FDA-cleared profile

icobrain aria

icometrix NV

icobrain aria is a computer-assisted detection (CADe) and diagnosis (CADx) software device to be used as a concurrent reading aid to help trained radiologists in the detection, assessment and characterization of Amyloid Related Imaging Abnormalities (ARIA) from a set of brain MR images. The software provides information about the presence, location, size, severity and changes of ARIA-E (brain edema or sulcal effusions) and ARIA-H (hemosiderin deposition, including microhemorrhage and superficial siderosis). Patient management decisions should not be made solely on the basis of analysis by icobrain aria.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
icobrain aria is a computer-assisted detection (CADe) and diagnosis (CADx) software device to be used as a concurrent reading aid to help trained radiologists in the detection, assessment and characterization of Amyloid Related Imaging Abnormalities (ARIA) from a set of brain MR images. The software provides information about the presence, location, size, severity and changes of ARIA-E (brain edema or sulcal effusions) and ARIA-H (hemosiderin deposition, including microhemorrhage and superficial siderosis). Patient management decisions should not be made solely on the basis of analysis by icobrain aria.
Public source Exact FDA submission Checked 2026-09-01

Exact-submission indication text from the ACR AI Central model card. FDA labeling remains controlling.

Intended users
Radiologist
Public source Exact FDA submission Checked 2026-09-01

Intended-user text reported in the exact-submission ACR model card.

Care setting and population
Adult
Public source Exact FDA submission Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Detection, localization, or diagnostic support
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Required input
MR images or data. Inclusion criteria: All input must be brain MRI scans (FLAIR and/or T2* GRE for volumetric analysis) with minimal artifacts, proper cranial coverage, appropriate slice thickness (~4–5 mm), and correct DICOM formatting. This ensures reliable segmentation and volumetric assessments.
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Output and human action
The software outputs are annotated DICOM images and an eletronic report summarizing the automated finding. These outputs are reviewed by radiologists concurrently alongside the raw input images in an image review software. Finally, the radiologist summarizes all findings (according to his/her best interpretation after reviewing all images and software outputs) in a structured report, which is then forwarded to the PACS.
Public source Exact FDA submission Checked 2026-09-01

Clinical output reported in the exact-submission ACR model card.

Limitations
Exclusion criteria: Scans with severe artifacts, incomplete brain coverage, corrupted DICOM data, or missing key sequences (such as T1 or FLAIR) are excluded. Excessively thick slices or major deviations from the recommended acquisition protocols also render data unusable.
Public source Exact FDA submission Checked 2026-09-01

Limitations and supported-acquisition details reported in the exact-submission ACR model card.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K240712
FDA source Exact FDA submission Checked 2026-08-31
Modality
MRI
Public source Exact FDA submission Checked 2026-09-01

Imaging modality reported in the exact-submission ACR AI Central model card.

Anatomy
Brain
Public source Exact FDA submission Checked 2026-09-01

Anatomy normalized from explicit exact-submission ACR model-card wording.

Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2024-11-07
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact-submission ACR model card explicitly references DICOM and PACS in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
Public source Exact FDA submission Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Deployment and data flow
On-premise
Public source Exact FDA submission Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA authorization does not establish the current commercial security posture. Request data-flow diagrams, hosting and retention details, access controls, audit logging, attestations, and incident-response documentation.

Training and support
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Training, onboarding, competency, and support commitments are commercial implementation details that require current vendor documentation.

Monitoring and change control
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Local acceptance testing, production monitoring, drift review, update control, and incident escalation require a current vendor and site-specific governance plan.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Result destination

PACS is explicitly referenced as a result destination.

exact submission · checked 2026-09-01

Confirm supported products, versions, routing behavior, and display requirements during local evaluation.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
Study type: Reader Study. Reference standard: null. Reader-study cases: 199. Readers: 16. Reader-study sensitivity: ARIA-E 70.9% (unassisted) to 86.5% (assisted) ARIA-H detection 68.7% to 79.0% ARIA-H microhemorrhages detection 69.3% to 79.6% ARIA-H superficial siderosis detection 49.7% to 59.9%. Reader-study specificity: specificity remained above 80% for the detection of all ARIA types, except for a specificity of 76.7% for the detection of (mild) ARIA-H microhemorrhages
Public source Exact FDA submission Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Reported sensitivity
70.9%; 70.2%
FDA source Exact FDA submission Checked 2026-08-13

Values reported in the exact-submission materials. Consult the linked source for endpoint, threshold, population, and confidence interval before comparison.

Reported specificity
80%; 76.7%; 79.0%
FDA source Exact FDA submission Checked 2026-08-13

Values reported in the exact-submission materials. Consult the linked source for endpoint, threshold, population, and confidence interval before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

icobrain aria regulatory performance study summary

Reader Study · Adult

exact submission · Tested version: 1 · Independence not established

K240712

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA submission materials do not establish current commercial pricing. Request the pricing unit, term, minimum commitment, implementation fees, and included support from the vendor.

Reimbursement and coding
Research pending
Needs research Exact FDA submission Checked 2026-08-31

Determine whether any product-specific CPT/HCPCS pathway exists and separate it from billing for the underlying imaging service.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-04Last searched
20Fields reviewed
6Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-02.

Alternatives

Related tools to compare

FDA-cleared K261405

ANDI 2.2

Imeka Solutions, Inc.

ANDI 2.2 processes brain MR images for atlas-assisted visualization, segmentation and volumetric quantification, presenting adjunctive DICOM-readable information for trained...

Modality
MRI
Anatomy
Brain
Decision
2026-05-28
FDA-cleared K260032

Morph

BeauBrain Healthcare, Inc.

Morph automatically labels, visualizes, and volumetrically quantifies segmentable brain structures from MR images and may compare measurements with reference percentile data; it...

Modality
MRI
Anatomy
Brain
Decision
2026-05-13
FDA-cleared K253057

AI-Rad Companion Brain MR

Siemens Healthcare GmbH

Brain MRI morphometry and white matter hyperintensity quantification, including normative comparison and longitudinal change assessment for clinician review.

Modality
MRI
Anatomy
Brain
Decision
2026-01-22
FDA-cleared K253489

Swoop® Portable MR Imaging® System

Hyperfine, Inc.

Swoop Portable MR Imaging System provides ultra-low-field point-of-care head MRI for all ages and includes deep-learning T1, T2, FLAIR, and DWI reconstruction.

Modality
MRI
Anatomy
Brain
Decision
2025-12-12