Authorized purpose: NeuroQuant automatically labels, visualizes, and volumetrically quantifies segmentable brain structures and lesions from MR images, optionally using T2 FLAIR and T2* GRE/SWI for lesion analysis; outputs require radiologist/neuroradiologist review and may be compared with reference percentiles.
Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.
Product overview
Start with the essentials
Authorized purpose: NeuroQuant automatically labels, visualizes, and volumetrically quantifies segmentable brain structures and lesions from MR images, optionally using T2 FLAIR and T2* GRE/SWI for lesion analysis; outputs require radiologist/neuroradiologist review and may be compared with reference percentiles.
Overview status: Source-reviewed overview
What does this software do?Authorized purpose: NeuroQuant automatically labels, visualizes, and volumetrically quantifies segmentable brain structures and lesions from MR images, optionally using T2 FLAIR and T2* GRE/SWI for lesion analysis; outputs require radiologist/neuroradiologist review and may be compared with reference percentiles.
Which images or clinical context are in scope?Modality: MRI Anatomy or clinical context: Brain Input: 3D T1-weighted MR images; optional T2 FLAIR and T2* GRE/SWI series; DICOM Storage input from an MRI scanner or PACS; specified protocols are recommended.
Who uses it?Intended users: Medical professionals; clinical outputs must be reviewed by a radiologist or neuroradiologist and may be forwarded to the referring neurologist.
What does the software produce?Output: Numerical brain and lesion volumes, grayscale intensity maps, color overlays, segmented image series, and morphometric reports in standard DICOM for display on commercial DICOM workstations/PACS; age/gender-matched normative percentiles and change analysis may be provided.
What is its most important limitation?Key limitation: Clinical output requires radiologist/neuroradiologist review; the device is prescription-only; specified acquisition protocols are recommended; and the output is not a substitute for full clinical interpretation.
Clinical takeaway
Clinical role: NeuroQuant automatically labels, visualizes, and volumetrically quantifies segmentable brain structures and lesions from MR images, optionally using T2 FLAIR and T2* GRE/SWI for lesion analysis; outputs require radiologist/neuroradiologist review and may be compared with reference percentiles.
Evidence takeaway
Evidence from the reviewed source record: FDA performance testing covered brain segmentation, FLAIR lesion segmentation, and MCH detection models; FLAIR mean Dice 0.70 (SD 0.14) and MCH median F1 0.60 were reported, with separate test sets from training.
Clinical details
What this tool is for
Start with the authorized purpose, then verify how it fits your service line and reading workflow.
Exact purpose
NeuroQuant automatically labels, visualizes, and volumetrically quantifies segmentable brain structures and lesions from MR images, optionally using T2 FLAIR and T2* GRE/SWI for lesion analysis; outputs require radiologist/neuroradiologist review and may be compared with reference percentiles.
3D T1-weighted MR images; optional T2 FLAIR and T2* GRE/SWI series; DICOM Storage input from an MRI scanner or PACS; specified protocols are recommended.
Numerical brain and lesion volumes, grayscale intensity maps, color overlays, segmented image series, and morphometric reports in standard DICOM for display on commercial DICOM workstations/PACS; age/gender-matched normative percentiles and change analysis may be provided.
Clinical output requires radiologist/neuroradiologist review; the device is prescription-only; specified acquisition protocols are recommended; and the output is not a substitute for full clinical interpretation.
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
Use these fields to structure a vendor demo, security review, and implementation estimate.
Integration
DICOM Storage receives MR series from an MRI scanner or PACS and returns DICOM image series and reports to DICOM workstations/PACS; transient data storage only is described.
The reviewed sources do not establish a scoped security audit, certification, or control set.
Training and support
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-07
Training, credentialing, and implementation support terms were not established.
Monitoring and change control
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-07
Current post-deployment monitoring and change-control practices were not established.
DICOM Storage and PACS routing
The exact submission describes DICOM Storage receipt from an MRI scanner or PACS and standard DICOM image-series/report output for commercial DICOM workstations.
The official NeuroQuant Installed Quick Start Guide describes 3D T1 DICOM input, AETitle/report selection, automated processing, and output to a configured PACS; the guide is a v3.0 family manual and is not treated as proof of the exact K241098 build.
product family · checked 2026-09-07
Family manual context; current exact-version compatibility requires confirmation.
CorTechs' current NeuroQuant page describes cloud-based brain segmentation, quantification, comparison, and adaptable reports that integrate with radiology workflows; this current family description is not used to override the exact FDA appliance deployment.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA performance testing covered brain segmentation, FLAIR lesion segmentation, and MCH detection models; FLAIR mean Dice 0.70 (SD 0.14) and MCH median F1 0.60 were reported, with separate test sets from training.
NeuroQuant model validation across brain, FLAIR, and MCH pipelines
DesignRetrospective model validation with separate training and test datasets
PopulationBrain segmentation: 1,473 training 3D T1 series from more than 16 institutions and 30-patient test set; FLAIR: 340 training series from 22 institutions and 63-patient test set; MCH: 463 training 2D T2*GRE/SWI series from more than 68 institutions and 117-patient test set.
Scopeexact submission
Tested versionNeuroQuant submitted in K241098; build not stated
IndependenceIndependence not established
K241098
The FDA establishes separation of training and test data, not independent investigators or sponsorship. The reported FLAIR and MCH values are model-specific and should not be generalized to every NeuroQuant output.
These are common procurement questions; unknown values remain visible until a source supports them.
Pricing and total cost
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-07
No public amount, range, currency, geography, or contract term was established.
Reimbursement and coding
Aetna Clinical Policy Bulletin 0739 classifies automated quantification of brain MRI as experimental, investigational, or unproven and lists Category III CPT codes 0865T and 0866T as not covered for the indications in the bulletin. This is service-level Aetna policy, not an exact NeuroQuant eligibility or universal coverage determination.
Not covered for indications listed in Aetna CPB 0739; automated quantification of brain MRI is classified as experimental, investigational, or unproven
Aetna · United States; plan and benefit terms vary
Policy last reviewed 2026-07-24. This service-level policy does not prove eligibility for a particular NeuroQuant version, member, indication, site, or claim.
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
Not established in reviewed sources
Reviewed sources checked
Exact FDA submission
Checked 2026-09-04
No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.
Buyer worksheet
Open questions to take to the vendor
Verify security and privacy controls: The reviewed sources do not establish a scoped security audit, certification, or control set.
Verify training: Training, credentialing, and implementation support terms were not established.
Verify pricing and total cost: No public amount, range, currency, geography, or contract term was established.
Verify local validation and lifecycle monitoring: Current post-deployment monitoring and change-control practices were not established.
Verify recalls and postmarket safety signals: No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.
This audit distinguishes completed source review from fields that have not yet been researched.
Human reviewedStatus
2026-09-07Last searched
24Fields reviewed
12Source classes checked
20Screened PubMed leads
5Screened trial leads
0Unreviewed FDA recall leads
The stored PubMed, ClinicalTrials.gov, and exact-submission recall metadata candidates were screened by identifier and count. The records were not treated as a full paper, trial, or postmarket review. The current discovery snapshot contains 20 PubMed and 5 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, vendor product page, vendor manual, payer policy, literature index, trial registry, fda device recall, pubmed, clinical trials, openfda device recall
Sources & history
How this profile was documented
Open the ledger for source dates, scope, and research-record updates.
View source ledger and history9 sources - 3 history items
ANDI 2.2 processes brain MR images for atlas-assisted visualization, segmentation and volumetric quantification, presenting adjunctive DICOM-readable information for trained...
Morph automatically labels, visualizes, and volumetrically quantifies segmentable brain structures from MR images and may compare measurements with reference percentile data; it...
Brain MRI morphometry and white matter hyperintensity quantification, including normative comparison and longitudinal change assessment for clinician review.
Adjunctive quantitative brain MRI analysis that compares patient measures with normative and longitudinal references; a physician reviews all results and makes the clinical...