Release and scope verification
Map the installed model, build, options, population, anatomy, inputs, and outputs to K213779 before acceptance.
exact submission · checked 2026-09-03 Later product-family features are not evidence that this cleared release includes them.
Interoperability acceptance
DICOM CT, 3D CAD visualization, implant library, server case management, and approval workflow; validate geometry, catalog version, revision control, physician sign-off, and surgical access.
exact submission · checked 2026-09-03 Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.
Clinical acceptance
Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Not for diagnosis, spine surgery, or implant and instrument design. The FDA summary reports nonclinical validation but does not disclose numerical segmentation, repeatability, or 3D-model accuracy results.
exact submission · checked 2026-09-03 Define stop-use, escalation, rollback, and revalidation triggers before production use.
Economic evaluation
Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.
product family · checked 2026-09-03 Vendor productivity claims are scenario inputs, not guaranteed local savings.
Lifecycle monitoring
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track the limits that matter for this product: Not for diagnosis, spine surgery, or implant and instrument design. The FDA summary reports nonclinical validation but does not disclose numerical segmentation, repeatability, or 3D-model accuracy results.
exact submission · checked 2026-09-03 Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.