FDA-cleared profile

AutoDensity

Eos Imaging

Software estimates L1-L4 spine bone-mineral density from EOSedge dual-energy AP radiographs for opportunistic orthopedic pre-surgical assessment.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
Software estimates L1-L4 spine bone-mineral density from EOSedge dual-energy AP radiographs for opportunistic orthopedic pre-surgical assessment.
FDA source Exact FDA submission Checked 2026-09-03
Intended users
Qualified imaging professionals and physicians trained for the represented anatomy and workflow.
FDA source Exact FDA submission Checked 2026-09-03
Care setting and population
Hospital or outpatient imaging analysis and reporting environments.
FDA source Exact FDA submission Checked 2026-09-03
Workflow role
Medical-image processing, measurement, quantification, visualization, and reporting support.
FDA source Exact FDA submission Checked 2026-09-03
Required input
Supported DICOM images and release-specific user inputs.
FDA source Exact FDA submission Checked 2026-09-03
Output and human action
Measurements, segmentations, visualizations, calculations, and reports for professional review.
FDA source Exact FDA submission Checked 2026-09-03
Limitations
AutoDensity is not a DXA replacement; patients with suspected low bone density require confirmation using the standard clinical pathway.
FDA source Exact FDA submission Checked 2026-09-03

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K251078
FDA source Exact FDA submission Checked 2026-08-31
Modality
X-ray/Radiography
FDA source Exact FDA submission Checked 2026-09-03
Anatomy
Spine
FDA source Exact FDA submission Checked 2026-09-03
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2025-11-14
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
DICOM import and output to PACS, reporting, or downstream systems; validate identifiers, series selection, measurements, units, geometry, routing, and round trips.
Public source Exact FDA submission Checked 2026-09-03
Deployment and data flow
Software on a workstation, server, or supported virtual environment with release-specific licenses.
Public source Exact FDA submission Checked 2026-09-03
Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified imaging professionals and physicians trained for the represented anatomy and workflow. should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: AutoDensity is not a DXA replacement; patients with suspected low bone density require confirmation using the standard clinical pathway.
Public source Exact FDA submission Checked 2026-09-03

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, and outputs to K251078 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

DICOM import and output to PACS, reporting, or downstream systems; validate identifiers, series selection, measurements, units, geometry, routing, and round trips.

exact submission · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. AutoDensity is not a DXA replacement; patients with suspected low bone density require confirmation using the standard clinical pathway.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

Vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software and model version, patient population, anatomy, protocol, operator, input quality, output edits or overrides, failed studies, repeat acquisition or rework, downstream impact, downtime, incidents, recalls, and updates. Track these product-specific limits: AutoDensity is not a DXA replacement; patients with suspected low bone density require confirmation using the standard clinical pathway.

exact submission · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
K251078 reports a 30-subject precision study and 253-subject agreement study. Within-subject CV was 2.23%; mean bias versus reference was 0.045 g/cm2 with limits of agreement from -0.088 to 0.178 g/cm2.
FDA source Exact FDA submission Checked 2026-09-03
Reported sensitivity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Sensitivity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Specificity is not the primary endpoint for this acquisition, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Within-subject coefficient of variation

2.23 %

Repeat AutoDensity precision in AutoDensity.

Tested version: AutoDensity release cleared in K251078

Mean BMD bias

0.045 g/cm2

Agreement with reference in AutoDensity.

Tested version: AutoDensity release cleared in K251078

K251078 FDA performance package for AutoDensity

Manufacturer precision and agreement studies · Adults undergoing EOSedge spine radiography and reference bone-density assessment

exact submission · Tested version: AutoDensity release cleared in K251078 · n=253 · Sponsor or vendor study

K251078

Manufacturer evidence submitted for clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, enabled features, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS coverage material. Evaluate the product within the applicable imaging, procedural, or surgical service line and verify payer, site-of-service, and coding assumptions independently.
Public source Not applicable Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K251078 by exact submission identifier. This does not exclude product-family or later safety information.
Public source Exact FDA submission Checked 2026-09-03

Exact-identifier review does not prove absence of all product-family safety signals.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
24Fields reviewed
9Source classes checked
5Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact FDA materials, available current product-family sources, reimbursement context, and exact-identifier postmarket results were reviewed. Pricing, contracted services, enabled options, final security architecture, and local interoperability remain organization-specific evidence. Automated exact-name discovery found 5 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-03.

Alternatives

Related tools to compare

FDA-cleared K241593

BoneMetrics (US)

Gleamer SAS

BoneMetrics is fully automated software that assists trained radiology professionals with Cobb-angle measurement on frontal spine radiographs in patients aged four years or older...

Modality
X-ray/Radiography
Anatomy
Spine
Decision
2025-02-05
FDA-cleared K231668

Spine CAMP™ (1.1)

Medical Metrics, Inc.

FDA product code QIH identifies this device class as Automated Radiological Image Processing Software. To provide automated radiological image processing and analysis tools....

Modality
X-ray/Radiography
Anatomy
Spine
Decision
2023-07-07
FDA-cleared K221632

Spine CAMP™

Medical Metrics, Inc.

FDA product code QIH identifies this device class as Automated Radiological Image Processing Software. To provide automated radiological image processing and analysis tools....

Modality
X-ray/Radiography
Anatomy
Spine
Decision
2022-10-18
FDA-cleared K212005

UNiD Spine Analyzer

MEDICREA International, Inc.

FDA product code LLZ identifies this device class as System, Image Processing, Radiological. Consult the linked FDA record for the exact authorized indications for use.

Modality
X-ray/Radiography
Anatomy
Spine
Decision
2022-01-12