FDA-cleared profile

AccuContour

Xiamen Manteia Technology LTD.

AccuContour generates initial organ-at-risk contours from supported multimodality images for physician review in radiation-treatment planning and evaluation workflows.

Evidence status: each field shows source quality, applicability, and review date. Research pending, information not established in reviewed sources, and vendor documentation pending are distinct outcomes.

Clinical fit

What this tool is for

Start with the authorized purpose, then verify how it fits your service line and reading workflow.

Exact purpose
AccuContour generates initial organ-at-risk contours from supported multimodality images for physician review in radiation-treatment planning and evaluation workflows.
FDA source Exact FDA submission Checked 2026-09-03
Intended users
Qualified radiation oncologists, medical physicists, dosimetrists, and radiation therapists.
FDA source Exact FDA submission Checked 2026-09-03
Care setting and population
Radiation-oncology departments using compatible planning, imaging, and treatment-delivery systems.
FDA source Exact FDA submission Checked 2026-09-03
Workflow role
Radiotherapy contouring, treatment planning, adaptation, or treatment-session management under professional review.
FDA source Exact FDA submission Checked 2026-09-03
Required input
Supported CT or MR images, structures, prescriptions, plans, registrations, and treatment-session data.
FDA source Exact FDA submission Checked 2026-09-03
Output and human action
Contours, treatment plans, adapted plans, dose or evaluation results, or session-management information requiring review and approval.
FDA source Exact FDA submission Checked 2026-09-03
Limitations
Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
FDA source Exact FDA submission Checked 2026-09-03

Regulatory identity

FDA record and catalog context

The FDA listing establishes the regulatory identity. It does not by itself establish local workflow fit, pricing, security, or performance in your environment.

FDA submission
K191928
FDA source Exact FDA submission Checked 2026-08-31
Modality
Radiotherapy Planning Imaging
FDA source Exact FDA submission Checked 2026-09-03
Anatomy
Multi-region
FDA source Exact FDA submission Checked 2026-09-03
Clearance type
510(k)
FDA source Exact FDA submission Checked 2026-08-31
Decision date
2020-02-28
FDA source Exact FDA submission Checked 2026-08-31
FDA status
FDA-cleared
FDA source Exact FDA submission Checked 2026-08-31

Implementation

Questions for IT, informatics, and operations

Use these fields to structure a vendor demo, security review, and implementation estimate.

Integration
DICOM RT and vendor workflow integration; validate patient identity, image geometry, structures, dose, plan transfer, approvals, version compatibility, and rollback.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K191928; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Deployment and data flow
Licensed software connected to compatible radiotherapy planning and delivery infrastructure.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K191928; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Security and privacy
Vendor documentation pending
Vendor confirmation Exact FDA submission Checked 2026-09-03

Public materials reviewed for this exact release do not establish the complete security posture. Request data-flow, identity, audit, encryption, retention, remote-access, vulnerability-management, backup, and incident-response evidence.

Training and support
Only qualified radiation oncologists, medical physicists, dosimetrists, and radiation therapists should operate or interpret the product. Local onboarding should cover the exact release, supported population, limitations, failure examples, escalation, downtime, and competency documentation.
FDA source Exact FDA submission Checked 2026-09-03
Monitoring and change control
Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.
Public source Product family Checked 2026-09-03

Current vendor materials are product-family evidence for K191928; confirm the exact cleared version, enabled options, deployment, and local compatibility.

Release and scope verification

Map the installed model, build, options, population, anatomy, inputs, outputs, and enabled AI functions to K191928 before acceptance.

exact submission · checked 2026-09-03

Later product-family features are not evidence that this cleared release includes them.

Interoperability acceptance

DICOM RT and vendor workflow integration; validate patient identity, image geometry, structures, dose, plan transfer, approvals, version compatibility, and rollback.

product family · checked 2026-09-03

Test representative identifiers, orientation, geometry, measurements, units, routing, failure handling, and round trips locally. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Clinical acceptance

Use a signed local test set for the supported population and workflow. Include common and difficult cases, exclusions, acquisition variation, failed or low-quality inputs, reviewer overrides, and downstream effects. Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.

exact submission · checked 2026-09-03

Define stop-use, escalation, rollback, and revalidation triggers before production use.

Economic evaluation

Model capital and recurring cost against net staff time after review, throughput, repeat imaging or rework, room time, consumables, infrastructure, training, service, downtime, upgrade, and replacement costs.

product family · checked 2026-09-03

No exact-product price was found in the reviewed public material; vendor productivity claims are scenario inputs, not guaranteed local savings.

Lifecycle monitoring

Monitor by exact hardware, software, model, and algorithm version; patient population; anatomy; protocol; operator; input quality; output edits or overrides; failed studies; repeat acquisition or rework; downstream impact; downtime; incidents; recalls; and updates. Track these product-specific limits: Generated or modified contours and plans require qualified review, editing when needed, independent checks, and final approval before treatment.

product family · checked 2026-09-03

Revalidate after material software, hardware, model, protocol, interface, or clinical-policy changes. Vendor statements are current product-family context; confirm the exact cleared release and installed options.

Evidence

Performance evidence

Metrics are shown only when they are tied to a source, endpoint, population, and tested product version.

Evidence summary
K191928 reports contouring, registration, software, and workflow verification and validation for the cleared release. The contour is a starting point for professional review and is not a treatment decision or a universal segmentation-performance claim.
FDA source Exact FDA submission Checked 2026-09-03
Reported sensitivity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Sensitivity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

Reported specificity
Not applicable
Not applicable Not applicable Checked 2026-09-03

Specificity is not the primary endpoint for this system, reconstruction, segmentation, guidance, or quantitative function. Use the task-specific evidence below.

No contextualized exact-version metric has completed evidence review. Regulatory-document values, when available, are shown above with their limitations.

K191928 FDA performance package for AccuContour

Manufacturer contouring, registration, software, and radiotherapy-workflow verification and validation · Radiation-treatment planning and evaluation using supported multimodality images

exact submission · Tested version: AccuContour · Sponsor or vendor study

K191928

Manufacturer evidence submitted for FDA clearance or approval; local clinical, technical, and workflow acceptance remains necessary.

Interobserver delineation variability of computed tomography-based radiomic features of the parotid gland.

Peer-reviewed literature candidate identified by exact-name search; full product and version applicability review pending · Multi-region or related Radiotherapy Planning Imaging workflow; exact product and version linkage not established

product family · Tested version: Candidate source; exact FDA submission and product version not established · Independence not established

38549385

Candidate source was found by automated exact-name search; product or version linkage and clinical endpoint were not independently established, so it must not be treated as exact-submission performance evidence.

Evaluation Exploration of Atlas-Based and Deep Learning-Based Automatic Contouring for Nasopharyngeal Carcinoma.

Peer-reviewed literature candidate identified by exact-name search; full product and version applicability review pending · Multi-region or related Radiotherapy Planning Imaging workflow; exact product and version linkage not established

product family · Tested version: Candidate source; exact FDA submission and product version not established · Independence not established

35433460

Candidate source was found by automated exact-name search; product or version linkage and clinical endpoint were not independently established, so it must not be treated as exact-submission performance evidence.

Consistency evaluation and performance optimization of deep learning-based auto-contouring for nasopharyngeal carcinoma.

Peer-reviewed literature candidate identified by exact-name search; full product and version applicability review pending · Multi-region or related Radiotherapy Planning Imaging workflow; exact product and version linkage not established

product family · Tested version: Candidate source; exact FDA submission and product version not established · Independence not established

41436593

Candidate source was found by automated exact-name search; product or version linkage and clinical endpoint were not independently established, so it must not be treated as exact-submission performance evidence.

Economics and lifecycle

Budget and ongoing governance

These are common procurement questions; unknown values remain visible until a source supports them.

Pricing and total cost
Not established in reviewed sources
Reviewed sources checked Exact FDA submission Checked 2026-09-03

No public list price for the exact cleared configuration was found. Obtain a written quote separating hardware or license basis, implementation, interfaces, infrastructure, training, service, upgrades, renewal, consumables, and exit costs.

Reimbursement and coding
No separate named-product Medicare payment was identified in the reviewed CMS radiation-oncology material. Evaluate net planning or delivery value within the radiotherapy service line.
Public source Not applicable Checked 2026-09-03

No product-specific code

No separate named-product payment identified in reviewed public material

Centers for Medicare & Medicaid Services · United States

Build a local total-cost and value model; do not infer reimbursement from FDA clearance or approval.

Safety and lifecycle

Postmarket record

Recall and adverse-event records are shown only after product matching. Adverse-event reports do not establish incidence or causality.

Postmarket safety review
The 2026-09-02 openFDA device-recall snapshot contained no record matched to K191928 by exact submission identifier. This does not exclude product-family, later, or adverse-event information.
Public source Exact FDA submission Checked 2026-09-03

Exact-identifier review does not prove absence of all product-family safety signals.

Buyer worksheet

Open questions to take to the vendor

Open evaluation checklist

Research record

What has been checked

This audit distinguishes completed source review from fields that have not yet been researched.

Human reviewedStatus
2026-09-04Last searched
25Fields reviewed
11Source classes checked
7Unreviewed PubMed leads
0Unreviewed trial leads
0Unreviewed FDA recall leads

Recovered from the prior exact-submission extraction and normalized under the current provenance rules. Exact FDA decision materials were reviewed for identity, indication, users, workflow, limitations, and submitted performance. Current vendor material is labeled product-family context where it does not establish the exact release. PubMed and ClinicalTrials.gov results are retained as candidate evidence with version applicability pending; CMS reimbursement context and exact-identifier openFDA postmarket results were also checked. Pricing, final security architecture, contracted services, enabled options, and local interoperability remain organization-specific evidence. Automated exact-name discovery found 7 PubMed and 0 ClinicalTrials.gov candidate records. Candidates require human product and version matching; zero candidates is not evidence that no studies exist. Native FDA recall identifiers produced 0 postmarket candidate records; 0 have been reviewed (0 published, 0 rejected) and 0 remain unreviewed.

Candidate leads remain unpublished until a human confirms the exact product and tested version.

Source classes: fda ai list, fda decision summary, fda decision material, vendor product page, peer reviewed literature, clinical trial registry, reimbursement policy, fda device recall, pubmed, clinical trials, openfda device recall

Sources

Source ledger

Sources accessed through 2026-09-03.

Alternatives

Related tools to compare

FDA-cleared K260509

AutoContour (RADAC V5)

Radformation, Inc.

AutoContour Model RADAC V5 assists radiation treatment planners with contouring and reviewing structures in medical images for radiation therapy treatment planning. It produces...

Modality
Radiotherapy Planning Imaging
Anatomy
Multi-region
Decision
2026-03-19
FDA-cleared K253091

ART-Plan+ (v3.1.0)

Therapanacea Sas

ART-Plan+ assists trained radiation-therapy professionals with multimodal image review, registration, contouring, and the release-specific planning functions described in K253091....

Modality
Radiotherapy Planning Imaging
Anatomy
Multi-region
Decision
2025-12-23
FDA-cleared K242748

Oncospace

Oncospace, Inc.

Oncospace supports trained radiation therapy professionals with protocol configuration, machine-learning achievable-dose prediction, planning-objective exchange, and plan review...

Modality
Radiotherapy Planning Imaging
Anatomy
Multi-region
Decision
2025-04-11