Authorized purpose: Provide quantitative support for physician diagnosis, follow-up examination, and documentation of pulmonary-tissue images or abnormalities from CT thoracic datasets.
Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.
Product overview
Start with the essentials
Authorized purpose: Provide quantitative support for physician diagnosis, follow-up examination, and documentation of pulmonary-tissue images or abnormalities from CT thoracic datasets.
Overview status: Source-reviewed overview
What does this software do?Authorized purpose: Provide quantitative support for physician diagnosis, follow-up examination, and documentation of pulmonary-tissue images or abnormalities from CT thoracic datasets.
Which images or clinical context are in scope?Modality: CT/CTA Anatomy or clinical context: Thorax and lungs, including left and right lungs, lobes, pulmonary tissue, and fissures. Input: DICOM CT thoracic datasets that may contain pulmonary abnormalities.
Who uses it?Intended users: Physicians; the FDA packet does not further specify user credentials.
What does the software produce?Output: Three-dimensional segmentation and isolation of the left and right lungs and lobes; lung/lobar volume, density-range and 15th-percentile measurements; fissure evaluation; and a report.
What is its most important limitation?Key limitation: The software is command-line and has no graphical user interface; the comparison table states that users cannot manually edit the segmentation. The reported equivalence criteria were volume difference <=10%, LAA-950HU agreement limits -1% to 1%, LAA-910HU limits -10% to 10%, and 15th-percentile limits -10 HU to 10 HU. The packet does not establish a disease-classification sensitivity or specificity endpoint.
Clinical takeaway
Clinical role: Provide quantitative support for physician diagnosis, follow-up examination, and documentation of pulmonary-tissue images or abnormalities from CT thoracic datasets.
Evidence takeaway
Evidence from the reviewed source record: FDA K173821 compared LungQ 1.1.0 with VIDA PW2 on 250 randomly selected COPDGene CT scans spanning subjects with and without COPD and multiple disease stages. Lung/lobar volume and density outputs were assessed against prespecified equivalence limits. A separate fissure study used 55 scans from a five-site multicenter study in Germany, the Netherlands, and Sweden and compared voxel-wise fissure completeness with a manual reference; mean Az was 0.95 for LungQ 1.1.0 versus 0.76 for the previously published version (p<0.001). Az is an AUC endpoint, not sensitivity or specificity.
Clinical details
What this tool is for
Start with the authorized purpose, then verify how it fits your service line and reading workflow.
Exact purpose
Provide quantitative support for physician diagnosis, follow-up examination, and documentation of pulmonary-tissue images or abnormalities from CT thoracic datasets.
Command-line post-processing and quantitative-analysis aid; it supplies segmentation, volume/density/fissure measurements, and reports and has no graphical user interface.
Three-dimensional segmentation and isolation of the left and right lungs and lobes; lung/lobar volume, density-range and 15th-percentile measurements; fissure evaluation; and a report.
The software is command-line and has no graphical user interface; the comparison table states that users cannot manually edit the segmentation. The reported equivalence criteria were volume difference <=10%, LAA-950HU agreement limits -1% to 1%, LAA-910HU limits -10% to 10%, and 15th-percentile limits -10 HU to 10 HU. The packet does not establish a disease-classification sensitivity or specificity endpoint.
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
Use these fields to structure a vendor demo, security review, and implementation estimate.
Integration
DICOM loading and command-line/Image Analysis Service processing are described; the packet does not name a PACS, RIS, EHR, routing interface, or current deployment topology.
Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.
ACR AI Central provides an exact-submission model card for LungQ Software; FDA labeling remains controlling for clinical use, limitations, and performance.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.
Evidence summary
FDA K173821 compared LungQ 1.1.0 with VIDA PW2 on 250 randomly selected COPDGene CT scans spanning subjects with and without COPD and multiple disease stages. Lung/lobar volume and density outputs were assessed against prespecified equivalence limits. A separate fissure study used 55 scans from a five-site multicenter study in Germany, the Netherlands, and Sweden and compared voxel-wise fissure completeness with a manual reference; mean Az was 0.95 for LungQ 1.1.0 versus 0.76 for the previously published version (p<0.001). Az is an AUC endpoint, not sensitivity or specificity.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
LungQ regulatory performance study summary
DesignStand-Alone Performance
PopulationPopulation described in the exact-submission ACR model card
Scopeexact submission
Tested versionExact product version not reported; regulatory study summarized for FDA submission K173821
Samplen=250
IndependenceIndependence not established
K173821
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
These are common procurement questions; unknown values remain visible until a source supports them.
Pricing and total cost
Vendor documentation pending
Vendor confirmation
Exact FDA submission
Checked 2026-09-09
No public price or quote for the exact K173821 configuration was established in this review.
Reimbursement and coding
Not established in reviewed sources
Reviewed sources checked
Exact FDA submission
Checked 2026-09-09
No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Safety and lifecycle
Postmarket record
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K173821. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
Verify security and privacy controls: Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.
Verify training: Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.
Verify pricing and total cost: No public price or quote for the exact K173821 configuration was established in this review.
Verify reimbursement and coding relevance: No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.
Verify local validation and lifecycle monitoring: Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-20Last searched
23Fields reviewed
9Source classes checked
7Screened PubMed leads
0Screened trial leads
0Unreviewed FDA recall leads
The refreshed discovery snapshot contains 7 PubMed and 0 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 7 PubMed and 0 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.
Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, pubmed, clinical trials, openfda device recall
Sources & history
How this profile was documented
Open the ledger for source dates, scope, and research-record updates.
View source ledger and history7 sources - 3 history items
Auto-Seg and Spine Auto-Seg accept spine and pelvis CT images and provide three-dimensional segmentation and labels to support spine-surgery planning and assessment by spine...
Acorn 3D Software and Acorn 3DP Model process CT, synthetic CT, and CT angiography for segmentation, visualization, measurement, treatment planning, and three-dimensional...
Modality
CT/CTA
Anatomy
Spine; craniomaxillofacial; musculoskeletal anatomy; and patient-specific printed models