FDA-cleared profile

RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A, RayStation 2024A SP3, RayPlan 2024A SP3

RaySearch Laboratories AB (PUBL)

RayStation and RayPlan releases covered by K240398 are configurable radiotherapy and medical-oncology software that proposes or supports treatment plans from user inputs, with authorized users required to review and approve plans before treatment.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
RayStation and RayPlan releases covered by K240398 are configurable radiotherapy and medical-oncology software that proposes or supports treatment plans from user inputs, with authorized users required to review and approve plans before treatment.
FDA source Exact FDA submission Checked 2026-09-03
Intended users
Qualified radiation oncologists, medical physicists, dosimetrists, and radiation therapists.
FDA source Exact FDA submission Checked 2026-09-03
Care setting and population
Radiation-oncology departments using compatible planning, imaging, and treatment-delivery systems.
FDA source Exact FDA submission Checked 2026-09-03
Workflow role
Radiotherapy contouring, treatment planning, adaptation, or treatment-session management under professional review.
FDA source Exact FDA submission Checked 2026-09-03
Required input
Supported CT or MR images, structures, prescriptions, plans, registrations, and treatment-session data.
FDA source Exact FDA submission Checked 2026-09-03
Output and human action
Contours, treatment plans, adapted plans, dose or evaluation results, or session-management information requiring review and approval.
FDA source Exact FDA submission Checked 2026-09-03
Limitations
Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
FDA source Exact FDA submission Checked 2026-09-03

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K240398
FDA source Exact FDA submission Checked 2026-08-31
Modality
Radiotherapy Planning Imaging
FDA source Exact FDA submission Checked 2026-09-03
Anatomy
Multi-region
FDA source Exact FDA submission Checked 2026-09-03
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2025-04-04
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
DICOM RT and vendor workflow integration; validate patient identity, image geometry, structures, dose, plan transfer, approvals, version compatibility, and rollback.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K240398; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Deployment and data flow
Licensed software connected to compatible radiotherapy planning and delivery infrastructure.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K240398; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified radiation oncologists, medical physicists, dosimetrists, and radiation therapists should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K240398; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled AI functions to K240398 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

DICOM RT and vendor workflow integration; validate patient identity, image geometry, structures, dose, plan transfer, approvals, version compatibility, and rollback.

product family · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

No exact-product price was found in the reviewed public material; vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.

product family · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
K240398 reports software, planning, dose, optimization, interface, and verification and validation testing for the listed releases. A plan proposal is not an autonomous treatment decision; the applicable machine, technique, image, structure, dose, and approval configuration must be checked before clinical use.
FDA source Exact FDA submission Checked 2026-09-03
Reported sensitivity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Sensitivity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Specificity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K240398 FDA performance package for RayStation 2023B, RayPlan 2023B, RayStation 2024A, RayPlan 2024A, RayStation 2024A SP3, RayPlan 2024A SP3

Manufacturer radiotherapy planning, dose, optimization, interface, and verification and validation · Radiotherapy planning and medical-oncology workflows using supported images, structures, prescriptions, and treatment-delivery interfaces

exact submission · Tested version: RayStation/RayPlan 2023B, 2024A, and 2024A SP3 · Sponsor or vendor study

K240398

Manufacturer evidence submitted for FDA clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS radiation-oncology material. Evaluate net planning or delivery value within the radiotherapy service line.
Public source Not applicable Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
2 exact-submission FDA recall records in the 2026-09-02 openFDA snapshot identify K240398. Affected model, version, unit, correction, and current manufacturer instructions still require local confirmation. FDA recall Z-1106-2026 is Open, Classified for Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 Software Version: RayStation 2023B, RayStation 2024A, RayStation 2024A SP3 Product Description: Radiation Therapy Treatment Planning System Component: No. FDA reason: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. FDA action: RaySearch Laboratories notified consignees via email on about 12/05/2025. Consignees were instructed to follow the outlined steps in the notification letter when adding geometry to an empty ROI with MO or type Bolus, Fixation or Support, educate planning staff and all users about the workaround, inspect units and identify all installed units with the affected software version numbers, and confirm receipt and understanding of the customer notification by responding to the notification email. The issue will be resolved in the next version of RayStation, scheduled for market release in December 2025 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. The upgrade to an unaffected software version will be scheduled in coordination with the authorized personnel, considering the specific circumstances and requirements of each individual case. The upgrade may be conducted remotely or onsite, as deemed appropriate. FDA recall Z-2495-2025 is Open, Classified for RayStation with the following product descriptions: 1. RayStation 2023B, Software Version: 14.0.0.3338. 2. RayStation 2024A, Software Version: 15.0.0.430. 3. RayStation 2024A SP3, Software Version: 15.1.3.10. Product Description: Radiation Therapy Treatment Planning System.. FDA reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. FDA action: RaySearch notified consignees on about 08/08/2025 via email. Consignees were informed that as temporary solution, affected objects can be re-exported via RayGateway to generate new UIDs. They were instructed to inform planning staff and all users about this workaround, inspect your product and identify all installed units with the above software version number(s), and complete and return the response form. The issue will be resolved in the next version of RayStation, scheduled for market release in December 2025 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.
FDA source Exact FDA submission Checked 2026-09-04

Each published event was manually matched to the exact FDA submission and affected product or version; a recall does not establish an AI-algorithm defect or current site exposure unless the event states that scope.

recall: Z-1106-2026

Open, Classified

2026-01-16

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-2495-2025

Open, Classified

2025-09-02

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; do not infer an AI-algorithm defect unless the FDA record says so. Confirm affected model, version, units, corrective action, and completion status locally.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
25Fields reviewed
11Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact FDA decision materials were reviewed for identity, indication, users, workflow, limitations, and submitted performance. Current vendor material is labeled product-family context where it does not establish the exact release. PubMed and ClinicalTrials.gov results are retained as candidate evidence with version applicability pending; CMS reimbursement context and exact-identifier openFDA postmarket results were also checked. Pricing, final security architecture, contracted services, enabled options, and local interoperability remain organization-specific evidence. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. 2 FDA recall candidates were manually reviewed: 2 published and 0 rejected as a neighboring or otherwise inapplicable version. Native FDA recall identifiers produced 2 postmarket candidate records; 2 have been reviewed (2 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-04.

Alternatives

Related tools to compare

FDA-cleared K242994

OncoStudio (OS-01)

OncoSoft. Co., Ltd.

OncoStudio OS-01 creates initial deep-learning organs-at-risk contours from CT images of the head and neck, thorax, abdomen, and pelvis for radiation-oncology review, editing, and...

Modality
Radiotherapy Planning Imaging
Anatomy
Multi-region
Decision
2025-02-24
FDA-cleared K232799

syngo.via RT Image Suite

Siemens Medical Solutions USA, Inc.

syngo.via RT Image Suite VB80 supports trained radiation-oncology professionals with multimodality visualization, registration, advanced contouring, treatment preparation,...

Modality
Radiotherapy Planning Imaging
Anatomy
Multi-region
Decision
2024-04-26