Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.
Product overview
Start with the essentials
Authorized purpose: Process and enhance CT images from any CT scanner manufacturer or model as an adjunct to the original reconstructed image.
Overview status: Source-reviewed overview
What does this software do?Authorized purpose: Process and enhance CT images from any CT scanner manufacturer or model as an adjunct to the original reconstructed image.
Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Multiple CT anatomies: abdomen, C-spine, cardiac, chest, chest-abdomen-pelvis, head, head/neck, L-spine, lower extremity, MSK, T-spine, and upper extremity. Input: Primary reconstructed DICOM CT images sent by a technologist through a compatible MDDS; FBP primary reconstructions at standard or higher dose were quantitatively validated. The packet also includes varied scanners, anatomies, contrast, reconstruction methods, and slice thicknesses, but warns that IR/DL reconstruction performance may be unknown.
Who uses it?Intended users: Radiologists.
What does the software produce?Output: Enhanced DICOM CT images returned through the compatible MDDS to PACS or another DICOM node; enhanced images coexist with the original reconstructed images; no user interface.
What is its most important limitation?Key limitation: Output interpretation is adjunctive to the original reconstructed image. Use with iterative reconstruction or deep-learning reconstructions may have unknown performance; advanced third-party post-processing or specialized non-standard filtering was not specifically characterized, and direct primary scanner reconstructions are recommended.
Clinical takeaway
Clinical role: Process and enhance CT images from any CT scanner manufacturer or model as an adjunct to the original reconstructed image.
Evidence takeaway
Evidence from the reviewed source record: FDA K254120 reports phantom performance validation using ACR, Digital CatPhan 600, and Mercury 4.0 phantoms plus a 270-case retrospective reader study. Phantom endpoints included detectability index, noise magnitude, NPS, CT-number accuracy, non-uniformity, and TTF, all reported as passing their stated criteria. Reader endpoints were diagnostic utility (average Likert score >=3) and artifact presence (average score >=2); no sensitivity/specificity or disease-detection accuracy was reported.
Clinical details
What this tool is for
Start with the authorized purpose, then verify how it fits your service line and reading workflow.
Exact purpose
Process and enhance CT images from any CT scanner manufacturer or model as an adjunct to the original reconstructed image.
Primary reconstructed DICOM CT images sent by a technologist through a compatible MDDS; FBP primary reconstructions at standard or higher dose were quantitatively validated. The packet also includes varied scanners, anatomies, contrast, reconstruction methods, and slice thicknesses, but warns that IR/DL reconstruction performance may be unknown.
Enhanced DICOM CT images returned through the compatible MDDS to PACS or another DICOM node; enhanced images coexist with the original reconstructed images; no user interface.
Output interpretation is adjunctive to the original reconstructed image. Use with iterative reconstruction or deep-learning reconstructions may have unknown performance; advanced third-party post-processing or specialized non-standard filtering was not specifically characterized, and direct primary scanner reconstructions are recommended.
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
Use these fields to structure a vendor demo, security review, and implementation estimate.
Integration
Technologist routes the study from the CT scanner to a compatible MDDS via DICOM; the MDDS supplies images to SubtleHD-CT and returns enhanced images to PACS or another DICOM node using an AE Title.
ACR AI Central provides an exact-submission model card for SubtleHD-CT (1.x); FDA labeling remains controlling for clinical use, limitations, and performance.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA K254120 reports phantom performance validation using ACR, Digital CatPhan 600, and Mercury 4.0 phantoms plus a 270-case retrospective reader study. Phantom endpoints included detectability index, noise magnitude, NPS, CT-number accuracy, non-uniformity, and TTF, all reported as passing their stated criteria. Reader endpoints were diagnostic utility (average Likert score >=3) and artifact presence (average score >=2); no sensitivity/specificity or disease-detection accuracy was reported.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
SubtleHD-CT (1.x) regulatory performance study summary
DesignReader Study
PopulationAdult and Pediatric
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K254120
IndependenceIndependence not established
K254120
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
These are common procurement questions; unknown values remain visible until a source supports them.
Pricing and total cost
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-08
No public price or quote for the exact K254120 configuration was established in this review.
Reimbursement and coding
Not established in reviewed sources
Reviewed sources checked
Exact FDA submission
Checked 2026-09-08
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Safety and lifecycle
Postmarket record
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K254120. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
Verify deployment and data flow: The reviewed sources do not establish the current exact-release hosting topology, data flow, region, tenancy, or disaster-recovery configuration.
Verify security and privacy controls: Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.
Verify training: Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Verify pricing and total cost: No public price or quote for the exact K254120 configuration was established in this review.
Verify reimbursement and coding relevance: No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Verify local validation and lifecycle monitoring: Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
23Fields reviewed
9Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, acr ai central product, fda decision summary, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Sources & history
How this profile was documented
Open the ledger for source dates, scope, and research-record updates.
View source ledger and history7 sources - 3 history items
AI-CVD provides automated quantitative CT-derived anatomical and density measurements for clinician review, including calcium scores, cardiac and vascular measurements, lung and...
uOmnispace.CT provides healthcare institutions with tools to view, manipulate, evaluate, and analyze medical CT images, including colon, lung, vessel, brain, cardiac, liver, rib,...
Modality
CT/CTA
Anatomy
Colon, lungs and airway, vessels, brain, heart, liver, ribs, spine, and other CT regions
uOmnispace.CT is software for viewing, manipulating, evaluating, and analyzing medical images within healthcare institutions. The FDA indication lists colon visualization, dental...
Modality
CT/CTA
Anatomy
Colon, jaw, lungs, airways, brain, heart, vessels, liver, spine, ribs, and other labeled CT regions
Auto-Seg and Spine Auto-Seg accept spine and pelvis CT images and provide three-dimensional segmentation and labels to support spine-surgery planning and assessment by spine...