FDA-cleared profile

SIGNA Prime

GE Healthcare (Tianjin) Company Limited

Authorized purpose: SIGNA Prime is a whole-body 1.5T MR system for qualified technologists and interpreting physicians to acquire, reconstruct, display, and export labeled diagnostic MR examinations using supported coils, sequences, and applications.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: SIGNA Prime is a whole-body 1.5T MR system for qualified technologists and interpreting physicians to acquire, reconstruct, display, and export labeled diagnostic MR examinations using supported coils, sequences, and applications.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: SIGNA Prime is a whole-body 1.5T MR system for qualified technologists and interpreting physicians to acquire, reconstruct, display, and export labeled diagnostic MR examinations using supported coils, sequences, and applications.

Evidence takeaway

Evidence from the reviewed source record: The K211980 FDA decision summary describes the SIGNA Prime whole-body 1.5T MR scanner, its 60 cm bore and release-specific software, unchanged or product-name-updated indications, DICOM functions, and sample clinical-image evaluation supporting substantial equivalence. It does not establish a general clinical sensitivity or specificity claim.

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
SIGNA Prime is a whole-body 1.5T MR system for qualified technologists and interpreting physicians to acquire, reconstruct, display, and export labeled diagnostic MR examinations using supported coils, sequences, and applications.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Intended users
Qualified MRI technologists and physicians trained in magnetic resonance imaging.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Care setting and population
Hospital and outpatient MRI departments operating the cleared field strength, coils, protocols, and accessories.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Workflow role
MRI acquisition, reconstruction, review, and release-specific workflow support.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Required input
Patient MRI signal data acquired with supported coils, pulse sequences, field strengths, and release-specific protocols.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Output and human action
Cross-sectional MR images, spectra, derived measurements, and reconstructed series for professional review.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Limitations
Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to the exact clearance and labeling.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K211980
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
MRI
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Anatomy
Multi-region
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2022-01-16
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Scanner-console and DICOM workflow to PACS or downstream applications; validate worklist, identifiers, series labels, geometry, measurements, routing, and archive completeness.
  • Public source
  • Product family
  • Checked 2026-09-03

Current vendor materials are product-family evidence for K211980; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Deployment and data flow
Installed MRI hardware and release-specific scanner software with licensed coils, sequences, reconstruction options, and accessories.
  • Public source
  • Product family
  • Checked 2026-09-03

Current vendor materials are product-family evidence for K211980; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-03

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified mri technologists and physicians trained in magnetic resonance imaging should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to the exact clearance and labeling.
  • Public source
  • Product family
  • Checked 2026-09-03

Current vendor materials are product-family evidence for K211980; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled AI functions to K211980 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

Scanner-console and DICOM workflow to PACS or downstream applications; validate worklist, identifiers, series labels, geometry, measurements, routing, and archive completeness.

product family · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to the exact clearance and labeling.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

No exact-product price was found in the reviewed public material; vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Field strength, coil, sequence, anatomy, population, implant, intervention, and optional-feature limits vary by model and release; map the installed configuration to the exact clearance and labeling.

product family · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
The K211980 FDA decision summary describes the SIGNA Prime whole-body 1.5T MR scanner, its 60 cm bore and release-specific software, unchanged or product-name-updated indications, DICOM functions, and sample clinical-image evaluation supporting substantial equivalence. It does not establish a general clinical sensitivity or specificity claim.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-03
Reported sensitivity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Sensitivity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
  • Not applicable
  • Not applicable
  • Checked 2026-09-03

Specificity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K211980 FDA performance package for SIGNA Prime

  • DesignManufacturer MR hardware, software, image-quality, sample-clinical-image, safety, EMC, usability, and DICOM verification and validation submitted in K211980
  • PopulationWhole-body diagnostic MR examinations across the cleared field strength, coils, sequences, anatomies, and supported patient workflows
  • Scopeexact submission
  • Tested versionSIGNA Prime (K211980)
  • IndependenceSponsor or vendor study

K211980

Manufacturer evidence submitted for FDA clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
  • Public source
  • Not applicable
  • Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
2 exact-submission FDA recall records in the 2026-09-02 openFDA snapshot identify K211980. Affected model, version, unit, correction, and current manufacturer instructions still require local confirmation. FDA recall Z-2479-2024 is Open, Classified for SIGNA Prime, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System. FDA reason: GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant. FDA action: GE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION to its consignees via letter on 06/14/2024. The letter explained the problem with the device, safety issue, and requested the following: "Actions to be taken by Customer/User: You can continue to use your MR system by following the instructions below: To ensure that the displayed B1+RMS value on the Scan Prescription User Interface does not exceed the user prescribed limit specified in the Low SAR Mode User Interface: " If your software version includes the Optimized T2 FLAIR Sequence option, enable this option by setting the value to 1.00. This option is present on the Advanced Tab of the Scan Prescription User Interface. " If your software version does not include the Optimized T2 FLAIR Sequence option, adjust scan parameters such as TR , # Slices or Auto Refocus Flip Angle until the displayed B1+RMS value on the Scan Prescription User Interface does not exceed the user prescribed limit. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions." GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. FDA recall Z-0983-2022 is Open, Classified for GE Healthcare MR Systems, Magnetic Resonance imaging (MRI). SIGNA Premier, SIGNA Architect, SIGNA Pioneer, Discovery MR750w 3.0T, Discovery MR750 3.0T, SIGNA PET/MR, 1.5T SIGNA Creator, 1.5T SIGNA Explorer, 1.5T SIGNA MR355, 1.5T SIGNA MR360, 1.49T SIGNA MR380, SIGNA Voyager, SIGNA Voyager Premier Edition, SIGNA Artist, Optima MR450w 1.5T, Optima MR450w GEM, Discovery MR450 1.5T, GE 1.5T and 3.0T SIGNA HDx, GE 1.5T and 3.0T SIGNA HDxt, 1.5T and 3.0T SIGNA HDxt Mobile, SIGNA HDi, SIGNA Vibrant, SIGNA 1.5T TwinSpeed, SIGNA Infinity Excite, SIGNA Infinity with Excite Technology, SIGNA OpenSpeed, SIGNA HFO, SIGNA HFO/I MR System, GE 0.7T SIGNA OpenSpeed, SIGNA OpenSpeed with Excite, SIGNA HDe, Brivo MR355, Optima MR360, SIGNA 3.0T MR System, SIGNA Excite 3T, 1.5T SIGNA Excite HD, 3.0T SIGNA Excite HD, SIGNA Contour, SIGNA Horizon Cx, SIGNA Horizon Lx, SIGNA Infinity, SIGNA MR/I, SmartSpeed, EchoSpeed Plus, HighSpeed Plus, MR Vectra, MR MAX, Optima MR430s, Optima MR430s 1.5T, Optima MR430s MRI Scanner, 1.0T MSK Extreme, SIGNA Profile EXCITE, SIGNA Ovation, 0.35T SIGNA Ovation with Excite, SIGNA Prime.. FDA reason: During the de-installation of an MR system magnet, if all rigging and hardware that attach to the magnet for transportation are not properly installed and secured, the magnet may fall, leading to potential injury. FDA action: The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters to U.S. consignees dated 3/4/2022 via FedEx. Letters to OUS consignees began on 3/10/2022 via FedEx. The letters explained the issue, the affected product (all GE Healthcare MR Systems), and informed the consignee the device can continue to be used. Actions to be taken by the customer/user are: 1) If you are planning to de-install your GE Healthcare MR System, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative prior to any activities so that GE Healthcare can provide you with guidance for de-installation. 2) Complete and return the attached response form to Recall.60985@ge.com A response form was enclosed for completion and return to GE via email or scanning the...
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-04

Each published event was manually matched to the exact FDA submission and affected product or version; a recall does not establish an AI-algorithm defect or current site exposure unless the event states that scope.

recall: Z-2479-2024

Open, Classified

2024-08-02

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; the FDA record does not identify an AI-algorithm defect unless it says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-0983-2022

Open, Classified

2022-04-21

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; the FDA record does not identify an AI-algorithm defect unless it says so. Confirm affected model, version, units, corrective action, and completion status locally.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
26Fields reviewed
13Source classes checked
0Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains no PubMed or ClinicalTrials.gov candidates; no external evidence was promoted. The current discovery snapshot contains 0 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. 2 FDA recall candidates were manually reviewed: 2 published and 0 rejected as a neighboring or otherwise inapplicable version. Native FDA recall identifiers produced 2 postmarket candidate records; 2 have been reviewed (2 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA 510(k) decision summary K211980: SIGNA Prime

    U.S. Food and Drug Administration - Published 2026-09-03 - Accessed 2026-09-03

    Scope: purpose, modality, anatomy, intendedUsers, input, output, limitations, evidenceSummary

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. FDA recall Z-2479-2024: SIGNA Prime, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System

    U.S. Food and Drug Administration - Published 2024-08-02 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  4. FDA recall Z-0983-2022: GE Healthcare MR Systems, Magnetic Resonance imaging (MRI). SIGNA Premier, SIGNA Architect, SIGNA Pioneer, Discovery MR750w 3.0T, Discovery MR750 3.0T, SIGNA PET/MR, 1.5T SIGNA Creator, 1.5T SIGNA Explorer, 1.5T SIGNA MR355, 1.5T SIGNA MR360, 1.49T SIGNA MR380, SIGNA Voyager, SIGNA Voyager Premier Edition, SIGNA Artist, Optima MR450w 1.5T, Optima MR450w GEM, Discovery MR450 1.5T, GE 1.5T and 3.0T SIGNA HDx, GE 1.5T and 3.0T SIGNA HDxt, 1.5T and 3.0T SIGNA HDxt Mobile, SIGNA HDi, SIGNA Vibrant, SIGNA 1.5T TwinSpeed, SIGNA Infinity Excite, SIGNA Infinity with Excite Technology, SIGNA OpenSpeed, SIGNA HFO, SIGNA HFO/I MR System, GE 0.7T SIGNA OpenSpeed, SIGNA OpenSpeed with Excite, SIGNA HDe, Brivo MR355, Optima MR360, SIGNA 3.0T MR System, SIGNA Excite 3T, 1.5T SIGNA Excite HD, 3.0T SIGNA Excite HD, SIGNA Contour, SIGNA Horizon Cx, SIGNA Horizon Lx, SIGNA Infinity, SIGNA MR/I, SmartSpeed, EchoSpeed Plus, HighSpeed Plus, MR Vectra, MR MAX, Optima MR430s, Optima MR430s 1.5T, Optima MR430s MRI Scanner, 1.0T MSK Extreme, SIGNA Profile EXCITE, SIGNA Ovation, 0.35T SIGNA Ovation with Excite, SIGNA Prime.

    U.S. Food and Drug Administration - Published 2022-04-21 - Accessed 2026-09-04

    Scope: postmarketSafety, affectedProduct, recallStatus, correctiveAction

  5. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  6. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  7. Medicare Coverage Database

    Centers for Medicare & Medicaid Services - Accessed 2026-09-01

    Scope: coverage, codingContext

  8. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  9. GE SIGNA Prime MRI

    GE HealthCare - Accessed 2026-09-03

    Scope: productFamilyContext, workflowRole, implementationCapabilities

Research history

  1. K211980 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    human reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K260746

S-scan Open (100001800)

Esaote, S.p.A.

S-scan Open is an Esaote MR system for diagnostic image acquisition, reconstruction, display, and professional interpretation within the labeled scanner, coil, sequence, and...

Modality
MRI
Anatomy
Multi-region
Decision
2026-03-27
FDA-cleared K252379

AIR Recon DL

Ge Medical Systems, LLC

AIR Recon DL adds a deep-learning phase-correction and reconstruction option to supported GE 1.5T, 3T, and 7T MR workflows for all labeled anatomies and ages to reduce noise or...

Modality
MRI
Anatomy
Multi-region
Decision
2025-12-23