FDA-cleared profile

Venue

GE Medical Systems

Venue provides mobile point-of-care diagnostic ultrasound; this release adds adult cNerve scouting, pediatric B-line and real-time ejection-fraction use, probe checks, catheter sizing, and workflow tools.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Venue provides mobile point-of-care diagnostic ultrasound; this release adds adult cNerve scouting, pediatric B-line and real-time ejection-fraction use, probe checks, catheter sizing, and workflow tools.
FDA source Exact FDA submission Checked 2026-09-03
Intended users
Qualified and trained physicians and point-of-care ultrasound professionals.
FDA source Exact FDA submission Checked 2026-09-03
Care setting and population
Hospitals and medical clinics, including space-constrained point-of-care settings.
FDA source Exact FDA submission Checked 2026-09-03
Workflow role
Broad diagnostic ultrasound acquisition and measurement with automated point-of-care workflow and scouting support.
Public source Exact FDA submission Checked 2026-09-03
Required input
Live ultrasound from supported linear, convex, and phased-array probes plus optional ECG gating input.
FDA source Exact FDA submission Checked 2026-09-03
Output and human action
Images, measurements, reports, DICOM studies, and release-specific nerve, B-line, ejection-fraction, and workflow outputs.
FDA source Exact FDA submission Checked 2026-09-03
Limitations
cNerve supports adult nerve scouting before needle insertion and is not live interventional guidance. Pediatric B-line and real-time EF exclude neonates; ECG is for image synchronization, not monitoring or diagnosis.
FDA source Exact FDA submission Checked 2026-09-03

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K220851
FDA source Exact FDA submission Checked 2026-08-31
Modality
Ultrasound
FDA source Exact FDA submission Checked 2026-09-03
Anatomy
Multi-region
FDA source Exact FDA submission Checked 2026-09-03
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2022-06-27
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
DICOM storage, archiving, printing, wireless communication, barcode and optional RFID workflows; test worklist, patient matching, archive, probe inventory, ECG gating, and role-based access.
Public source Exact FDA submission Checked 2026-09-03
Deployment and data flow
Mobile battery-powered ultrasound system with cleanable touch interface, articulated display, supported probes, and network options.
Public source Exact FDA submission Checked 2026-09-03
Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03

Public materials reviewed for this batch do not establish the complete security posture for the exact release. Request data-flow, identity, audit, encryption, retention, support-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified and trained physicians and point-of-care ultrasound professionals. should operate or interpret the system. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track the limits that matter for this product: cNerve supports adult nerve scouting before needle insertion and is not live interventional guidance. Pediatric B-line and real-time EF exclude neonates; ECG is for image synchronization, not monitoring or diagnosis.
Public source Exact FDA submission Checked 2026-09-03

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, and outputs to K220851 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

DICOM storage, archiving, printing, wireless communication, barcode and optional RFID workflows; test worklist, patient matching, archive, probe inventory, ECG gating, and role-based access.

exact submission · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. cNerve supports adult nerve scouting before needle insertion and is not live interventional guidance. Pediatric B-line and real-time EF exclude neonates; ECG is for image synchronization, not monitoring or diagnosis.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

Vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track the limits that matter for this product: cNerve supports adult nerve scouting before needle insertion and is not live interventional guidance. Pediatric B-line and real-time EF exclude neonates; ECG is for image synchronization, not monitoring or diagnosis.

exact submission · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
K220851 reports independent cNerve verification on 3,776 frames from 124 sequences contributed by 44 adults aged 18-82 across the United States, Japan, and Israel. The study used prespecified sequence and frame-detection requirements and subgroup checks; no clinical study was required.
Public source Exact FDA submission Checked 2026-09-03
Reported sensitivity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Sensitivity is not the primary applicable endpoint for this acquisition, reconstruction, segmentation, planning, guidance, or quantitative-biomarker function. Use the task-specific endpoints and tested population reported below.

Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Specificity is not the primary applicable endpoint for this acquisition, reconstruction, segmentation, planning, guidance, or quantitative-biomarker function. Use the task-specific endpoints and tested population reported below.

cNerve sequences

124 sequences

Adult nerve-scouting detection in Venue.

Tested version: Venue release cleared in K220851

cNerve frames

3776 frames

Frame-level nerve detection in Venue.

Tested version: Venue release cleared in K220851

Successful detected-sequence criterion

80 %

Meaningfully detected sequences meeting frame criteria in Venue.

Tested version: Venue release cleared in K220851

K220851 FDA performance package for Venue

Manufacturer held-out multiregion cNerve sequence and frame verification against expert annotations · 44 adults aged 18-82 across supported nerve-block locations, probes, sex, age, and BMI groups

exact submission · Tested version: Venue release cleared in K220851 · n=44 · Sponsor or vendor study

K220851

Manufacturer evidence submitted for substantial equivalence; local clinical, technical, and workflow acceptance remains necessary.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, enabled features, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the device within the applicable imaging or surgical service line and verify payer, site-of-service, and coding assumptions independently.
Public source Not applicable Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K220851 by exact submission identifier.
Public source Exact FDA submission Checked 2026-09-03

A zero-result exact-identifier search does not prove that no recall, correction, adverse-event report, or product-family safety signal exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
24Fields reviewed
8Source classes checked
20Unreviewed PubMed leads
10Unreviewed trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact FDA materials, current product-family sources, selected clinical literature, reimbursement context, and exact-identifier recall candidates were reviewed. Pricing, contracted services, enabled options, final security architecture, and local interoperability remain organization-specific evidence. Automated exact-name discovery found 20 PubMed and 10 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, vendor product page, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-03.

Alternatives

Related tools to compare

FDA-cleared K251963

LOGIQ E10s

Ge Medical Systems Ultrasound and Primary Care Diagnostics

LOGIQ E10s is a general-purpose diagnostic ultrasound system supporting the labeled adult, pediatric, cardiac, vascular, abdominal, obstetric, small-organ, musculoskeletal,...

Modality
Ultrasound
Anatomy
Multi-region
Decision
2025-10-29
FDA-cleared K251985

LOGIQ E10

Ge Medical Systems Ultrasound and Primary Care Diagnostics

LOGIQ E10 is a general diagnostic ultrasound system for qualified users across fetal and obstetric, abdominal, gynecologic, pediatric, small-organ, breast, cardiac, vascular,...

Modality
Ultrasound
Anatomy
Multi-region
Decision
2025-10-29