FDA-cleared profile

EPIQ Series Diagnostic Ultrasound System

Philips Ultrasound LLC

The K231190 EPIQ submission adds the mL26-8 linear-array transducer for high-resolution superficial imaging, including labeled musculoskeletal, ophthalmic, pediatric, peripheral-vessel, breast, thyroid, and testicular applications, within the EPIQ diagnostic ultrasound system workflow.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
The K231190 EPIQ submission adds the mL26-8 linear-array transducer for high-resolution superficial imaging, including labeled musculoskeletal, ophthalmic, pediatric, peripheral-vessel, breast, thyroid, and testicular applications, within the EPIQ diagnostic ultrasound system workflow.
FDA source Exact FDA submission Checked 2026-09-03
Intended users
Qualified physicians, sonographers, and other trained ultrasound professionals.
FDA source Exact FDA submission Checked 2026-09-03
Care setting and population
Hospitals, clinics, offices, operating rooms, and point-of-care environments within the cleared applications.
FDA source Exact FDA submission Checked 2026-09-03
Workflow role
Ultrasound acquisition, measurement, analysis, reporting, and release-specific AI assistance.
FDA source Exact FDA submission Checked 2026-09-03
Required input
Live or imported ultrasound data from supported probes, systems, host devices, and applications.
FDA source Exact FDA submission Checked 2026-09-03
Output and human action
Ultrasound images, measurements, calculations, overlays, and reports for clinician review.
FDA source Exact FDA submission Checked 2026-09-03
Limitations
AI-assisted measurements, views, and guidance do not replace image-quality assessment or clinician interpretation; supported probes, host devices, populations, and features differ by release.
FDA source Exact FDA submission Checked 2026-09-03

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K231190
FDA source Exact FDA submission Checked 2026-08-31
Modality
Ultrasound
FDA source Exact FDA submission Checked 2026-09-03
Anatomy
Multi-region
FDA source Exact FDA submission Checked 2026-09-03
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2023-05-12
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Validate supported probes and host devices, patient matching, DICOM or local export, PACS measurements, worklist, report transfer, and offline or downtime behavior.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K231190; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Deployment and data flow
Ultrasound hardware or software with supported probes, host devices, licensed options, and network configuration.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K231190; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified physicians, sonographers, and other trained ultrasound professionals should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: AI-assisted measurements, views, and guidance do not replace image-quality assessment or clinician interpretation; supported probes, host devices, populations, and features differ by release.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K231190; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled AI functions to K231190 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

Validate supported probes and host devices, patient matching, DICOM or local export, PACS measurements, worklist, report transfer, and offline or downtime behavior.

product family · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. AI-assisted measurements, views, and guidance do not replace image-quality assessment or clinician interpretation; supported probes, host devices, populations, and features differ by release.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

No exact-product price was found in the reviewed public material; vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: AI-assisted measurements, views, and guidance do not replace image-quality assessment or clinician interpretation; supported probes, host devices, populations, and features differ by release.

product family · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
The K231190 FDA decision summary documents the mL26-8 transducer addition, its frequency range and superficial high-resolution applications, and the unchanged EPIQ diagnostic-ultrasound indications. The FDA package does not establish a general AI diagnostic sensitivity or specificity claim.
FDA source Exact FDA submission Checked 2026-09-03
Reported sensitivity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Sensitivity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Specificity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K231190 FDA performance package for EPIQ Series Diagnostic Ultrasound System

Manufacturer transducer, acoustic, image-quality, software, electrical-safety, EMC, and usability verification and validation submitted in K231190 · EPIQ examinations using the mL26-8 linear-array transducer for labeled superficial, ophthalmic, pediatric, vascular, and small-organ applications

exact submission · Tested version: EPIQ Ultrasound Diagnostic System with mL26-8 transducer (K231190) · Sponsor or vendor study

K231190

Manufacturer evidence submitted for FDA clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
Public source Not applicable Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
4 exact-submission FDA recall records in the 2026-09-02 openFDA snapshot identify K231190. Affected model, version, unit, correction, and current manufacturer instructions still require local confirmation. FDA recall Z-2386-2025 is Open, Classified for mL26-8 Transducer Probe. FDA reason: To provide clarification and labelling to define the useful life of ultrasound transducers in the field. FDA action: An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377. FDA recall Z-0283-2024 is Completed for Philips EPIQ Diagnostic Ultrasound System Models: EPIQ 7G(model # 795200), EPIQ 7C (Model # 795201), EPIQ 7W (Model # 795202), EPIQ 5G (model # 795204), EPIQ 5W (model # 795206), EPIQ CVx (Model # 795231), EPIQ Elite (model # 795234). FDA reason: Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image. FDA action: On October 9, 2023, Philips issued a Urgent Medical Device Correction notification to affected consignees. Philips asked consignees to take the following actions: 1. Identify if your firm has an affected product. 2. In the 2D tab in the system touchscreen, disable the AutoSCAN feature every time a transducer or a Transducer preset is selected to prevent the system from potentially experiencing image lag issue. The AutoSCAN feature is disabled when the button is not highlighted in amber. 3. Please complete and return the attached response form to Philips promptly and no later than 15 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 4. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) and ensure the letter is in a place likely to be seen/viewed. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax. FDA recall Z-0284-2024 is Completed for Philips Affiniti 70, Diagnostic Ultrasound System Model # 795210. FDA reason: Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image. FDA action: On October 9, 2023, Philips issued a Urgent Medical Device Correction notification to affected consignees. Philips asked consignees to take the following actions: 1. Identify if your firm has an affected product. 2. In the 2D tab in the system touchscreen, disable the AutoSCAN feature every time a transducer or a Transducer preset is selected to prevent the system from potentially experiencing image lag issue. The AutoSCAN feature is disabled when the button is not highlighted in amber. 3. Please complete and return the attached response form to Philips promptly and no later than 15 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 4. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) and ensure the letter is in a place likely to be seen/viewed. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax. FDA recall Z-1007-2026 is Completed for Philips EPIQ Ultrasound Systems (various models) used in conjunction with the X5-1c transducer REF: , , 795200, 795201, 795231, 795232, 795234 REF: 795117, 795122. FDA reason: Retroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer. FDA action: On July 18, 2022, Phillips issued a product notification to affected consignees. Philips ask consignees to take the following actions: 1. If the X5-1c transducer is connected to the right-most port (when viewing from the front), this problem will not occur. Please ensure your X5-1c is connected to the right-most connector until the software update (FCO79500561) has been installed. 2. Please retain this letter with your product documentation until the software update (FCO79500561) has been implemented on your system. 3. Please circulate this notice to all users of this device so that they are aware of the issue. 4. A Philips Field Service Engineer will contact you to schedule the installation of the software update to resolve this issue (reference FCO79500561).
FDA source Exact FDA submission Checked 2026-09-04

Each published event was manually matched to the exact FDA submission and affected product or version; a recall does not establish an AI-algorithm defect or current site exposure unless the event states that scope.

recall: Z-2386-2025

Open, Classified

2025-08-22

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; the FDA record does not identify an AI-algorithm defect unless it says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-0283-2024

Completed

2023-11-09

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; the FDA record does not identify an AI-algorithm defect unless it says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-0284-2024

Completed

2023-11-09

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; the FDA record does not identify an AI-algorithm defect unless it says so. Confirm affected model, version, units, corrective action, and completion status locally.

recall: Z-1007-2026

Completed

2025-12-31

Exact native submission match in the checked openFDA snapshot. Published as a product, system, labeling, or postmarket issue; the FDA record does not identify an AI-algorithm defect unless it says so. Confirm affected model, version, units, corrective action, and completion status locally.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
25Fields reviewed
11Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact FDA decision materials were reviewed for identity, indication, users, workflow, limitations, and submitted performance. Current vendor material is labeled product-family context where it does not establish the exact release. PubMed and ClinicalTrials.gov results are retained as candidate evidence with version applicability pending; CMS reimbursement context and exact-identifier openFDA postmarket results were also checked. Pricing, final security architecture, contracted services, enabled options, and local interoperability remain organization-specific evidence. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. 4 FDA recall candidates were manually reviewed: 4 published and 0 rejected as a neighboring or otherwise inapplicable version. Native FDA recall identifiers produced 4 postmarket candidate records; 4 have been reviewed (4 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-04.

Alternatives

Related tools to compare

FDA-cleared K251985

LOGIQ E10

Ge Medical Systems Ultrasound and Primary Care Diagnostics

LOGIQ E10 is a general diagnostic ultrasound system for qualified users across fetal and obstetric, abdominal, gynecologic, pediatric, small-organ, breast, cardiac, vascular,...

Modality
Ultrasound
Anatomy
Multi-region
Decision
2025-10-29
FDA-cleared K172385

Clarius Ultrasound System

Clarius Mobile Health Corp.

Clarius Ultrasound Scanner is a software-based diagnostic ultrasound imaging system for trained healthcare professionals across labeled abdominal, thoracic, musculoskeletal,...

Modality
Ultrasound
Anatomy
Multi-region
Decision
2017-09-14