate Swoop test data reported midline-shift mean absolute error of 0.75-1.01 mm and Dice overlap of 82-96% across listed structures and sequences; diagnostic sensitivity and specificity were not reported.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA packet performance summary.

Reported sensitivity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

Reported specificity
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

BrainInsight regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationAdult and Pediatric
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K223268
  • IndependenceIndependence not established

K223268

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K223268. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-09
Pricing & contracts Public price, contract terms, and reimbursement

Price not public here; enter your quote

Not verified: Pricing information is recorded but not verified against a resolving source.

Reimbursement

IT implementation Deployment, interfaces, data handling, and support

System and version relationships

Match the AI release, system build, configuration and region. Source documentation does not establish acceptance of an installed system.

No specific system/version pairing has completed review for this FDA record. A broad DICOM or product-family statement does not establish a named viewer build's support.

Review Merge, Visage, Sectra, AGFA and IntelePACS context

Local implementation checklist

Agree the evidence and acceptance owner with your team and suppliers. Checkmarks apply to this page only and are not acceptance receipts.

Open the printable evaluation checklist

Deployment
On-cloud
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Integration
The exact-submission ACR model card explicitly references DICOM and PACS in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
  • Exact FDA submission
  • Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Security
Hyperfine's current certifications page says the Swoop System device services, security services and Hyperfine PACS services (hosted on AWS, Ashburn, Virginia) meet HITRUST CSF v9.4 r2 criteria, and that a SOC 2 Type II report on the Swoop system (security criteria only) is available under NDA. The page gives no certification dates; neither report was read.
  • Vendor-wide
  • Checked 2026-10-25

Vendor-wide statement, not an independent attestation. It does not establish certificate or report scope for this exact release, expiry, data location, retention, identity and access, or incident-response terms for a deployed configuration. Obtain these from the vendor.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Result destination

PACS is explicitly referenced as a result destination.

exact submission · checked 2026-09-01

Confirm supported products, versions, routing behavior, and display requirements during local evaluation.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for BrainInsight; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-09
Sources & review details FDA identity, research record, source ledger, and history

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K223268
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
MRI
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

Low-field structural MR normalized to the controlled MRI bucket.

Anatomy
Brain
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-15

FDA brain-structure and ventricle scope.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2022-12-16
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
25Fields reviewed
11Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, vendor product page, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Field review details

Recorded state, verification date, notes and sources for fields without a publishable value. Conflicting records stay explicitly unsettled here; no plain settled value is shown for them.

  • Public pricing: Not verified - checked 2026-09-09 - Exact FDA submission

    No public price or quote for the exact K223268 configuration was established in this review.

  • Pricing model: Not yet reviewed
  • Setup and integration fees: Not yet reviewed
  • Recurring fees: Not yet reviewed
  • Minimum commitment: Not yet reviewed
  • Contract duration: Not yet reviewed
  • Renewal terms: Not yet reviewed
  • Cancellation terms: Not yet reviewed
  • Support and service terms: Not yet reviewed
  • Trial or pilot terms: Not yet reviewed
  • Reimbursement: Not publicly available - checked 2026-09-09 - Exact FDA submission

    No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

  • Data hosting location: Not yet reviewed
  • Data retention: Not yet reviewed
  • Documented data use: Not yet reviewed
  • Hardware requirements: Not yet reviewed
  • Network requirements: Not yet reviewed
  • Identity and access: Not yet reviewed
  • Installation timeline: Not yet reviewed
  • Training: Not verified - checked 2026-09-09 - Exact FDA submission

    Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

  • Monitoring: Not verified - checked 2026-09-09 - Exact FDA submission

    Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

8 sources - 3 history items

Source ledger

  1. FDA K223268: BrainInsight

    U.S. Food and Drug Administration - Published 2022-12-16 - Accessed 2026-09-09

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. BrainInsight model card (K223268)

    American College of Radiology - Published 2026-08-17 - Accessed 2026-09-01

    Scope: intendedUsers, careSetting, output, limitations, evidenceSummary, integration, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K223268

    U.S. Food and Drug Administration - Accessed 2026-09-09

    Scope: postmarketSafety

  8. Hyperfine Certifications

    Hyperfine, Inc. - Accessed 2026-10-25

    Scope: security

Research history

  1. K223268 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K261405

ANDI 2.2

Imeka Solutions, Inc.

Authorized purpose: ANDI 2.2 processes brain MR images for atlas-assisted visualization, segmentation and volumetric quantification, presenting adjunctive DICOM-readable information for trained medical professionals; it does not predict, diagnose or interpret brain health.

Modality
MRI
Anatomy
Brain
Decision
2026-05-28
FDA-cleared K260032

Morph

BeauBrain Healthcare, Inc.

Authorized purpose: Morph automatically labels, visualizes, and volumetrically quantifies segmentable brain structures from MR images and may compare measurements with reference percentile data; it is adjunctive and not the sole basis for final decisions.

Modality
MRI
Anatomy
Brain
Decision
2026-05-13
FDA-cleared K253057

AI-Rad Companion Brain MR

Siemens Healthcare GmbH

Authorized purpose: Brain MRI morphometry and white matter hyperintensity quantification, including normative comparison and longitudinal change assessment for clinician review.

Modality
MRI
Anatomy
Brain
Decision
2026-01-22
FDA-cleared K252298

ANDI 2.0

Imeka Solutions, Inc.

Authorized purpose: Adjunctive quantitative brain MRI analysis that compares patient measures with normative and longitudinal references; a physician reviews all results and makes the clinical decision.

Modality
MRI
Anatomy
Brain
Decision
2025-10-22