FDA-cleared profile

VELYS™ Hip Navigation

Depuy Ireland UC

Authorized purpose: VELYS Hip Navigation analyzes intraoperative X-ray or fluoroscopy for total hip arthroplasty with surgeon-confirmed landmark suggestions and measurements of cup orientation, leg length, offset, and implant constructs.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: VELYS Hip Navigation analyzes intraoperative X-ray or fluoroscopy for total hip arthroplasty with surgeon-confirmed landmark suggestions and measurements of cup orientation, leg length, offset, and implant constructs.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: VELYS Hip Navigation analyzes intraoperative X-ray or fluoroscopy for total hip arthroplasty with surgeon-confirmed landmark suggestions and measurements of cup orientation, leg length, offset, and implant constructs.

Evidence takeaway

Evidence from the reviewed source record: K253551 reports landmark, measurement, workflow, software, and usability validation with human confirmation of suggested landmarks.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
VELYS Hip Navigation analyzes intraoperative X-ray or fluoroscopy for total hip arthroplasty with surgeon-confirmed landmark suggestions and measurements of cup orientation, leg length, offset, and implant constructs.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Clinical-use display mirrors the exact FDA-supported purpose reconciled from the decision-summary packet.

Intended users
Healthcare professionals.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication names healthcare professionals and does not establish the current generalized qualified-role list.

Care setting and population
Hospital imaging, interventional, operating-room, or procedure-suite workflows within the cleared indication.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Workflow role
Image reconstruction, localization, navigation, or procedural guidance under direct professional control.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Required input
Preoperative and intraoperative radiographic patient images of the pelvis.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact comparison table states this image-input scope and does not support CT or ultrasound inputs.

Output and human action
Software provides leg length and offset data for selected implant constructs and alternatives; software provides inclination and anteversion of the trial or final acetabular implant; manual landmark selection or AI assisted landmarking.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact comparison table states these outputs and landmark modes, not the current generalized navigation-guidance list.

Limitations
Clinical judgment and experience are required to properly use the software. The software is not for primary image interpretation. The software is not for use on mobile phones. This feature is available in Windows Version only.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact packet states these use boundaries; it does not state that the surgeon remains responsible for implant selection and placement.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K253551
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
Fluoroscopy/Angiography
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact device description states x-ray or fluoroscopy images and does not establish angiography or CT/ultrasound input. Controlled directory display reconciled from exact FDA wording: X-ray or fluoroscopy images. -> Fluoroscopy/Angiography.

Anatomy
Hip
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2026-03-06
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Validate compatible imaging hardware, acquisition geometry, displays, workstation, exported objects, latency, calibration, and fallback workflow.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-03
Deployment and data flow
Procedure-room hardware and software or a connected workstation with supported imaging equipment and accessories.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-03
Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-03

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified physicians, technologists, or procedural staff trained for the specified intervention. should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: The surgeon must review and confirm landmarks and remains responsible for implant selection and placement. The software is not for primary image interpretation or mobile-phone use; exact K253551 AI use is Windows-based.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-03

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, and outputs to K253551 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

Validate compatible imaging hardware, acquisition geometry, displays, workstation, exported objects, latency, calibration, and fallback workflow.

exact submission · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The surgeon must review and confirm landmarks and remains responsible for implant selection and placement. The software is not for primary image interpretation or mobile-phone use; exact K253551 AI use is Windows-based.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

Vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: The surgeon must review and confirm landmarks and remains responsible for implant selection and placement. The software is not for primary image interpretation or mobile-phone use; exact K253551 AI use is Windows-based.

exact submission · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
K253551 reports landmark, measurement, workflow, software, and usability validation with human confirmation of suggested landmarks.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Sensitivity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Specificity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K253551 FDA performance package for VELYS™ Hip Navigation

  • DesignManufacturer image-analysis, landmark, measurement, workflow, and usability validation
  • PopulationAdult total hip arthroplasty intraoperative radiographic workflows
  • Scopeexact submission
  • Tested versionVELYS Hip Navigation release cleared in K253551
  • IndependenceSponsor or vendor study

K253551

Manufacturer evidence submitted for clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, enabled features, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
  • Public source
  • Not applicable
  • Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K253551 by exact submission identifier. This does not exclude product-family or later safety information.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-03

Exact-identifier review does not prove absence of all product-family safety signals.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
27Fields reviewed
11Source classes checked
0Screened PubMed leads
1Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 0 PubMed and 1 ClinicalTrials.gov candidate record. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 0 PubMed and 1 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision material, vendor product page, reimbursement policy, fda device recall, fda decision summary packet, literature index, trial registry, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K253551: VELYS™ Hip Navigation

    U.S. Food and Drug Administration - Published 2026-03-06 - Accessed 2026-09-19

    Scope: purpose, anatomy, evidenceSummary, modality, intendedUsers, input, output, limitations, clinicalUse

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  4. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  5. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. VELYS Hip Navigation solutions

    Johnson & Johnson MedTech - Accessed 2026-09-03

    Scope: integration, deployment, monitoring, productFamilyContext

Research history

  1. K253551 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K260077

OrthoGrid Hip AI® 4.0

OrthoGrid Systems, Inc.

OrthoGrid Hip AI 4.0 analyzes two-dimensional C-arm fluoroscopic images to assist positioning of hip-replacement components during direct-anterior total hip arthroplasty.

Modality
Fluoroscopy/Angiography
Anatomy
Hip
Decision
2026-05-29
FDA-cleared K252500

CARA System

Cara Medical, Ltd.

CARA assists adult structural-heart and conduction-system procedure planning by segmenting preprocedural CT, visualizing cardiac chambers and conduction landmarks, and overlaying...

Modality
Fluoroscopy/Angiography
Anatomy
Heart/Cardiovascular
Decision
2026-02-20
FDA-cleared K243769

QFR (3.0)

QFR Solutions bv

QFR (3.0) uses paired coronary X-ray angiographic DICOM acquisitions to calculate three-dimensional vessel and lesion measurements and pressure-drop quantification for...

Modality
Fluoroscopy/Angiography
Anatomy
Heart/Cardiovascular
Decision
2025-04-04
FDA-cleared K252503

Intelligent NR

Canon, Inc.

Canon Intelligent NR is a noise-reduction function for fluoroscopic images acquired on compatible Canon radiography and fluoroscopy systems.

Modality
Fluoroscopy/Angiography
Anatomy
Whole body; diagnostic and interventional anatomy
Decision
2026-04-30