FDA-cleared profile

AI-Rad Companion (Pulmonary)

Siemens Healthcare GmBh

AI-Rad Companion (Pulmonary) is image processing software that provides quantitative and qualitative analysis from previously acquired Computed Tomography DICOM images to support radiologists and physicians from emergency medicine, specialty care, urgent care, and general practice in the evaluation and assessment of disease of the lungs. It provides the following functionality: Segmentation and measurements of complete lung and lung lobes Identification of areas with lower Hounsfield values in comparison to a predefined threshold for complete lung and lung lobes Providing an interface to external Medical Device syngo.CT Lung CAD Segmentation and measurements of solid lung nodules. Dedication of found lung nodules to corresponding lung lobe. Correlation of segmented lung nodules of current scan with known priors and quantitative assessment of changes of the correlated data. Identification of areas with elevated Hounsfield values, where areas with elevated versus high opacities are distinguished. The software has been validated for data from Siemens Healthineers (filtered back projection and iterative reconstruction),GE Healthcare (filtered back projection reconstruction), and Philips (filtered back projection reconstruction).Only DI COM images of adult patients are considered to be valid input.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
AI-Rad Companion (Pulmonary) is image processing software that provides quantitative and qualitative analysis from previously acquired Computed Tomography DICOM images to support radiologists and physicians from emergency medicine, specialty care, urgent care, and general practice in the evaluation and assessment of disease of the lungs. It provides the following functionality: Segmentation and measurements of complete lung and lung lobes Identification of areas with lower Hounsfield values in comparison to a predefined threshold for complete lung and lung lobes Providing an interface to external Medical Device syngo.CT Lung CAD Segmentation and measurements of solid lung nodules. Dedication of found lung nodules to corresponding lung lobe. Correlation of segmented lung nodules of current scan with known priors and quantitative assessment of changes of the correlated data. Identification of areas with elevated Hounsfield values, where areas with elevated versus high opacities are distinguished. The software has been validated for data from Siemens Healthineers (filtered back projection and iterative reconstruction),GE Healthcare (filtered back projection reconstruction), and Philips (filtered back projection reconstruction).Only DI COM images of adult patients are considered to be valid input.
Public source Exact FDA submission Checked 2026-09-01

Exact-submission indication text from the ACR AI Central model card. FDA labeling remains controlling.

Intended users
Radiologists, Physicians
Public source Exact FDA submission Checked 2026-09-01

Intended-user text reported in the exact-submission ACR model card.

Care setting and population
Adult
Public source Exact FDA submission Checked 2026-09-01

Patient-population text reported in the exact-submission ACR model card; institutional setting may still require vendor confirmation.

Workflow role
Segmentation or contouring
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Required input
CT images or data. Inclusion criteria: DICOM images of adult patients
Public source Exact FDA submission Checked 2026-09-01

Conservative normalization of explicit wording in an exact-submission source.

Output and human action
Segmentation and measurements of complete lung and lung lobes Identification of areas with lower Hounsfield values in comparison to a predefined threshold for complete lung and lung lobes Providing an interface to external Medical Device syngo.CT Lung CAD Segmentation and measurements of solid lung nodules Dedication of found lung nodules to corresponding lung lobe Correlation of segmented lung nodules of current scan with known priors and quantitative assessment of changes of the correlated data. Identification of areas with elevated Hounsfield values, where areas with elevated versus high opacities are distinguished
Public source Exact FDA submission Checked 2026-09-01

Clinical output reported in the exact-submission ACR model card.

Limitations
Scanner manufacturers: All
Public source Exact FDA submission Checked 2026-09-01

Limitations and supported-acquisition details reported in the exact-submission ACR model card.

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K213713
FDA source Exact FDA submission Checked 2026-08-31
Modality
CT/CTA
Public source Exact FDA submission Checked 2026-09-01

Imaging modality reported in the exact-submission ACR AI Central model card.

Anatomy
Chest/Lung
Public source Exact FDA submission Checked 2026-09-01

Anatomy normalized from explicit exact-submission ACR model-card wording.

Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2022-08-11
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
The exact-submission ACR model card explicitly references DICOM in the described input, output, or workflow; exact interfaces, versions, and local compatibility require confirmation.
Public source Exact FDA submission Checked 2026-09-01

Interoperability terms explicitly reported in the exact-submission ACR model card; interface versions and local compatibility still require confirmation.

Deployment and data flow
Hybrid
Public source Exact FDA submission Checked 2026-08-13

Deployment category preserved from the prior exact-submission review; hosting region, data flow, and current commercial configuration still require vendor confirmation.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA authorization does not establish the current commercial security posture. Request data-flow diagrams, hosting and retention details, access controls, audit logging, attestations, and incident-response documentation.

Training and support
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Training, onboarding, competency, and support commitments are commercial implementation details that require current vendor documentation.

Monitoring and change control
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

Local acceptance testing, production monitoring, drift review, update control, and incident escalation require a current vendor and site-specific governance plan.

Input or output interoperability

DICOM is explicitly referenced in the described input, output, or workflow.

exact submission · checked 2026-09-01

The model card does not establish every supported DICOM object, transfer method, or local interface.

Supported acquisition environment

Scanner manufacturers: All

exact submission · checked 2026-09-01

Confirm currently supported software versions, protocols, and site-specific acquisition configurations.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
Study type: Stand-Alone Performance. Cases: 200. Reported sensitivity: 94.30%. Additional reported metrics: PPV: 99.1%
Public source Exact FDA submission Checked 2026-09-01

Structured study metadata from the exact-submission ACR model card; this is not an independent evidence synthesis.

Reported sensitivity
94.30%
Public source Exact FDA submission Checked 2026-09-01

Model sensitivity reported by ACR AI Central. Review endpoint, threshold, study population, and tested version before comparison.

Reported specificity
Research pending
Needs research Exact FDA submission Checked 2026-08-31

This field has not been curated yet.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

AI-Rad Companion (Pulmonary) regulatory performance study summary

Stand-Alone Performance · Adult

exact submission · Tested version: Exact product version not reported; regulatory study summarized for FDA submission K213713 · n=200 · Independence not established

K213713

ACR AI Central summarizes the regulatory study; consult the exact FDA materials before comparative use.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-08-13

FDA submission materials do not establish current commercial pricing. Request the pricing unit, term, minimum commitment, implementation fees, and included support from the vendor.

Reimbursement and coding
Research pending
Needs research Exact FDA submission Checked 2026-08-31

Determine whether any product-specific CPT/HCPCS pathway exists and separate it from billing for the underlying imaging service.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-04

No native submission-identifier match was found in the checked openFDA radiology recall snapshot. Product-name, manufacturer, and adverse-event searches remain necessary; this does not establish that no recall or safety report exists.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Exact FDA record reviewedStatus
2026-09-04Last searched
19Fields reviewed
6Source classes checked
20Unreviewed PubMed leads
1Unreviewed trial leads
0Unreviewed FDA recall leads

Exact-submission ACR model-card fields were normalized under the current provenance rules; vendor and independent-study confirmation remain distinct. Automated exact-name discovery found 20 PubMed and 1 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, acr ai central product, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-02.

Alternatives

Related tools to compare

FDA-cleared K254075

Synapse Lung Nodule AI

Fujifilm Corporation

FDA product code OEB identifies this device class as Lung Computed Tomography System, Computer-Aided Detection. To assist radiologists in the review of multi-slice computed...

Modality
CT/CTA
Anatomy
Chest/Lung
Decision
2026-05-27
FDA-cleared K251203

AVIEW Lung Nodule CAD

Coreline Soft Co.,Ltd

FDA product code OEB identifies this device class as Lung Computed Tomography System, Computer-Aided Detection. To assist radiologists in the review of multi-slice computed...

Modality
CT/CTA
Anatomy
Chest/Lung
Decision
2025-12-03