Briefcase-Quantification is a radiological image management and processing system software indicated for use in the analysis of CT exams with contrast, that include the abdominal aorta, in adults or transitional adolescents aged 18 and older. The device is intended to assist appropriately trained medical specialists by providing the user with the maximum abdominal aortic diameter measurement of cases that include the abdominal aorta (MAbdAo ). Briefcase-Quantification is indicated to evaluate normal and aneurysmal abdominal aortas and is not intended to evaluate postoperative aortas. The Briefcase-Quantification results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of cases. These measurements are unofficial, are not final, and are subject to change after review by a radiologist. For final clinically approved measurements, please refer to the official radiology report. Clinicians are responsible for viewing full images per the standard of care.
Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.
Clinical fit
What this tool is for
Start with the authorized purpose, then verify how it fits your service line and reading workflow.
Exact purpose
Briefcase-Quantification is a radiological image management and processing system software indicated for use in the analysis of CT exams with contrast, that include the abdominal aorta, in adults or transitional adolescents aged 18 and older. The device is intended to assist appropriately trained medical specialists by providing the user with the maximum abdominal aortic diameter measurement of cases that include the abdominal aorta (MAbdAo ). Briefcase-Quantification is indicated to evaluate normal and aneurysmal abdominal aortas and is not intended to evaluate postoperative aortas. The Briefcase-Quantification results are not intended to be used on a stand-alone basis for clinical decision-making or otherwise preclude clinical assessment of cases. These measurements are unofficial, are not final, and are subject to change after review by a radiologist. For final clinically approved measurements, please refer to the official radiology report. Clinicians are responsible for viewing full images per the standard of care.
Public source Exact FDA submission Checked 2026-09-01
Exact-submission indication text from the ACR AI Central model card. FDA labeling remains controlling.
CT images or data. Inclusion criteria: CT exams with contrast that include the abdominal aorta; adults or transitional adolescents aged 18 and older; normal and aneurysmal abdominal aortas
Public source Exact FDA submission Checked 2026-09-01
Conservative normalization of explicit wording in an exact-submission source.
The AI model outputs the maximum abdominal aortic diameter measurement. It produces a preview image annotated with the maximum diameter measurement, presenting the key slice with the long and short axis of the aorta in the reformatted slice composed of the plane orthogonal to the blood flow. The output is exported in DICOM format and sent to PACS for review. The measurement is unofficial and not final.
Public source Exact FDA submission Checked 2026-09-01
Clinical output reported in the exact-submission ACR model card.
The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.
FDA submission
K241112
FDA source Exact FDA submission Checked 2026-08-31
Use these fields to structure a vendor demo, security review, and implementation estimate.
Integration
The exact-submission ACR model card explicitly references DICOM and PACS in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
Public source Exact FDA submission Checked 2026-09-01
Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.
Public source Exact FDA submission Checked 2026-08-13
Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.
Vendor confirmation Exact FDA submission Checked 2026-08-13
FDA authorization does not establish the current commercial security posture. Request data-flow diagrams, hosting and retention details, access controls, audit logging, attestations, and incident-response documentation.
Training and support
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13
Training, onboarding, competency, and support commitments are commercial implementation details that require current vendor documentation.
Monitoring and change control
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13
Local acceptance testing, production monitoring, drift review, update control, and incident escalation require a current vendor and site-specific governance plan.
Input or output interoperability
DICOM is explicitly referenced in the described input, output, or workflow.
exact submission · checked 2026-09-01
The model card does not establish every supported DICOM object, transfer method, or local interface.
Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.
Evidence summary
Study type: Stand-Alone Performance. Cases: 162. Sites: 6. Reference standard: Interpretation by Reviewing Clinician. Additional reported metrics: The algorithm performance showed that the mean absolute error between the ground truth measurement and algorithm was 1.52 mm (95% CI: 1.20 mm, 1.83 mm).
Public source Exact FDA submission Checked 2026-09-01
Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.
Needs research Exact FDA submission Checked 2026-08-31
This field has not been curated yet.
Reported specificity
Research pending
Needs research Exact FDA submission Checked 2026-08-31
This field has not been curated yet.
No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.
BriefCase-Quantification of the Abdominal Aortic Measurement (M-AbdAo) regulatory performance study summary
Stand-Alone Performance · Population described in the exact-submission ACR model card
exact submission · Tested version: 2 · n=162 · 6 sites · Independence not established
K241112
ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.
These are common procurement questions; unknown values remain visible until a source supports them.
Pricing and total cost
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13
FDA submission materials do not establish current commercial pricing. Request the pricing unit, term, minimum commitment, implementation fees, and included support from the vendor.
Reimbursement and coding
Research pending
Needs research Exact FDA submission Checked 2026-08-31
Determine whether any product-specific CPT/HCPCS pathway exists and separate it from billing for the underlying imaging service.
Safety and lifecycle
Postmarket record
Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.
Postmarket safety review
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04
No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.
Buyer worksheet
Open questions to take to the vendor
Verify care setting and population: Confirm the intended care setting, patient population, and service line.
Verify security and privacy controls: FDA authorization does not establish the current commercial security posture. Request data-flow diagrams, hosting and retention details, access controls, audit logging, attestations, and incident-response documentation.
Verify training: Training, onboarding, competency, and support commitments are commercial implementation details that require current vendor documentation.
Verify pricing and total cost: FDA submission materials do not establish current commercial pricing. Request the pricing unit, term, minimum commitment, implementation fees, and included support from the vendor.
Verify reimbursement and coding relevance: Determine whether any product-specific CPT/HCPCS pathway exists and separate it from billing for the underlying imaging service.
Verify local validation and lifecycle monitoring: Local acceptance testing, production monitoring, drift review, update control, and incident escalation require a current vendor and site-specific governance plan.
Verify recalls and postmarket safety signals: No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.
This audit distinguishes completed source review from fields that have not yet been researched.
Exact FDA record reviewedStatus
2026-09-04Last searched
17Fields reviewed
6Source classes checked
0Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads
Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Automated exact-name discovery found 0 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.
Candidate leads remain unpublished until a human confirms the exact product and tested version.
Source classes: fda ai list, fda decision summary, acr ai central product, pubmed, clinical trials, openfda device recall
PRAEVAorta 2 assists clinical specialists, physicians, and other licensed practitioners with automatic vascular segmentation and maximum-diameter measurement on adult CT images;...
TruPlan provides CT image processing, visualization, measurements, and cardiac procedure-planning workflow support for the labeled use; exact procedure scope and outputs are...