FDA-cleared profile

UltraSight AI Guidance

UltraSight Inc.

UltraSight AI Guidance assists non-expert healthcare professionals with acquisition of ten standard adult two-dimensional transthoracic echocardiography views by providing real-time transducer-position and orientation guidance on a compatible ultrasound system.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
UltraSight AI Guidance assists non-expert healthcare professionals with acquisition of ten standard adult two-dimensional transthoracic echocardiography views by providing real-time transducer-position and orientation guidance on a compatible ultrasound system.
FDA source Exact FDA submission Checked 2026-09-03
Intended users
Qualified physicians, surgeons, technologists, or procedural staff trained for the specified intervention.
FDA source Exact FDA submission Checked 2026-09-03
Care setting and population
Hospital imaging, interventional, operating-room, or procedure-suite workflows within the cleared indication.
FDA source Exact FDA submission Checked 2026-09-03
Workflow role
Image reconstruction, localization, navigation, or procedural guidance under direct professional control.
FDA source Exact FDA submission Checked 2026-09-03
Required input
Supported radiographic, fluoroscopic, CT, or ultrasound images plus procedure-specific user inputs.
FDA source Exact FDA submission Checked 2026-09-03
Output and human action
Images, landmarks, measurements, overlays, or navigation guidance for real-time professional review.
FDA source Exact FDA submission Checked 2026-09-03
Limitations
Guidance supports image acquisition and does not replace a qualified cardiac examination, image-quality review, interpretation, or clinical decision. Confirm compatible ultrasound host, ten-view scope, adult population, and fallback workflow.
FDA source Exact FDA submission Checked 2026-09-03

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K223347
FDA source Exact FDA submission Checked 2026-08-31
Modality
Ultrasound
FDA source Exact FDA submission Checked 2026-09-03
Anatomy
Heart/Cardiovascular
FDA source Exact FDA submission Checked 2026-09-03
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2023-07-24
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Validate compatible imaging hardware, acquisition geometry, displays, workstation, exported objects, latency, calibration, and fallback workflow.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Deployment and data flow
Procedure-room hardware and software or a connected workstation with supported imaging equipment and accessories.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified physicians, surgeons, technologists, or procedural staff trained for the specified intervention should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Guidance supports image acquisition and does not replace a qualified cardiac examination, image-quality review, interpretation, or clinical decision. Confirm compatible ultrasound host, ten-view scope, adult population, and fallback workflow.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K223347; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled AI functions to K223347 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

Validate compatible imaging hardware, acquisition geometry, displays, workstation, exported objects, latency, calibration, and fallback workflow.

product family · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Guidance supports image acquisition and does not replace a qualified cardiac examination, image-quality review, interpretation, or clinical decision. Confirm compatible ultrasound host, ten-view scope, adult population, and fallback workflow.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

No exact-product price was found in the reviewed public material; vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Guidance supports image acquisition and does not replace a qualified cardiac examination, image-quality review, interpretation, or clinical decision. Confirm compatible ultrasound host, ten-view scope, adult population, and fallback workflow.

product family · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
K223347 reports summative usability validation in which 61 subjects were scanned twice, comparing non-expert acquisition with AI guidance against cardiac-sonographer acquisition without AI guidance. This is an acquisition-guidance and usability evaluation, not evidence that the system independently diagnoses cardiac disease.
FDA source Exact FDA submission Checked 2026-09-03
Reported sensitivity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Sensitivity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Specificity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Usability-validation subjects

61 subjects

Adult TTE acquisition comparison in UltraSight AI Guidance.

Tested version: UltraSight AI Guidance

K223347 FDA performance package for UltraSight AI Guidance

Summative usability validation of AI-guided adult transthoracic echocardiography acquisition · 61 subjects scanned twice by non-experts with AI guidance and by a cardiac sonographer without AI guidance

exact submission · Tested version: UltraSight AI Guidance · Sponsor or vendor study

K223347

Manufacturer evidence submitted for FDA clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

ENABLE-HCM - AI-ENabled Echocardiography With Ultrasound Beyond the Echo Lab for Better HCM Imaging and Expanded Access

ClinicalTrials.gov candidate identified by exact-name search; full product and version applicability review pending · Heart/Cardiovascular or related Ultrasound workflow; exact product and version linkage not established

product family · Tested version: Candidate source; exact FDA submission and product version not established · Independence not established

NCT07155434

Candidate source was found by automated exact-name search; product or version linkage and clinical endpoint were not independently established, so it must not be treated as exact-submission performance evidence.

Human Factors Study of Ultrasound Navigation Software for Cardiac Imaging

ClinicalTrials.gov candidate identified by exact-name search; full product and version applicability review pending · Heart/Cardiovascular or related Ultrasound workflow; exact product and version linkage not established

product family · Tested version: Candidate source; exact FDA submission and product version not established · Independence not established

NCT05660044

Candidate source was found by automated exact-name search; product or version linkage and clinical endpoint were not independently established, so it must not be treated as exact-submission performance evidence.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
Public source Not applicable Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K223347 by exact submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source Exact FDA submission Checked 2026-09-03

Exact-identifier review does not prove absence of all product-family safety signals.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
25Fields reviewed
11Source classes checked
0Unreviewed PubMed leads
3Unreviewed trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact FDA decision materials were reviewed for identity, indication, users, workflow, limitations, and submitted performance. Current vendor material is labeled product-family context where it does not establish the exact release. PubMed and ClinicalTrials.gov results are retained as candidate evidence with version applicability pending; CMS reimbursement context and exact-identifier openFDA postmarket results were also checked. Pricing, final security architecture, contracted services, enabled options, and local interoperability remain organization-specific evidence. Automated exact-name discovery found 0 PubMed and 3 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-03.

Alternatives

Related tools to compare

FDA-cleared K260780

HeartFocus

Deski

Guides medical professionals acquiring ten standard 2D transthoracic echocardiographic views in adults requiring a cardiac ultrasound examination; acquired images require...

Modality
Ultrasound
Anatomy
Heart/Cardiovascular
Decision
2026-06-03
FDA-cleared K254161

Automated Aortic Stenosis Software (AutoAS)

GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC

FDA product code POK identifies this device class as Computer-Assisted Diagnostic Software For Lesions Suspicious For Cancer. Assist users in characterizing lesions identified on...

Modality
Ultrasound
Anatomy
Heart/Cardiovascular
Decision
2026-03-27
FDA-cleared K253535

Ligence Heart

Ligence Uab

FDA product code QIH identifies this device class as Automated Radiological Image Processing Software. To provide automated radiological image processing and analysis tools....

Modality
Ultrasound
Anatomy
Heart/Cardiovascular
Decision
2026-02-27