- Integration
- DICOM CT workflow with advanced image objects; validate PACS, RIS, worklist, quantitative consistency, reconstruction, and downstream software support.
Public source Exact FDA submission Checked 2026-09-03
- Deployment and data flow
- Installed photon-counting CT system with syngo CT VB10 and corresponding detector, reconstruction, siting, and service infrastructure.
Public source Exact FDA submission Checked 2026-09-03
- Security and privacy
- Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03
Public materials reviewed for this batch do not establish the complete security posture for the exact release. Request data-flow, identity, audit, encryption, retention, support-access, vulnerability-management, backup, and incident-response evidence.
- Training and support
- Only qualified ct technologists and physicians trained in diagnostic and interventional ct workflows. should operate or interpret the system. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
- Monitoring and change control
- Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track the limits that matter for this product: K233657 explicitly does not support radiation-therapy planning, unlike K243523. The scanner is not the principal guidance method for invasive procedures, and product-family literature does not establish every exact-version outcome.
Public source Exact FDA submission Checked 2026-09-03
Release and scope verification
Map the installed model, build, options, population, anatomy, inputs, and outputs to K233657 before acceptance.
exact submission · checked 2026-09-03 Later product-family features are not evidence that this cleared release includes them.
Interoperability acceptance
DICOM CT workflow with advanced image objects; validate PACS, RIS, worklist, quantitative consistency, reconstruction, and downstream software support.
exact submission · checked 2026-09-03 Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.
Clinical acceptance
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. K233657 explicitly does not support radiation-therapy planning, unlike K243523. The scanner is not the principal guidance method for invasive procedures, and product-family literature does not establish every exact-version outcome.
exact submission · checked 2026-09-03 Define stop-use, escalation, rollback, and revalidation triggers before production use.
Economic evaluation
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-03 Vendor productivity claims are scenario inputs, not guaranteed local savings.
Lifecycle monitoring
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track the limits that matter for this product: K233657 explicitly does not support radiation-therapy planning, unlike K243523. The scanner is not the principal guidance method for invasive procedures, and product-family literature does not establish every exact-version outcome.
exact submission · checked 2026-09-03 Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.