FDA-cleared profile

BlineSlide

Deep Breathe, Inc.

The device is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients aged 18 years or older with suspected disease.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
The device is intended for noninvasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients aged 18 years or older with suspected disease.
Public source Exact FDA submission Checked 2026-09-01

Exact-submission indication text from the ACR AI Central model card. FDA labeling remains controlling.

Intended users
Physicians credentialed to use lung ultrasound clinically, such as intensivists, emergency physicians, pulmonologists, general practitioners, or other clinicians interpreting the source lung ultrasound cines.
Public source Exact FDA submission Checked 2026-09-01

Intended-user text reported in the exact-submission ACR model card.

Care setting and population
Adult
Public source Exact FDA submission Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Measurement or quantification
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Required input
US images or data. Inclusion criteria: User-uploaded B-Mode lung ultrasound (LUS) video clips (cines) in MP4 format acquired from the anterior or anterolateral chest regions using non-linear array ultrasound transducers, acquired at 18 or more frames per second, requiring no more than 2048 megabytes of memory, and between 2600 and 7800 milliseconds in duration.
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Output and human action
The AI model outputs detection of the presence or absence of B-line artifacts in lung ultrasound cines. A table summarizing the device outputs for each uploaded cine is displayed to the user. If the cine does not meet input criteria, a 'Cannot Evaluate' output is returned.
Public source Exact FDA submission Checked 2026-09-01

Clinical output reported in the exact-submission ACR model card.

Limitations
Exclusion criteria: Cines acquired from linear array ultrasound transducers; cines acquired at less than 18 frames per second; cines that require more than 2048 megabytes of memory; cines that are less than 2600 milliseconds in duration; cines that are greater than 7800 milliseconds in duration.
Public source Exact FDA submission Checked 2026-09-01

Limitations and supported-acquisition details reported in the exact-submission ACR model card.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K252007
FDA source Exact FDA submission Checked 2026-08-31
Modality
Ultrasound
Public source Exact FDA submission Checked 2026-09-01

Imaging modality reported in the exact-submission ACR AI Central model card.

Anatomy
Chest/Lung
Public source Exact FDA submission Checked 2026-09-01

Anatomy normalized from explicit exact-submission ACR model-card wording.

Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2025-10-06
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Standards-based DICOM/PACS integration was identified in the FDA submission materials; compatibility with a specific PACS, RIS, EHR, or orchestrator requires confirmation.
Public source Exact FDA submission Checked 2026-08-13

Normalized from the prior exact-submission review. Confirm interface versions and local compatibility during technical evaluation.

Deployment and data flow
On-cloud
Public source Exact FDA submission Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA authorization does not establish the current commercial security posture. Request data-flow diagrams, hosting and retention details, access controls, audit logging, attestations, and incident-response documentation.

Training and support
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Training, onboarding, competency, and support commitments are commercial implementation details that require current vendor documentation.

Monitoring and change control
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Local acceptance testing, production monitoring, drift review, update control, and incident escalation require a current vendor and site-specific governance plan.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
Study type: Stand-Alone Performance. Cases: 1005. Reported sensitivity: 91% (95% CI: 88%, 94%). Reported specificity: 84% (95% CI: 81%, 86%)
Public source Exact FDA submission Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Reported sensitivity
91% (95% CI: 88%, 94%)
Public source Exact FDA submission Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
84% (95% CI: 81%, 86%)
Public source Exact FDA submission Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

BlineSlide regulatory performance study summary

Stand-Alone Performance · Adult

exact submission · Tested version: Exact product version not reported; regulatory study summarized for FDA submission K252007 · n=1005 · Independence not established

K252007

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA submission materials do not establish current commercial pricing. Request the pricing unit, term, minimum commitment, implementation fees, and included support from the vendor.

Reimbursement and coding
Research pending
Needs research Exact FDA submission Checked 2026-08-31

Determine whether any product-specific CPT/HCPCS pathway exists and separate it from billing for the underlying imaging service.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-04Last searched
19Fields reviewed
6Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-02.

Alternatives

Related tools to compare

FDA-cleared K251368

FETOLY

Diagnoly

FETOLY analyzes fetal ultrasound images and sequences to detect labeled fetal brain and heart views, assess image completeness or quality, and provide fetal biometric measurements...

Modality
Ultrasound
Anatomy
Fetal heart and fetal brain
Decision
2025-09-12