VisAble.IO is CT and MR image-processing software for liver-ablation procedures. It imports images from CT and MR scanners and facility PACS systems, assists physicians with ablation-target identification and virtual ablation-needle placement, and assists confirmation of ablation zones.
Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.
Clinical fit
What this tool is for
Start with the authorized purpose, then verify how it fits your service line and reading workflow.
Exact purpose
VisAble.IO is CT and MR image-processing software for liver-ablation procedures. It imports images from CT and MR scanners and facility PACS systems, assists physicians with ablation-target identification and virtual ablation-needle placement, and assists confirmation of ablation zones.
FDA source Exact FDA submission Checked 2026-09-04
The FDA indication states that VisAble.IO is not intended for diagnosis and does not predict ablation volumes or ablation success. Planning and treatment-confirmation outputs require physician review and do not replace procedural judgment.
FDA source Exact FDA submission Checked 2026-09-04
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
FDA submission
K240773
FDA source Exact FDA submission Checked 2026-08-31
Use these fields to structure a vendor demo, security review, and implementation estimate.
Integration
The FDA indication describes CT and MR import from scanners and facility PACS systems for planning and confirmation; validate patient identity, registration, image geometry, source selection, export, and procedure-room fallback locally.
FDA source Exact FDA submission Checked 2026-09-04
The exact FDA material describes this input or interface boundary; local interoperability acceptance remains necessary.
Reviewed sources checked Exact FDA submission Checked 2026-09-04
The exact FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Security and privacy
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04
Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.
Training and support
Only qualified surgeons, radiologists, dentists, radiation professionals, and other trained users for the labeled planning workflow should operate or interpret this product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-04
Monitor by exact hardware, software, model, algorithm, reconstruction, or configuration version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The FDA indication states that VisAble.IO is not intended for diagnosis and does not predict ablation volumes or ablation success. Planning and treatment-confirmation outputs require physician review and do not replace procedural judgment.
FDA source Exact FDA submission Checked 2026-09-04
Map installed model, build, options, population, anatomy, inputs, outputs, and enabled functions to K240773 before acceptance. Later family features are not evidence for this exact submission.
The FDA indication describes CT and MR import from scanners and facility PACS systems for planning and confirmation; validate patient identity, registration, image geometry, source selection, export, and procedure-room fallback locally. Test representative identifiers, geometry, measurements, units, routing, failure handling, and round trips locally.
The exact FDA material does not establish hosting, deployment topology, availability, upgrade path, or service responsibility. Obtain site-specific evidence before procurement.
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The FDA indication states that VisAble.IO is not intended for diagnosis and does not predict ablation volumes or ablation success. Planning and treatment-confirmation outputs require physician review and do not replace procedural judgment.
exact submission · checked 2026-09-04
Define stop-use, escalation, rollback, and revalidation triggers before production use.
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs. No exact-product price was found.
Monitor by exact hardware, software, model, algorithm, reconstruction, or configuration version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: The FDA indication states that VisAble.IO is not intended for diagnosis and does not predict ablation volumes or ablation success. Planning and treatment-confirmation outputs require physician review and do not replace procedural judgment.
exact submission · checked 2026-09-04
Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.
Evidence summary
The exact FDA 510(k) decision material for K240773 identifies VisAble.IO and describes the indicated use summarized in this dossier. It is regulatory evidence for the exact submission, not independent clinical validation. No performance or outcome claim is promoted without exact product and version linkage.
FDA source Exact FDA submission Checked 2026-09-04
Sensitivity is not a product-wide endpoint for this system or software function. Compare only an exact endpoint, threshold, population, version, and source if later verified.
Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-04
Specificity is not a product-wide endpoint for this system or software function. Compare only an exact endpoint, threshold, population, version, and source if later verified.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
K240773 FDA decision evidence for VisAble.IO
FDA 510(k) substantial-equivalence review of the exact submission, device description, indications, and verification or validation documentation. · Liver, hepatic and portal veins, ablation targets, and ablation zones
exact submission · Tested version: VisAble.IO · Sponsor or vendor study
K240773
The FDA decision material is manufacturer-submitted regulatory evidence. It is not independent clinical validation, and local technical, clinical, privacy, security, and workflow acceptance remains necessary.
These are common procurement questions; unknown values remain visible until a source supports them.
Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04
No public list price for the exact cleared or approved configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, surgical, or diagnostic service line and verify payer, site-of-service, and coding assumptions independently.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K240773 by exact native submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source Exact FDA submission Checked 2026-09-04
Exact-identifier review does not prove absence of all product-family safety signals.
Verify deployment and data flow: The exact FDA decision material does not establish the complete deployment topology, hosting responsibility, availability, upgrade path, or service contract.
Verify security and privacy controls: Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.
Verify pricing and total cost: No public list price for the exact cleared or approved configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-04Last searched
32Fields reviewed
10Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads
Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact-name and exact-submission searches in the checked indexes returned no PubMed records and no ClinicalTrials.gov records for K240773; no published exact-submission recall was promoted. Candidate literature, registry, and non-native recall records remain quarantined unless exact product, version, and submission linkage is established. The FDA decision material remains the identity and intended-use anchor for VisAble.IO. Pricing, complete security architecture, contracted deployment, exact reimbursement, and local interoperability remain unverified. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads remain unpublished until a human confirms the exact product and tested version.
Source classes: fda ai list, fda decision summary, fda decision material, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall
ADAS 3D supports visualization and analysis of MR and CT images of the heart for individual patients with cardiovascular disease, including patients with myocardial scar from...
Modality
Radiological images; modality varies or is not specified in FDA list
Cercare Neurosuite with Capillary Function and Oncology Virtual Expert processes CT or MR information to provide perfusion, tissue-blood-volume, diffusion, labeling, and...
Modality
Radiological images; modality varies or is not specified in FDA list
HistoSonics Planning Tool is a stand-alone visualization and comparison tool for CT and MR studies used to plan histotripsy target localization, treatment volume, acoustic path,...
Modality
Radiological images; modality varies or is not specified in FDA list