FDA-cleared profile

BriefCase

Aidoc Medical, Ltd.

Authorized purpose: BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced CT and CTPA images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of Intracranial Hemorrhage and Pulmonary Embolism pathologies. For the PE pathology, the software is only intended to be used on single-energy exams.

Evidence status: each field states its source quality, applicability, and review date. Research pending, information not established, and vendor documentation pending remain distinct outcomes.

Product overview

Start with the essentials

Authorized purpose: BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced CT and CTPA images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of Intracranial Hemorrhage and Pulmonary Embolism pathologies. For the PE pathology, the software is only intended to be used on single-energy exams.

Overview status: Source-reviewed overview

Clinical takeaway

Clinical role: BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced CT and CTPA images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of Intracranial Hemorrhage and Pulmonary Embolism pathologies. For the PE pathology, the software is only intended to be used on single-energy exams.

Evidence takeaway

Evidence from the reviewed source record: Study type: Stand-Alone Performance. Cases: 184. Sites: 3. Reported sensitivity: 90.6% (95% CI: 82.2%-95.9%). Reported specificity: 89.9% (95% CI: 82.2%-95.1%)

Clinical details

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of non-enhanced CT and CTPA images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive findings of Intracranial Hemorrhage and Pulmonary Embolism pathologies. For the PE pathology, the software is only intended to be used on single-energy exams.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-08
Intended users
Radiologist
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Care setting and population
Research pending
  • Needs research
  • Exact FDA submission
  • Checked 2026-08-31

Confirm the intended care setting, patient population, and service line.

Workflow role
Triage or notification support
  • Public source
  • Exact FDA submission
  • Checked 2026-08-31

Conservative normalization of explicit wording in an exact-submission source.

Required input
non-enhanced head CT and CTPA images; for the PE pathology, the software is only intended to be used on single-energy exams.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication states the two image types and the single-energy boundary for PE.

Output and human action
Triage and flag suspected positive findings of Intracranial Hemorrhage and Pulmonary Embolism pathologies.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Limitations
Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

The exact indication states the preview-image, diagnostic-use, and unchanged-image boundaries.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K190072
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Modality
CT/CTA
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Anatomy
Brain; Chest/Lung
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Clearance type
510(k)
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
Decision date
2019-04-15
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31
FDA status
FDA-cleared
  • FDA source
  • Exact FDA submission
  • Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
Standards-based DICOM/PACS integration was identified in the FDA submission materials; compatibility with a specific PACS, RIS, EHR, or orchestrator requires confirmation.
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Normalized from the prior exact-submission review. Confirm interface versions and local compatibility during technical evaluation.

Deployment and data flow
On-cloud
  • Public source
  • Exact FDA submission
  • Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Exact-release security controls, data retention, hosting boundaries, and scoped assurance documentation require vendor confirmation.

Training and support
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current training prerequisites, competency checks, and support commitments for the exact configuration require vendor confirmation.

Monitoring and change control
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

Current drift, quality, uptime, alert, escalation, and incident-response commitments for the exact configuration require vendor confirmation.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

ACR exact-submission catalog context

ACR AI Central provides an exact-submission model card for BriefCase; FDA labeling remains controlling for clinical use, limitations, and performance.

exact submission · checked 2026-09-08

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version. Candidate literature without exact product and version linkage is not shown as product evidence.

Evidence summary
Study type: Stand-Alone Performance. Cases: 184. Sites: 3. Reported sensitivity: 90.6% (95% CI: 82.2%-95.9%). Reported specificity: 89.9% (95% CI: 82.2%-95.1%)
  • FDA source
  • Exact FDA submission
  • Checked 2026-09-19

Exact FDA decision-summary packet reviewed on 2026-09-19; existing wording is supported without expansion.

Reported sensitivity
90.6% (95% CI: 82.2%-95.9%)
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
89.9% (95% CI: 82.2%-95.1%)
  • Public source
  • Exact FDA submission
  • Checked 2026-09-01

Model specificity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

BriefCase-ICH and PE triage regulatory performance study summary

  • DesignStand-Alone Performance
  • PopulationPopulation described in the exact-submission ACR model card
  • Scopeexact submission
  • Tested versionExact product version not reported; regulatory study summarized for FDA submission K190072
  • Samplen=184
  • Sites3
  • IndependenceIndependence not established

K190072

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Costs and setup

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
  • Vendor confirmation
  • Exact FDA submission
  • Checked 2026-09-08

No public price or quote for the exact K190072 configuration was established in this review.

Reimbursement and coding
Not established in reviewed sources
  • Reviewed sources checked
  • Exact FDA submission
  • Checked 2026-09-08

No exact-product payer policy, coding instruction, or payment determination was established; eligibility must be evaluated by payer, site of service, and use case.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The exact-identifier FDA recall query returned no native recall record for K190072. This limited result does not establish absence of recalls or other safety information and does not cover later versions, family records, MAUDE reports, corrections, or field notices.
  • Public source
  • Exact FDA submission
  • Checked 2026-09-08

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-20Last searched
26Fields reviewed
10Source classes checked
20Screened PubMed leads
6Screened trial leads
0Unreviewed FDA recall leads

The refreshed discovery snapshot contains 20 PubMed and 6 ClinicalTrials.gov candidate records. Discovery metadata alone does not establish exact tested-product, version, population, and endpoint applicability, so all candidates remain quarantined and none was promoted. The current discovery snapshot contains 20 PubMed and 6 ClinicalTrials.gov candidate records; its dated, count-matched human screening remains current. Unlinked candidates remain quarantined. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads were screened against the exact product and tested version. Unlinked records remain quarantined.

Source classes: fda ai list, fda decision summary, acr ai central product, fda device recall, literature index, trial registry, fda decision summary packet, pubmed, clinical trials, openfda device recall

Sources & history

How this profile was documented

Open the ledger for source dates, scope, and research-record updates.

View source ledger and history

Source ledger

  1. FDA K190072: BriefCase

    U.S. Food and Drug Administration - Published 2019-04-15 - Accessed 2026-09-19

    Scope: identity, clinicalUse, modality, anatomy, workflow, evidence, implementation, purpose, intendedUsers, output, evidenceSummary, input, limitations

  2. Artificial Intelligence-Enabled Medical Devices

    U.S. Food and Drug Administration - Accessed 2026-08-31

    Scope: regulatoryIdentity, modalityContext

  3. BriefCase-ICH and PE triage model card (K190072)

    American College of Radiology - Published 2026-06-26 - Accessed 2026-09-01

    Scope: intendedUsers, output, limitations, reportedSensitivity, reportedSpecificity, evidenceSummary, anatomy, evidenceStudies, implementationCapabilities

  4. PubMed biomedical literature index

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery

  5. ClinicalTrials.gov study registry

    U.S. National Library of Medicine - Accessed 2026-09-01

    Scope: evidenceDiscovery, studyStatus

  6. openFDA Device Recall API

    U.S. Food and Drug Administration - Accessed 2026-09-02

    Scope: recallDiscovery, nativeSubmissionMatching

  7. openFDA exact recall query for K190072

    U.S. Food and Drug Administration - Accessed 2026-09-08

    Scope: postmarketSafety

Research history

  1. K190072 FDA decision

    Regulatory record

    510(k)

  2. Source-review record updated

    Research record

    fda record reviewed

  3. Reader overview source-reviewed

    Editorial review

    Source-reviewed editorial overlay

Alternatives

Related tools to compare

FDA-cleared K203508

BriefCase

Aidoc Medical, Ltd.

BriefCase analyzes supported noncontrast head CT, head CTA, noncontrast cervical-spine CT, and CTPA examinations to triage suspected intracranial hemorrhage, large-vessel...

Modality
CT/CTA
Anatomy
Adult head, brain, cervical spine, pulmonary arteries, and thorax
Decision
2020-12-29
FDA-cleared K252421

JLK-NCCT

JLK, Inc.

Notification-only triage of adult non-contrast head CT for suspected intracranial hemorrhage and large-vessel occlusion, with parallel communication to trained clinicians during...

Modality
CT/CTA
Anatomy
Brain and intracranial vessels: intracranial hemorrhage and LVO involving ICA, MCA-M1, and MCA-M2
Decision
2026-03-24
FDA-cleared K251983

Brainomix 360 Triage Stroke

Brainomix Limited

Triage and notification of suspected intracranial hemorrhage or anterior-circulation LVO on acute non-contrast head CT; not primary diagnosis.

Modality
CT/CTA
Anatomy
Brain/head, including intracranial ICA and M1 for LVO and intracranial hemorrhage findings
Decision
2025-08-26