- Integration
- Scanner-console and DICOM workflow to PACS or downstream applications; validate worklist, identifiers, series labels, geometry, measurements, routing, and archive completeness.
Public source Exact FDA submission Checked 2026-09-03
- Deployment and data flow
- Installed MRI hardware and release-specific scanner software with licensed coils, sequences, and options.
Public source Exact FDA submission Checked 2026-09-03
- Security and privacy
- Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03
Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.
- Training and support
- Only qualified mri technologists and physicians trained in magnetic resonance imaging. should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
- Monitoring and change control
- Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: The clearance does not establish broad whole-body proton MRI. Apply the exact nucleus, anatomy, coil, sequence, and accessory restrictions for each examination.
Public source Exact FDA submission Checked 2026-09-03
Release and scope verification
Map the installed model, build, options, population, anatomy, inputs, and outputs to K232322 before acceptance.
exact submission · checked 2026-09-03 Later product-family features are not evidence that this cleared release includes them.
Interoperability acceptance
Scanner-console and DICOM workflow to PACS or downstream applications; validate worklist, identifiers, series labels, geometry, measurements, routing, and archive completeness.
exact submission · checked 2026-09-03 Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.
Clinical acceptance
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. The clearance does not establish broad whole-body proton MRI. Apply the exact nucleus, anatomy, coil, sequence, and accessory restrictions for each examination.
exact submission · checked 2026-09-03 Define stop-use, escalation, rollback, and revalidation triggers before production use.
Economic evaluation
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-03 Vendor productivity claims are scenario inputs, not guaranteed local savings.
Lifecycle monitoring
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: The clearance does not establish broad whole-body proton MRI. Apply the exact nucleus, anatomy, coil, sequence, and accessory restrictions for each examination.
exact submission · checked 2026-09-03 Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.